Randomized Trial of Infusion Set Function: Steel Versus Teflon
Open accessAbstract
Background: This study compared infusion set function for up to 1 week using either a Teflon ® (Dupont ™ , Wilmington, DE) catheter or a steel catheter for insulin pump therapy in type 1 diabetes mellitus. Subjects and Methods: Twenty subjects participating in a randomized, open-labeled, crossover study were asked to wear two Quick-Set ® and two Sure-T ® infusion sets (both from Medtronic Minimed, Northridge, CA) until the infusion set failed or was worn for 1 week. All subjects wore a MiniMed continuous glucose monitoring system for the duration of the study. Results: One subject withdrew from the study. There were 38 weeks of Sure-T wear and 39 weeks of Quick-Set wear with no difference in the survival curves of the infusion sets. There was, however, a 15% initial failure rate with the Teflon infusion set. After 7 days, both types of infusion sets had a 64% failure rate. Overall, 30% failed because of hyperglycemia and a failed correction dose, 13% were removed for pain, 10% were pulled out by accident, 10% had erythema and/or induration of>10 mm, 5% fell out because of loss of adhesion, and 4% were removed for infection. The main predictor of length of wear was the individual subject. There was no increase in hyperglycemia or daily insulin requirements when an infusion set was successfully used for 7 days ( n =25 of 77 weeks). Conclusions: We found no difference between steel and Teflon infusion sets in their function over 7 days, although 15% of Teflon sets failed because of kinking on insertion. The strongest predictor of prolonged 7-day infusion set function was the individual subject, not the type of infusion set.
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The overall EIS survival rate by the end of day 7 was greater than previously reported survival rates for standard IIS devices worn for similar durations (69%, 63%, 43%, 33%, and 32%).
“The overall EIS survival rate of 77.8% (95% CI: 76.2–79.3) by the end of day 7 was greater than previously reported survival rates of 69% (n = 40 total participants and 80 total steel cannula sets),9 63% (n = 40 total participants, 80 total soft cannula sets),9 43% (n = 20 total participants, 40 total sets in non-lipohypertrophied tissue),18 33% (n = 20 total participants and 40 total sets in lipo”
From Evaluation of Extended Infusion Set Performance in Adults with Type 1 Diabetes: Infusion Set Survival Rate and Glycemic Outcomes from a Pivotal Trial (2022) · cites this paper as [8]
The citation is attached to the claimed 69% steel and 63% soft-cannula survival rates. The paper’s Results instead report that, at the end of day 7, 32% of steel sets and 33% of Teflon sets were functioning (page unavailable). (LLM, conf=0.88)
Checked by a reasoning model — may be wrong.
Experimental — accuracy not yet measuredSilent occlusions contribute to the occurrence of unexplained hyperglycemia and glycemic variability in patients using CSII.
“Based on our findings as well as previous reports,15,16,18 we believe that silent occlusions contribute to the occurrence of unexplained hyperglycemia and glycemic variability in patients using CSII.”
From Reduced Silent Occlusions with a Novel Catheter Infusion Set (BD FlowSmart): Results from Two Open-Label Comparative Studies (2016) · cites this paper as [15]
The Discussion (page not available) describes unexplained hyperglycemia in CSII users that resolves after changing the infusion set, and reports unexpected hyperglycemia and/or infusion-set occlusion. It does not identify those occlusions as silent or establish that they contribute to glycemic variability. (LLM, conf=0.84)
Experimental — accuracy not yet measuredA 15% failure rate of polymer catheters over the 8-h period following insertion was observed by Patel et al., and nearly 9% of soft cannula insertions failed within a 6-h period by Renard et al.
“Furthermore, a 15% failure rate of polymer catheters (blood glucose value of >250 mg/dL and not decreasing at least 50 mg in 1 h post–correction bolus) over the 8-h period following insertion was observed by Patel et al.15 and in nearly 9% of soft cannula insertions within a 6-h period by Renard et al,16 suggesting that existing polymer catheters have a tendency to fail upon insertion.”
From Reduced Silent Occlusions with a Novel Catheter Infusion Set (BD FlowSmart): Results from Two Open-Label Comparative Studies (2016) · cites this paper as [15]
The Results report a 15% failure rate on initial insertion of Teflon catheters due to kinking, detected hours later. They do not specify that all 15% failed within 8 hours; the Results describe those failures as occurring on the first day. The Introduction also gives a different primary failure threshold (>300 mg/dL), though the analysis included eight events above 250 mg/dL with an unsuccessful correction bolus. The Renard finding has a separate citation and is not judged here. (LLM, conf=0.88)
Experimental — accuracy not yet measuredPatel et al. found that the strongest predictor of IIS failure was by individual rather than IIS type.
“However, in an open-label crossover study comparing 90° Teflon with steel, Patel et al. found that the strongest predictor of IIS failure was by individual rather than IIS type.3”
From Frequency and Detection of Insulin Infusion Site Failure in the Type 1 Diabetes Exchange Online Community (2023) · cites this paper as [3]
Detection of true causative associations between IIS failure rates and IIS characteristics will require prospective randomization, likely using each individual as their own control.
“Ultimately, detection of true causative associations between IIS failure rates and IIS characteristics will require prospective randomization, likely using each individual as their own control in light of the findings by Patel et al. 3”
From Frequency and Detection of Insulin Infusion Site Failure in the Type 1 Diabetes Exchange Online Community (2023) · cites this paper as [3]
The success of insulin pump therapy may be dependent on the robustness of insertion methods to avoid catheter kinking and failed delivery of insulin into the tissue.
