Interlaboratory standardization of glycated hemoglobin determinations.

Little, R R, England, J D, Wiedmeyer, H M, McKenzie, E M, Mitra, R, Erhart, P M, Durham, J B, Goldstein, D E · 1986 · Clinical Chemistry
Cited by 63

The DOI matches its Crossref record.

Abstract

As the clinical utility of glycated hemoglobin (gHb) measurement increases, so does the need for standardization of values between different methods and different laboratories. Using three different methods, we examined the feasibility of interlaboratory standardization of gHb measurement. A liquid-chromatographic (HPLC) system from our research laboratory was designated the reference method. For gHb standards we used erythrocyte hemolysates prepared from blood samples from nondiabetic and diabetic subjects. Values assigned to each standard were based on the mean of multiple gHb determinations by the HPLC method. A clinical laboratory routinely prepared hemolysates and assayed gHb by commercially available ion-exchange ("mini column") and affinity chromatographic methods. For each assay a standard curve was constructed and gHb values were derived from these curves. Samples analyzed in the clinical laboratory were also analyzed in the research laboratory and the curve-derived values were compared with the HPLC-measured values, to determine the accuracy of our interlaboratory standardization procedure. Correlations were excellent (r = 0.99). The lack of significant differences between calculated and HPLC-measured values indicates that interlaboratory standardization is feasible.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1093/clinchem/32.2.358

Cites

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cited by

See all 63

Metadata

Authors
Little, R. R., England, J. D., Wiedmeyer, H. M., McKenzie, E. M., Mitra, R., Erhart, P. M., Durham, J. B., & Goldstein, D. E.
Venue
Clinical Chemistry
Published
1986-02-01
Volume
32
Issue
2
Pages
358-360
ISSN
  • 0009-9147 (ISSN-L)
  • 1530-8561
OpenAlex
W4299857163
Cited by
63
References
0
FWCI
2.42
Type
journal article
Language
en
Open access
closed

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Interlaboratory standardization of glycated hemoglobin determinations. https://app.cochat.ai/papers/10.1093/clinchem/32.2.358
The paper
Little, R. R., England, J. D., Wiedmeyer, H. M., McKenzie, E. M., Mitra, R., Erhart, P. M., Durham, J. B., & Goldstein, D. E. (1986). Interlaboratory standardization of glycated hemoglobin determinations. Clinical Chemistry. https://doi.org/10.1093/clinchem/32.2.358

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.