Comparison of Continuous Glucose Monitoring Accuracy Between Abdominal and Upper Arm Insertion Sites

Steineck, Isabelle Isa Kristin, Mahmoudi, Zeinab, Ranjan, Ajenthen, Schmidt, Signe, Jørgensen, John Bagterp, Nørgaard, Kirsten · 2019 · Diabetes Technology & Therapeutics
Cited by 40

The DOI matches its Crossref record; 4 of the 4 references it cites were found in a registry or an index.

Abstract

Background: The aim was to compare the accuracy of the Dexcom ® G4 Platinum continuous glucose monitor (CGM) sensor inserted on the upper arm and the abdomen in adults. Methods: Fourteen adults with type 1 diabetes wore two CGMs, one placed on the upper arm and one placed on the abdomen. Three in-clinic visits of 5 h with YSI (2300 STAT, Yellow Springs Instrument) measurements as comparator were performed. Each visit was followed by 4 days with seven-point self-monitoring of blood glucose (SMBG) in free-living conditions. Accuracy analyses on the paired CGM-YSI and CGM-SMBG measurements of the two CGM sensors were performed. Results: Using YSI as comparator, the overall Mean Absolute Relative Difference (MARD) for the CGM abd was 12.3% and CGM arm was 12.0%. The percentage of the CGM measurements in zone A of Clarke error grid analysis for the CGM abd was 85.6% and CGM arm was 86.0%. The hypoglycemia sensitivity for the CGM abd and CGM arm was 69.3%. Using SMBG as comparator, the overall MARD for the CGM abd was 12.5% and CGM arm was 12.0%. The percentage of the CGM measurements in zone A for the CGM abd was 84.1% and the CGM arm was 85.0%. The hypoglycemia sensitivity for the CGM abd was 60.0% and the CGM arm was 71.1%. All the P -values from the comparisons between the accuracy of CGM abd and CGM arm were >0.05. Conclusion: The accuracy of a Dexcom G4 Platinum CGM sensor placed on the upper arm was not different from the accuracy of the sensor placed on the abdomen in adults with type 1 diabetes.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1089/dia.2019.0014

Cites

4 of 4 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Show all 4 references

Cited by

See all 40

Metadata

Authors
Steineck, I. I. K., Mahmoudi, Z., Ranjan, A., Schmidt, S., Jørgensen, J. B., & Nørgaard, K.
Venue
Diabetes Technology & Therapeutics
Published
2019-05
Volume
21
Issue
5
Pages
295-302
ISSN
  • 1520-9156 (ISSN-L)
  • 1557-8593
OpenAlex
W2936517447
Cited by
40
References
4
FWCI
3.44
Type
journal article
Language
en
Open access
closed

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Comparison of Continuous Glucose Monitoring Accuracy Between Abdominal and Upper Arm Insertion Sites. https://app.cochat.ai/papers/10.1089/dia.2019.0014
The paper
Steineck, I. I. K., Mahmoudi, Z., Ranjan, A., Schmidt, S., Jørgensen, J. B., & Nørgaard, K. (2019). Comparison of Continuous Glucose Monitoring Accuracy Between Abdominal and Upper Arm Insertion Sites. Diabetes Technology & Therapeutics. https://doi.org/10.1089/dia.2019.0014

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.