“The success of insulin pump therapy may further be dependent on the robustness of insertion methods to avoid issues such as catheter kinking and therefore failed delivery of insulin into the tissue.”
From Detailed Analysis of Insulin Absorption Variability and the Tissue Response to Continuous Subcutaneous Insulin Infusion Catheter Implantation in Swine (2017) · cites this paper as [19]
Patients commonly experience an insulin bolus that fails to produce the expected glucose lowering effect, leading to persistent hyperglycemia.
“Patients commonly experience an insulin bolus that fails to produce the expected glucose lowering effect, leading to persistent hyperglycemia.6,19”
From Detailed Analysis of Insulin Absorption Variability and the Tissue Response to Continuous Subcutaneous Insulin Infusion Catheter Implantation in Swine (2017) · cites this paper as [19]
Prior rates of IIS failure due to unexplained hyperglycemia ranged from 19% to 30%.
“rates of IIS failure due to “unexplained hyperglycemia” (i.e., hyperglycemia >250 mg/dL and a failed correction bolus due to suspected IIS occlusion) ranged from a relatively high 19%9 to 30%.8”
From Evaluation of Extended Infusion Set Performance in Adults with Type 1 Diabetes: Infusion Set Survival Rate and Glycemic Outcomes from a Pivotal Trial (2022) · cites this paper as [8]
The EIS failure rate due to unexplained hyperglycemia or both unexplained hyperglycemia and device-related SevHyper was substantially lower than the 19%, 23%, 30%, and 35% reported to be attributable or partly attributable to unexplained hyperglycemia in studies investigating extended standard IIS wear.
“The EIS failure rate due to unexplained hyperglycemia or both unexplained hyperglycemia and device-related SevHyper was quite low (≤2%) and substantially lower than the 19%,9 23%,18 30%,8 and 35%18 reported to be attributable or partly attributable to unexplained hyperglycemia in the studies investigating extended standard IIS wear.”
From Evaluation of Extended Infusion Set Performance in Adults with Type 1 Diabetes: Infusion Set Survival Rate and Glycemic Outcomes from a Pivotal Trial (2022) · cites this paper as [8]
There were multiple definitions of a failed correction dose, with 1 h or 60-90 min as the elapsed time by when a single correction dose was required to decrease BG by 50 mg/dL.
“There were also multiple definitions of a failed correction dose where 1 h8,18 or 60–90 min9 was the elapsed time by when a single correction dose was required to decrease BG by 50 mg/dL.”
From Evaluation of Extended Infusion Set Performance in Adults with Type 1 Diabetes: Infusion Set Survival Rate and Glycemic Outcomes from a Pivotal Trial (2022) · cites this paper as [8]
Some of the aforementioned studies reported that standard IIS survival appeared to depend heavily on the individual user, as ∼42% of Patel et al.'s study participants were able to wear an IIS for 6-7 days and 75% of those in Waldenmaier et al.'s study wore at least one standard IIS for 7 days.
“Interestingly, some of the aforementioned studies reported that standard IIS survival appeared to depend heavily on the individual user, as ∼42% of the Patel et al.'s study participants were able to wear an IIS for 6–7 days and 75% of those in Waldenmaier et al.'s study wore at least one standard IIS for 7 days.”
From Evaluation of Extended Infusion Set Performance in Adults with Type 1 Diabetes: Infusion Set Survival Rate and Glycemic Outcomes from a Pivotal Trial (2022) · cites this paper as [8]
The observed infusion set duration is comparable to infusion set durations observed in previous studies.
“The observed infusion set duration is comparable to infusion set durations observed in previous studies. 15,29”
From Real-Time Detection of Infusion Site Failures in a Closed-Loop Artificial Pancreas (2018) · cites this paper as [29]
Informed alarms dictating the appropriate patient-directed intervention potentially increase safety in advanced diabetes technology by improving the interactions between patients and technology.
“Informed alarms dictating the appropriate patient-directed intervention potentially increase safety in advanced diabetes technology by improving the interactions between patients and technology. 29”
From Real-Time Detection of Infusion Site Failures in a Closed-Loop Artificial Pancreas (2018) · cites this paper as [29]
The Introduction (pp. 15–19; passage page unspecified) describes patients checking glucose and changing infusion sets when glucose did not improve. A search for alarms and alerts found no passage addressing whether alarms that specify patient interventions improve safety or patient–technology interactions. (LLM, conf=0.73)
Experimental — accuracy not yet measuredPrior overall findings determined that 7-day wear, irrespective of steel or Teflon cannula, did not adversely influence glycemic control.
“Overall findings from the latter determined that 7-day wear (irrespective of steel or Teflon cannula) did not adversely influence glycemic control”
From Evaluation of Extended Infusion Set Performance in Adults with Type 1 Diabetes: Infusion Set Survival Rate and Glycemic Outcomes from a Pivotal Trial (2022) · cites this paper as [8]
Few randomized studies have reported failure and survival rates when IIS devices are worn out to 7 days.
