Comparison of Continuous Glucose Monitoring Accuracy Between Abdominal and Upper Arm Insertion Sites
The DOI matches its Crossref record; 4 of the 4 references it cites were found in a registry or an index.
Abstract
Background: The aim was to compare the accuracy of the Dexcom ® G4 Platinum continuous glucose monitor (CGM) sensor inserted on the upper arm and the abdomen in adults. Methods: Fourteen adults with type 1 diabetes wore two CGMs, one placed on the upper arm and one placed on the abdomen. Three in-clinic visits of 5 h with YSI (2300 STAT, Yellow Springs Instrument) measurements as comparator were performed. Each visit was followed by 4 days with seven-point self-monitoring of blood glucose (SMBG) in free-living conditions. Accuracy analyses on the paired CGM-YSI and CGM-SMBG measurements of the two CGM sensors were performed. Results: Using YSI as comparator, the overall Mean Absolute Relative Difference (MARD) for the CGM abd was 12.3% and CGM arm was 12.0%. The percentage of the CGM measurements in zone A of Clarke error grid analysis for the CGM abd was 85.6% and CGM arm was 86.0%. The hypoglycemia sensitivity for the CGM abd and CGM arm was 69.3%. Using SMBG as comparator, the overall MARD for the CGM abd was 12.5% and CGM arm was 12.0%. The percentage of the CGM measurements in zone A for the CGM abd was 84.1% and the CGM arm was 85.0%. The hypoglycemia sensitivity for the CGM abd was 60.0% and the CGM arm was 71.1%. All the P -values from the comparisons between the accuracy of CGM abd and CGM arm were >0.05. Conclusion: The accuracy of a Dexcom G4 Platinum CGM sensor placed on the upper arm was not different from the accuracy of the sensor placed on the abdomen in adults with type 1 diabetes.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
International consensus on use of continuous glucose monitoring
Danne, T., Nimri, R., Battelino, T., Bergenstal, R. M., Close, K. L., DeVries, J. H., Garg, S., Heinemann, L., Hirsch, I., Amiel, S. A., et al. (2017) · Diabetes Care · doi:10.2337/dc17-1600 · Cited by 2,069
Self-citationBeck, R. W., Riddlesworth, T., Ruedy, K., Ahmann, A., Bergenstal, R., Haller, S., Kollman, C., Kruger, D., McGill, J. B., Polonsky, W., et al. (2017) · JAMA · doi:10.1001/jama.2016.19975 · Cited by 1,185
Comparative accuracy of 17 point-of-care glucose meters
Ekhlaspour, L., Mondesir, D., Lautsch, N., Balliro, C., Hillard, M., Magyar, K., Radocchia, L. G., Esmaeili, A., Sinha, M., & Russell, S. J. (2017) · Journal of Diabetes Science and Technology · doi:10.1177/1932296816672237 · Cited by 127
Clinical Use of Continuous Glucose Monitoring in Adults with Type 1 Diabetes
Slattery, D., & Choudhary, P. (2017) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2017.0051 · Cited by 87
Cited by
Moser, O., Riddell, M. C., Eckstein, M. L., Adolfsson, P., Rabasa-Lhoret, R., van den Boom, L., Gillard, P., Nørgaard, K., Oliver, N. S., Zaharieva, D. P., et al. (2020) · Diabetologia · doi:10.1007/s00125-020-05263-9 · Cited by 214
Self-citationAccuracy and Safety of Dexcom G7 Continuous Glucose Monitoring in Adults with Diabetes
Garg, S. K., Kipnes, M., Castorino, K., Bailey, T. S., Akturk, H. K., Welsh, J. B., Christiansen, M. P., Balo, A. K., Brown, S. A., Reid, J. L., et al. (2022) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2022.0011 · Cited by 179
Longo, R. R., Elias, H., Khan, M., & Seley, J. J. (2022) · Journal of Diabetes Science and Technology · doi:10.1177/19322968211008446 · Cited by 89
Freckmann, G., Eichenlaub, M., Waldenmaier, D., Pleus, S., Wehrstedt, S., Haug, C., Witthauer, L., Jendle, J., Hinzmann, R., Thomas, A., et al. (2023) · Journal of Diabetes Science and Technology · doi:10.1177/19322968231190941 · Cited by 86
Faulds, E. R., Boutsicaris, A., Sumner, L., Jones, L., McNett, M., Smetana, K. S., May, C. C., Buschur, E., Exline, M. C., Ringel, M. D., et al. (2021) · The Journal of Clinical Endocrinology & Metabolism · doi:10.1210/clinem/dgab409 · Cited by 71
Metadata
- Authors
- Steineck, I. I. K., Mahmoudi, Z., Ranjan, A., Schmidt, S., Jørgensen, J. B., & Nørgaard, K.
- Venue
- Diabetes Technology & Therapeutics
- Publisher
- SAGE Publications
- Published
- 2019-05
- Volume
- 21
- Issue
- 5
- Pages
- 295-302
- ISSN
- 1520-9156 (ISSN-L)
- 1557-8593
- OpenAlex
- W2936517447
- Cited by
- 40
- References
- 4
- FWCI
- 3.44
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Comparison of Continuous Glucose Monitoring Accuracy Between Abdominal and Upper Arm Insertion Sites. https://app.cochat.ai/papers/10.1089/dia.2019.0014
- The paper
- Steineck, I. I. K., Mahmoudi, Z., Ranjan, A., Schmidt, S., Jørgensen, J. B., & Nørgaard, K. (2019). Comparison of Continuous Glucose Monitoring Accuracy Between Abdominal and Upper Arm Insertion Sites. Diabetes Technology & Therapeutics. https://doi.org/10.1089/dia.2019.0014
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.