Insulin Pump Infusion Set Failures Associated with Prolonged Hyperglycemia: Frequency and Relationship to Age and Type of Infusion Set During 22,741 Infusion Set Wears

Kanapka, Lauren G., Lum, John W., Beck, Roy W.·2022·Diabetes Technology & Therapeutics

Abstract

Background: Although insulin pump infusion set failures are common, studies assessing the failure rate are limited. Methods: Data were analyzed from two clinical trials, in which 263 participants aged 6-72 years used 22,741 infusion sets. The frequency of removal due to prolonged hyperglycemia (continuous glucose monitor measuring >300 mg/dL immediately before removal and >250 mg/dL continuously for at least 2 h before removal with at least 90 min >300 mg/dL out of the prior 120 min) was determined. Differences in failure rates among age groups and infusion set types were evaluated. Results: Among 22,741 infusion sets, 748 (3.3%) were removed before 72 h in association with prolonged hyperglycemia. The percentage replaced within 48 h and within 24 h with prolonged hyperglycemia were 1.8% and 1.0%, respectively. Mean duration of continuous time >250 mg/dL before removal was 5.1 ± 3.7 h. Using a less restrictive definition of failure related to hyperglycemia, 1688 (7.4%) sets were removed before 72 h with a glucose level >300 mg/dL at the time of removal. The frequency of insulin set failure with prolonged hyperglycemia was lower in adults ≥18 years old (1.9%) than in those 14-17 years old (5.8%, P P = 0.002). The 90° Teflon sets had the highest frequency of prolonged hyperglycemia failure within 72 h (4.0%) compared with the angled Teflon set frequency (1.3%, P = 0.01) or the steel set frequency (1.9%, P = 0.006). Conclusions: Based on the data from these 22,741 infusion sets,…

Integrity signals

Registry

Crossref · 10.1089/dia.2021.0548

Cites

3 of 3 references verified0 citation errors · 0 not checkable

Statistics

Recomputing p-values needs the paper’s full text.

Claims the paper makes

This paper’s full text isn’t openly available, so its claims and self-check weren’t run. The registry and citation checks still apply.

Claims other papers make about this one

Each is checked against this paper’s own text: does it say what the citing paper says it does?

  • Kanapka et al. found higher rates of IIS failure in association with 90° Teflon cannulas, but a difference between steel and angled Teflon cannulas was not apparent.

    “In their analysis, Kanapka et al. found higher rates of IIS failure in association with 90° Teflon cannulas, but a difference between steel and angled Teflon cannulas was not apparent.16”
  • Participants aged <18 years or with higher baseline HbA1c had greater odds of IIS failure using a definition of prolonged (≥2 h) hyperglycemia before IIS exchange.

    “In contrast, a recent posthoc analysis by Kanapka et al. examining 22,741 infusion site wears among 263 participants across two clinical trials found that participants aged <18 years or with higher baseline HbA 1c had greater odds of IIS failure using a definition of prolonged ( ‡2 h) hyperglycemia before IIS exchange.16”
  • Insulin infusion set failures are common among individuals using insulin pump therapy.

    “However, insulin infusion set (IIS) failures are common among individuals using insulin pump therapy.3–5”

Cites

Flagged first, then by citations

Cited by

Sorted by citations

Authors
Kanapka, L. G., Lum, J. W., & Beck, R. W.
Affiliations
Jaeb Center for Health Research (US)
Corresponding author
Beck, Roy W.
Venue
Diabetes Technology & Therapeutics
Published
2022-05-31
Volume
24
Issue
6
Pages
396-402
ISSN
  • 1520-9156 (ISSN-L)
  • 1557-8593
OpenAlex
W4210734128
Cited by
35
References
3
Citations by year
2026: 8 · 2024: 7 · 2023: 8 · 2022: 12
FWCI
5.02
Type
journal article
Language
en
Open access
closed
Locations
  • Diabetes Technology & Therapeutics (publishedVersion)
  • PubMed (publishedVersion)
Topics
  • Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Research
  • Physical Sciences › Engineering › Biomedical Engineering › Intravenous Infusion Technology and Safety
  • Health Sciences › Health Professions › Emergency Medical Services › Patient Safety and Medication Errors
Keywords
continuous glucose monitoring, insulin pump therapy, age differences, pediatric diabetes, failure rate, diabetes management
MeSH terms
Adolescent, Adult, Blood Glucose, Child, Diabetes Mellitus, Type 1, Humans, Hyperglycemia, Hypoglycemic Agents, Insulin, Insulin Infusion Systems, Polytetrafluoroethylene, Steel

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review