“few randomized studies have reported failure and survival rates when IIS devices are worn out to 7 days.8,9”
From Evaluation of Extended Infusion Set Performance in Adults with Type 1 Diabetes: Infusion Set Survival Rate and Glycemic Outcomes from a Pivotal Trial (2022) · cites this paper as [8]
While early standard IIS device studies confirmed manufacturer indications for 2- to 3-day use due to increased risk of IIS site reactions or compromised glycemic control, more recent studies with greater numbers of study participants have reported minimal or no issues with glycemia over time or fewer adverse events associated with extended IIS wear.
“Thus, while early standard IIS device studies confirmed manufacturer indications for 2- to 3-day use due to increased risk of IIS site reactions or compromised glycemic control, more recent studies with greater numbers of study participants have reported minimal or no issues with glycemia over time or fewer adverse events associated with extended IIS wear.8,9,18,19”
From Evaluation of Extended Infusion Set Performance in Adults with Type 1 Diabetes: Infusion Set Survival Rate and Glycemic Outcomes from a Pivotal Trial (2022) · cites this paper as [8]
Patel et al. indicated that kinking at initial insertion of Teflon-cannula IISs contributed to 15% of the total IIS failure rate, although investigational staff inserted all IIS devices during the study.
“For instance, Patel et al., indicated that kinking at initial insertion of Teflon-cannula IISs (MiniMed Quick-Set infusion set [Medtronic]) contributed to 15% of the total IIS failure rate, although investigational staff inserted all IIS devices during the study.8”
From Evaluation of Extended Infusion Set Performance in Adults with Type 1 Diabetes: Infusion Set Survival Rate and Glycemic Outcomes from a Pivotal Trial (2022) · cites this paper as [8]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Cites
Pickup, J. C., & Sutton, A. J. (2008) · Diabetic Medicine · doi:10.1111/j.1464-5491.2008.02486.x · Cited by 618 · see citing works
Pickup, J. C., Mattock, M., & Kerry, S. (2002) · BMJ · doi:10.1136/bmj.324.7339.705 · Cited by 481 · see citing works
Gibney, M. A., Arce, C. H., Byron, K. J., & Hirsch, L. J. (2010) · Current Medical Research and Opinion · doi:10.1185/03007995.2010.481203 · Cited by 283 · see citing works
Acute Complications Associated With Insulin Infusion Pump Therapy
Mecklenburg, R. S. (1984) · JAMA · doi:10.1001/jama.252.23.3265 · Cited by 184 · see citing works
Dermatological Complications of Continuous Subcutaneous Insulin Infusion in Children and Adolescents
Conwell, L. S., Pope, E., Artiles, A. M., Mohanta, A., Daneman, A., & Daneman, D. (2008) · The Journal of Pediatrics · doi:10.1016/j.jpeds.2007.10.006 · Cited by 111 · see citing works
Durability of Insulin Pump Use in Pediatric Patients With Type 1 Diabetes
Wood, J. R., Moreland, E. C., Volkening, L. K., Svoren, B. M., Butler, D. A., & Laffel, L. M. B. (2006) · Diabetes Care · doi:10.2337/dc06-1141 · Cited by 98 · see citing works
Schmid, V., Hohberg, C., Borchert, M., Forst, T., & Pfützner, A. (2010) · Journal of Diabetes Science and Technology · doi:10.1177/193229681000400429 · Cited by 97 · see citing works
van Bon, A. C., Bode, B. W., Sert-Langeron, C., DeVries, J. H., & Charpentier, G. (2011) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2010.0224 · Cited by 96 · see citing works
Presti, D. L., Ingegnosi, C., & Strauss, K. W. (2012) · Pediatric Diabetes · doi:10.1111/j.1399-5448.2012.00865.x · Cited by 83 · see citing works
Long-Term Metabolic Control with Insulin Pump Therapy
Mecklenburg, R. S., Benson, E. A., Benson, J. W., Blumenstein, B. A., Fredlund, P. N., Guinn, T. S., Metz, R. J., & Nielsen, R. L. (1985) · New England Journal of Medicine · doi:10.1056/nejm198508223130802 · Cited by 68 · see citing works
Thethi, T. K., Rao, A., Kawji, H., Mallik, T., Yau, C. L., Christians, U., & Fonseca, V. (2010) · Journal of Diabetes and its Complications · doi:10.1016/j.jdiacomp.2009.03.002 · Cited by 53 · see citing works
Schober, E., & Rami-Merhar, B. M. (2008) · Pediatric Diabetes · doi:10.1111/j.1399-5448.2008.00477.x · Cited by 51 · see citing works
Renard, E., Guerci, B., Leguerrier, A.-M., & Boizel, R. (2010) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2010.0073 · Cited by 42 · see citing works
Metabolic decompensation in pump users due to lispro insulin precipitation
Wolpert, H. A., Faradji, R. N., Bonner-Weir, S., & Lipes, M. A. (2002) · BMJ · doi:10.1136/bmj.324.7348.1253 · Cited by 42 · see citing works
Højbjerre, L., Skov-Jensen, C., Kaastrup, P., Pedersen, P. E., & Stallknecht, B. (2009) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2008.0061 · Cited by 24 · see citing works
Patient management of long‐term continuous subcutaneous insulin infusion
Johansson, U., Adamson, U., LINS, P.-E. S., & Wredling, R. (2005) · Journal of Advanced Nursing · doi:10.1111/j.1365-2648.2005.03475.x · Cited by 20 · see citing works
Toxic-shock syndrome in a patient using a continuous subcutaneous insulin infusion pump--Idaho.
(1983) · PubMed · Cited by 9 · see citing works
Flagged first, then by citations
Cited by
Sorted by citations
New insulin delivery recommendations
Frid, A. H., Kreugel, G., Grassi, G., Halimi, S., Hicks, D., Hirsch, L. J., Smith, M. J., Wellhoener, R., Bode, B. W., Hirsch, I. B., et al. (2016) · Mayo Clinic Proceedings · doi:10.1016/j.mayocp.2016.06.010 · Cited by 286 · see citing works
Advances in technology for management of type 1 diabetes
Beck, R. W., Bergenstal, R. M., Laffel, L. M., & Pickup, J. C. (2019) · The Lancet · doi:10.1016/s0140-6736(19)31142-0 · Cited by 242 · see citing works
Latres, E., Finan, D. A., Greenstein, J. L., Kowalski, A. J., & Kieffer, T. J. (2019) · Cell Metabolism · doi:10.1016/j.cmet.2019.02.007 · Cited by 118 · see citing works
Domingo‐Lopez, D. A., Lattanzi, G., Schreiber, L. H. J., Wallace, E. J., Wylie, R., O’Sullivan, J., Dolan, E. B., & Duffy, G. P. (2022) · Advanced Drug Delivery Reviews · doi:10.1016/j.addr.2022.114280 · Cited by 103 · see citing works
Clinical review: insulin pump-associated adverse events in adults and children
Ross, P., Milburn, J., Reith, D. M., Wiltshire, E. J., & Wheeler, B. J. (2015) · Acta Diabetologica · doi:10.1007/s00592-015-0784-2 · Cited by 95 · see citing works
Nimri, R., Nir, J., & Phillip, M. (2019) · American Journal of Therapeutics · doi:10.1097/mjt.0000000000001097 · Cited by 73 · see citing works
Insulin Infusion Set Use: European Perspectives and Recommendations
Deiss, D., Adolfsson, P., Alkemade-van Zomeren, M., Bolli, G. B., Charpentier, G., Cobelli, C., Danne, T., Girelli, A., Mueller, H., Verderese, C. A., et al. (2016) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2016.07281.sf · Cited by 62 · see citing works
Bionic Pancreas Research Group, Messer, L. H., Buckingham, B. A., Cogen, F., Daniels, M., Forlenza, G., Jafri, R. Z., Mauras, N., Muir, A., Wadwa, R. P., et al. (2022) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2022.0201.pub · Cited by 59 · see citing works
Self-citationForlenza, G. P., Deshpande, S., Ly, T. T., Howsmon, D. P., Cameron, F. M., Baysal, N., Mauritzen, E., Marcal, T., Towers, L., Bequette, B. W., et al. (2017) · Diabetes Care · doi:10.2337/dc17-0500 · Cited by 58 · see citing works
Self-citationPfützner, A., Sachsenheimer, D., Grenningloh, M., Heschel, M., Walther-Johannesen, L., Gharabli, R., & Klonoff, D. (2015) · Journal of Diabetes Science and Technology · doi:10.1177/1932296815604438 · Cited by 53 · see citing works
Fault Detection and Safety in Closed-Loop Artificial Pancreas Systems
Bequette, B. W. (2014) · Journal of Diabetes Science and Technology · doi:10.1177/1932296814543661 · Cited by 50 · see citing works
Tarafdar, A., Xie, J., Gowen, A., O'Higgins, A. C., & Xu, J. (2024) · The Science of The Total Environment · doi:10.1016/j.scitotenv.2024.172648 · Cited by 51 · see citing works
Weber, C., Kammerer, D., Streit, B., & Licht, A. H. (2014) · Toxicology Reports · doi:10.1016/j.toxrep.2014.11.019 · Cited by 48 · see citing works
A Review of Safety and Hazards Associated With the Artificial Pancreas
Ramkissoon, C. M., Aufderheide, B., Bequette, B. W., & Vehı́, J. (2017) · IEEE Reviews in Biomedical Engineering · doi:10.1109/rbme.2017.2749038 · Cited by 41 · see citing works
Bionic Pancreas Research Group, Beck, R. W., Russell, S. J., Damiano, E. R., El-Khatib, F. H., Ruedy, K. J., Balliro, C., Li, Z., & Calhoun, P. (2022) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2022.0167 · Cited by 40 · see citing works
Hauzenberger, J. R., Münzker, J., Kotzbeck, P., Asslaber, M., Bubalo, V., Joseph, J. I., & Pieber, T. R. (2018) · Scientific Reports · doi:10.1038/s41598-017-18790-0 · Cited by 40 · see citing works
Hauzenberger, J. R., Hipszer, B. R., Loeum, C., McCue, P. A., DeStefano, M., Torjman, M. C., Kaner, M. T., Dinesen, A. R., Chervoneva, I., Pieber, T. R., et al. (2017) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2017.0175 · Cited by 39 · see citing works
“The success of insulin pump therapy may further be dependent on the robustness of insertion methods to avoid issues such as catheter kinking and therefore failed delivery of insulin into the tissue.(18,19) We hypothesize that the flow of insulin from a CSII catheter into adjacent vascular tissue is affected by the cha…”
BackgroundGlucose-responsive insulin delivery for type 1 diabetes: The artificial pancreas story
Bally, L., Thabit, H., & Hovorka, R. (2017) · International Journal of Pharmaceutics · doi:10.1016/j.ijpharm.2017.12.022 · Cited by 40 · see citing works
Inducible Pluripotent Stem Cells as a Potential Cure for Diabetes
Verhoeff, K., Henschke, S., Marfil-Garza, B., Dadheech, N., & Shapiro, A. M. (2021) · Cells · doi:10.3390/cells10020278 · Cited by 21 · see citing works
Gibney, M., Xue, Z., Swinney, M., Bialonczyk, D., & Hirsch, L. (2016) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2015.0342 · Cited by 32 · see citing works
Karlin, A. W., Ly, T. T., Pyle, L., Forlenza, G. P., Messer, L., Wadwa, R. P., DeSalvo, D. J., Payne, S. L., Hanes, S., Clinton, P., et al. (2016) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2015.0432 · Cited by 32 · see citing works
Self-citationReal-Time Detection of Infusion Site Failures in a Closed-Loop Artificial Pancreas
Howsmon, D. P., Baysal, N., Buckingham, B. A., Forlenza, G. P., Ly, T. T., Maahs, D. M., Marcal, T., Towers, L., Mauritzen, E., Deshpande, S., et al. (2018) · Journal of Diabetes Science and Technology · doi:10.1177/1932296818755173 · Cited by 32 · see citing works
Self-citationKanapka, L. G., Lum, J. W., & Beck, R. W. (2022) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2021.0548 · Cited by 35 · see citing works
Improving Patient Experience With Insulin Infusion Sets
Evert, A. B., Bode, B. W., Buckingham, B. A., Nardacci, E., Verderese, C. A., Wolff-McDonagh, P., Walsh, J. S. A., & Hirsch, I. B. (2016) · The Science of Diabetes Self-Management and Care · doi:10.1177/0145721716642526 · Cited by 29 · see citing works
Self-citationDouglass, M. E., Hopkins, S. P., Chug, M. K., Kim, G., Garren, M. R. S., Ashcraft, M. M., Nguyen, D. T., Tayag, N., Handa, H., & Brisbois, E. J. (2021) · ACS Applied Materials & Interfaces · doi:10.1021/acsami.1c18190 · Cited by 29 · see citing works
Transcutaneous Flexible Sensor forIn VivoPhotonic Detection of pH and Lactate
Nguyen, D., Lawrence, M. M., Berg, H., Lyons, M. A., Shreim, S., Keating, M. T., Weidling, J., & Botvinick, E. L. (2022) · ACS Sensors · doi:10.1021/acssensors.1c01720 · Cited by 29 · see citing works
Insulin Infusion Sets: A Critical Reappraisal
Heinemann, L. (2016) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2016.0013 · Cited by 29 · see citing works
Brazg, R., Garg, S. K., Bhargava, A., Thrasher, J. R., Latif, K., Bode, B. W., Bailey, T. S., Horowitz, B. S., Cavale, A., Kudva, Y. C., et al. (2022) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2021.0540 · Cited by 28 · see citing works
Self-citationInsulin Pump and CGM Usage in the United States and Germany
Walsh, J. S. A., Roberts, R., Weber, D., Faber-Heinemann, G., & Heinemann, L. (2015) · Journal of Diabetes Science and Technology · doi:10.1177/1932296815588945 · Cited by 28 · see citing works
Regittnig, W., Tschaikner, M., Tuca, A., Simic, A., Feiel, J., Schaller‐Ammann, R., Licht, A. H., Jungklaus, M., & Pieber, T. R. (2022) · Diabetes, Obesity and Metabolism · doi:10.1111/dom.14594 · Cited by 27 · see citing works
Continuous Glucose Monitoring Enables the Detection of Losses in Infusion Set Actuation (LISAs)
Howsmon, D., Cameron, F., Baysal, N., Ly, T., Forlenza, G., Maahs, D., Buckingham, B., Hahn, J., & Bequette, B. (2017) · Sensors · doi:10.3390/s17010161 · Cited by 26 · see citing works
Self-citationRumpler, M., Mader, J. K., Fischer, J., Thar, R., Granger, J., Deliane, F., Klimant, I., Aberer, F., Sinner, F. M., Pieber, T. R., et al. (2016) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2016.08.039 · Cited by 25 · see citing works
Advancing continuous subcutaneous insulin infusion in vivo: New insights into tissue challenges
Kesserwan, S., Mulka, A., Sharafieh, R., Qiao, Y., Wu, R., Kreutzer, D. L., & Klueh, U. (2020) · Journal of Biomedical Materials Research Part A · doi:10.1002/jbm.a.37097 · Cited by 25 · see citing works
Taleb, N., Messier, V., Ott-Braschi, S., Ardilouze, J.-L., & Rabasa-Lhoret, R. (2018) · Diabetes Research and Clinical Practice · doi:10.1016/j.diabres.2018.07.035 · Cited by 20 · see citing works
Tissue Response to Subcutaneous Infusion Catheter
Zhang, E., & Cao, Z. (2019) · Journal of Diabetes Science and Technology · doi:10.1177/1932296819837972 · Cited by 24 · see citing works
Waldenmaier, D., Zschornack, E., Buhr, A., Pleus, S., Haug, C., & Freckmann, G. (2020) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2019.0445 · Cited by 21 · see citing works
Advances in Insulin Pump Infusion Sets Symposium Report
Zhang, J. Y., Shang, T., CHATTARAJ, S., Cohen, O., Heschel, M., Vigersky, R. A., Heinemann, L., Nørgaard, K., Svensson, J., Buckingham, B. A., et al. (2021) · Journal of Diabetes Science and Technology · doi:10.1177/1932296821999080 · Cited by 23 · see citing works
Self-citationA Glucose‐Responsive Cannula for Automated and Electronics‐Free Insulin Delivery
Fuchs, S., Caserto, J. S., Liu, Q., Wang, K., Shariati, K., Hartquist, C. M., Zhao, X., & Ma, M. (2024) · Advanced Materials · doi:10.1002/adma.202403594 · Cited by 22 · see citing works
Longevity of the novel <scp>ConvaTec</scp> infusion set with Lantern technology
Lal, R. A., Hsu, L., Zhang, J., Schøndorff, P. K., Heschel, M., & Buckingham, B. (2021) · Diabetes, Obesity and Metabolism · doi:10.1111/dom.14395 · Cited by 22 · see citing works
Self-citationMulka, A., Lewis, B. E., Mao, L., Sharafieh, R., Kesserwan, S., Wu, R., Kreutzer, D. L., & Klueh, U. (2021) · ACS Pharmacology & Translational Science · doi:10.1021/acsptsci.1c00047 · Cited by 22 · see citing works
We Need More Research and Better Designs for Insulin Infusion Sets
Heinemann, L., Walsh, J. S. A., & Roberts, R. (2014) · Journal of Diabetes Science and Technology · doi:10.1177/1932296814523882 · Cited by 22 · see citing works
EISLER, G. J., Kastner, J. R., Torjman, M. C., Khalf, A., Diaz, D., Dinesen, A. R., Loeum, C., Thakur, M. L., Strasma, P. J., & Joseph, J. I. (2019) · BMJ Open Diabetes Research & Care · doi:10.1136/bmjdrc-2019-000881 · Cited by 18 · see citing works
Bonato, L., Taleb, N., Gingras, V., Messier, V., Gobeil, F., Ménard, J., Ardilouze, J., & RABASA-LHORET, R. P. R. (2018) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2018.0110 · Cited by 17 · see citing works
Tschaikner, M., Powell, K., Jungklaus, M., Fritz, M., Ellmerer, M., Hovorka, R., Lane, S., Pieber, T. R., & Regittnig, W. (2019) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2019.2925434 · Cited by 18 · see citing works
Hughes, M. S., Douvas, J. L., Layfield-Bryan, M., Blanco, L. E., Gray, J. C., Zapotoczny, G., Espinoza, J., Wilcox, J. H., & Lal, R. A. (2023) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2023.0005 · Cited by 16 · see citing works
Development of the Extended Infusion Set and Its Mechanism of Action
ZHANG, G., Cohen, O., & CHATTARAJ, S. (2022) · Journal of Diabetes Science and Technology · doi:10.1177/19322968221112120 · Cited by 16 · see citing works
An Artificial Pancreas for Automated Blood Glucose Control in Patients With Type 1 Diabetes
Schmidt, S., Boiroux, D., Ranjan, A. G., J�rgensen, J., Madsen, H., & Nørgaard, K. (2015) · Therapeutic Delivery · doi:10.4155/tde.15.12 · Cited by 16 · see citing works
Randomized Cross-Over Study Comparing Two Infusion Sets for CSII in Daily Life
Freckmann, G., Arndt, S., Fießelmann, A., Klausmann, G., Pralle, K., Künsting, T., & Petersen, B. (2017) · Journal of Diabetes Science and Technology · doi:10.1177/1932296816667510 · Cited by 14 · see citing works
New advances in insulin products
Rajpoot, K., Tekade, M., Sharma, M. C., Molugulu, N., & Tekade, R. K. (2020) · Elsevier eBooks · doi:10.1016/b978-0-12-814455-8.00014-1 · Cited by 15 · see citing works
Rabbone, I., Minuto, N., Toni, S., Lombardo, F., Iafusco, D., Marigliano, M., Schiaffini, R., Maltoni, G., Frongia, A. P., Scardapane, M., et al. (2018) · Diabetes Obesity and Metabolism · doi:10.1111/dom.13419 · Cited by 15 · see citing works
Feasibility study of a prototype extended‐wear insulin infusion set in adults with type 1 diabetes
Kastner, J. R., Venkatesh, N., Brown, K., Muchmore, D. B., Ekinci, E. I., Fourlanos, S., Joseph, J. I., Shafeeq, M., Shi, L., Strange, P., et al. (2022) · Diabetes Obesity and Metabolism · doi:10.1111/dom.14685 · Cited by 14 · see citing works
McVey, E., Keith, S., Herr, J. K., Sutter, D. E., & Pettis, R. J. (2015) · Journal of Diabetes Science and Technology · doi:10.1177/1932296815598327 · Cited by 14 · see citing works
Graf, A., McAuley, S. A., Sims, C. M., Ulloa, J., Jenkins, A. J., Voskanyan, G. R., & O'Neal, D. N. (2016) · Journal of Diabetes Science and Technology · doi:10.1177/1932296816682762 · Cited by 14 · see citing works
Swinney, M. R., Cox, A. L., Hawkins, E. D., Xue, J., Garhyan, P., Stanley, J. R. L., Sule, S. V., Adragni, K. P., & Michael, M. D. (2020) · Journal of Pharmaceutical Sciences · doi:10.1016/j.xphs.2020.12.008 · Cited by 13 · see citing works
Two case reports of retained steel insulin pump infusion set needles
Plager, P., Murati, M. A., Moran, A., & Sunni, M. (2015) · Pediatric Diabetes · doi:10.1111/pedi.12266 · Cited by 13 · see citing works
Irwin, T. D., Speirs, A., & Merrett, C. G. (2021) · Journal of the mechanical behavior of biomedical materials/Journal of mechanical behavior of biomedical materials · doi:10.1016/j.jmbbm.2021.104660 · Cited by 12 · see citing works
Kastner, J. R., EISLER, G. J., Torjman, M. C., Khalf, A., Diaz, D., Dinesen, A. R., Loeum, C., Strasma, P. J., & Joseph, J. I. (2022) · Journal of Diabetes Science and Technology · doi:10.1177/19322968221093362 · Cited by 11 · see citing works
Kwa, T., Zhang, G., Shepard, K., Wherry, K., & Chattaraj, S. (2021) · Journal of Medical Economics · doi:10.1080/13696998.2021.1945784 · Cited by 10 · see citing works
Inflammation at Site of Insulin Infusion Diminishes Glycemic Control
Kesserwan, S., Lewis, B. E., Mao, L., Sharafieh, R., Atwood, T., Kreutzer, D. L., & Klueh, U. (2022) · Journal of Pharmaceutical Sciences · doi:10.1016/j.xphs.2021.12.025 · Cited by 10 · see citing works
Case Report: Insulin hypersensitivity in youth with type 1 diabetes
Alkhatib, E. H., Grundman, J., Adamusiak, A., Bellin, M. D., Brooks, J. P., Buckley, K. S., Janssen, E., Kitcharoensakkul, M., McNerney, K. P., Pfeifer, T. L., et al. (2023) · Frontiers in Endocrinology · doi:10.3389/fendo.2023.1226231 · Cited by 9 · see citing works
Improving the Patient Experience With Longer Wear Infusion Sets Symposium Report
Nguyen, K. T., Xu, N. Y., Buckingham, B. A., CHATTARAJ, S., Cohen, O., Heinemann, L., Pickup, J. C., Svensson, J., Vigersky, R. A., Wong, J. C., et al. (2022) · Journal of Diabetes Science and Technology · doi:10.1177/19322968221078884 · Cited by 8 · see citing works
Self-citationAdvances in Insulin Infusion Set in the New Era of Automated Insulin Delivery: A Systematic Review
ZHANG, G., ANSELMO, E., Kwa, T., Cohen, O., Vigersky, R. A., & CHATTARAJ, S. (2022) · Journal of Diabetes Science and Technology · doi:10.1177/19322968221145731 · Cited by 7 · see citing works
The DERMIS Study: Methodologies, Results, and Implications for the Future
Hirsch, I. B., Khakpour, D., Joseph, J. I., Shinohara, M. M., Wang, R. K., Klueh, U., Kreutzner, D., Riveline, J., Jacquemier, P., Maier, L. E., et al. (2024) · Journal of Diabetes Science and Technology · doi:10.1177/19322968241298005 · Cited by 6 · see citing works
Insulin infusion sets associated with adverse events: strategies for improved diabetes education
Neves, A. L. D., Martins, L. E. G., Gabbay, M. A. L., Pascali, P. M., de Oliveira, T., Martinazzo, A., Dib, S. A., Casarini, D. E., Arêdes, S. V., Tenório, F. S., et al. (2023) · Frontiers in Medicine · doi:10.3389/fmed.2023.1275394 · Cited by 6 · see citing works
Hendel, K., Stumpe, T., & Özer, K. (2022) · Journal of Diabetes Science and Technology · doi:10.1177/19322968221138076 · Cited by 5 · see citing works
Mast Cell Deficiency in Mice Attenuates Insulin Phenolic Preservative-Induced Inflammation
Kesserwan, S., Sadagurski, M., Mao, L., & Klueh, U. (2023) · Biomedicines · doi:10.3390/biomedicines11082258 · Cited by 5 · see citing works
Modelling and simulation of occlusions in insulin pumps
Formo, M. W., Stavdahl, O., & Fougner, A. L. (2021) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc46164.2021.9630219 · Cited by 5 · see citing works
Yoshimura, K., Urai, S., Hozumi, K., Nakatsuji, M., Saito, S., Nishikage, S., Yamamoto, A., Takayoshi, T., Ogawa, W., & Hirota, Y. (2025) · Journal of Diabetes Science and Technology · doi:10.1177/19322968251345837 · Cited by 3 · see citing works
Kowalczyk, E., & Szypowska, A. (2024) · International Journal of Diabetes in Developing Countries · doi:10.1007/s13410-024-01310-y · Cited by 3 · see citing works
Messer, L. H., Buckingham, B. A., Cogen, F. R., Daniels, M., Forlenza, G. P., Jafri, R. Z., Mauras, N., MUIR, A. B., Wadwa, R. P., White, P. C., et al. (2022) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2022.0201 · Cited by 3 · see citing works
Self-citationRetractedBest Ways and Practices to Avoid Insulin Pump Catheter Occlusions
Slover, R. H. (2016) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2016.0025 · Cited by 2 · see citing works
Insulin Delivery Hardware: Pumps and Pens
Lal, R., & Leelarathna, L. (2023) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2023.2503 · Cited by 2 · see citing works
Renard, É., & Boizel, R. (2014) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2014.0058 · Cited by 2 · see citing works
Pachołek, A., Oniszczuk, E. R. S. , Mierzwa, J., Wróbel, A., Piotrowicz, K., Gąsowski, J., & Tomasik, T. (2023) · Folia Medica Cracoviensia · doi:10.24425/fmc.2023.145426 · Cited by 2 · see citing works
Zhang, Y., Kuo, L., Woodhouse, K. A., & Fitzpatrick, L. E. (2024) · ACS Pharmacology & Translational Science · doi:10.1021/acsptsci.4c00363 · Cited by 1 · see citing works
Analysis of “Randomized Cross-Over Study Comparing Two Infusion Sets for CSII in Daily Life”
Renard, É. (2016) · Journal of Diabetes Science and Technology · doi:10.1177/1932296816675177 · Cited by 1 · see citing works
Insulin Delivery Hardware: Pumps and Pens
Lal, R., & Leelarathna, L. (2024) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2024.2503 · Cited by 1 · see citing works
Device‐Enabled Drug Infusion Therapies
Hildebrand, K. R. (2015) · doi:10.1002/9781119002956.ch07 · Cited by 1 · see citing works
Clinical utility of the twiist Automated Insulin Delivery System: first study with twiist
Ruiz, A., Mcdonough, A., Snow, K., Bode, B., & Mitri, J. (2026) · BMJ Open Diabetes Research & Care · doi:10.1136/bmjdrc-2026-006307
Skin Oil Staining to Avoid Infusion Set Cannula Crimping
Bossi, A. C. (2016) · Journal of Diabetes Science and Technology · doi:10.1177/1932296816634358
Forlenza, G. P. (2017) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2017.0048
Insulin Delivery Hardware: Pumps and Pens
Lal, R., & Leelarathna, L. (2021) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2021.2503
Kastner, J. R., Poettler, T., Kopanz, J., Hochfellner, D. A., Romey, M., Baumann, P. M., Muchmore, D. B., Strasma, P. J., & Mader, J. K. (2022) · Diabetes Obesity and Metabolism · doi:10.1111/dom.14935 · Cited by 1 · see citing works
Ucieklak, D., Mrozińska, S., Wojnarska, A., Malecki, M. T., Klupa, T., & Matejko, B. (2023) · Folia Medica Cracoviensia · doi:10.24425/fmc.2022.144079
Bi, Q., Lai, M., Chen, K., Liu, J. M., Tang, H., Teng, X., & Guo, Y. (2023) · Advances in Production Engineering & Management · doi:10.14743/apem2023.1.456
New Technologies for Insulin Administration
Forde, H. E., & Choudhary, P. (2024) · Wiley eBooks · doi:10.1002/9781119697473.ch33
Barnett, M. J. L., Prater, J., Green, L., & Mayrin, J. V. (2024) · Cureus · doi:10.7759/cureus.73900
Ansong, M., & Friedman, S. H. (2025) · Molecular Pharmaceutics · doi:10.1021/acs.molpharmaceut.5c01159
Groisman, L., Thorson, T. J., Saudi, B., Myrick, R. J., Gonthier, A. R., Nelson, G., Loeum, C., Torjman, M. C., Joseph, J. I., Mohraz, A., et al. (2026) · Biomaterials Advances · doi:10.1016/j.bioadv.2026.215155
- Authors
- Patel, P. J., Benasi, K., Ferrari, G., Evans, M. G., Shanmugham, S., Wilson, D. M., & Buckingham, B. A.
- Affiliations
- Stanford Medicine (US)
- Stanford Medical Center
- Venue
- Diabetes Technology & Therapeutics
- Publisher
- SAGE Publications
- Published
- 2014-01
- Volume
- 16
- Issue
- 1
- Pages
- 15-19
- ISSN
- 1520-9156 (ISSN-L)
- 1557-8593
- PMC
- PMC3887416
- OpenAlex
- W1972348237
- Cited by
- 90
- References
- 17
- Citations by year
- 2026: 2 · 2025: 2 · 2024: 8 · 2023: 8 · 2022: 15 · 2021: 10 · 2020: 5 · 2019: 6 · 2018: 5 · 2017: 6 · 2016: 11 · 2015: 8 · 2014: 4
- FWCI
- 6.04
- Type
- journal article
- Language
- en
- Open access
- green
- www.ncbi.nlm.nih.gov
- Locations
- Diabetes Technology & Therapeutics (publishedVersion)
- Europe PMC (PubMed Central) (publishedVersion)
- PubMed Central (publishedVersion)
- PubMed (publishedVersion)
- Funders
- Genentech
- Topics
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Research
- Physical Sciences › Engineering › Biomedical Engineering › Intravenous Infusion Technology and Safety
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Education
- Keywords
- Teflon catheter, insulin pump therapy, type 1 diabetes, continuous glucose monitoring, hyperglycemia, insulin requirement
- MeSH terms
- Adolescent, Adult, Blood Glucose, Catheterization, Child, Diabetes Mellitus, Type 1, Equipment Design, Female, Glycated Hemoglobin, Humans, Hyperglycemia, Hypoglycemic Agents, Infusions, Intravenous, Insulin, Male, Monitoring, Physiologic, Polytetrafluoroethylene, Steel, Time Factors, Blood Glucose Self-Monitoring, Body Mass Index, Cross-Over Studies
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review