CoChat

Evaluation of a Homogeneous Direct LDL-Cholesterol Assay in Diabetic Patients: Effect of Glycemic Control

Open access
Ragland, Brian D, Konrad, Robert J, Chaffin, Carolyn, Robinson, C Andrew, Hardy, Robert W · 2000 · Clinical Chemistry
Cited by 28

The DOI matches its Crossref record; 1 of the 1 references it cites were found in a registry or an index.

Abstract

Increased serum LDL-cholesterol (LDL-C) is associated with increased risk for ischemic heart disease, and lowering of LDL-C has been shown to decrease mortality in patients with known coronary heart disease (1)(2)(3). Persons with diabetes have a greatly increased risk for atherosclerosis and its complications. Many of these patients have increased plasma cholesterol (including LDL-C) and triglycerides (TGs). Additionally, glycation and oxidation of circulating LDL particles may further increase the risk for atherosclerotic disease in patients with diabetes (4)(5). Currently used methods for LDL-C include calculations [based on total cholesterol (TC), HDL-cholesterol (HDL-C), and TG concentrations], ultracentrifugation, and most recently, direct LDL-C assays. Calculation methods, such at the Friedewald formula (6), are well known to have a prominent negative bias in patients with TG concentrations >4.5 mmol/L (400 mg/dL) (6)(7). In addition, because TGs are measured, fasting blood samples are preferred. This can present a problem for certain patient populations, including many people with diabetes. Ultracentrifugation methods are time-consuming and expensive and generally are performed only in reference laboratories. Direct LDL-C assays have been developed recently and have been shown to provide accurate and precise measurements of LDL-C (8)(9)(10)(11)(12). They overcome the TG and fasting limitations of calculation methods, are readily adapted to clinical laboratories, are less expensive than ultracentrifugation methods, and provide greatly improved turnaround time. Direct LDL-C assays make use of either an immunoseparation step or specific detergents to separate LDL particles from other lipoproteins, followed by measurement of cholesterol by conventional enzymatic reactions. The N-geneousTM LDL-C assay is a solute-based homogeneous assay that agrees well with ultracentrifugation (11). The purpose of the present study was to evaluate the effect of glycemic control (as gauged by hemoglobin A1c) on the accuracy of the N-geneous assay. Additionally, we compared this direct LDL-C assay to density gradient ultracentrifugation (13) and to the Friedewald calculation (6), and assessed its usefulness in determining a calculated VLDL-cholesterol (VLDL-C) concentration. Overnight fasting plasma samples were obtained from 52 patients. Direct LDL-C was measured using the N-geneous LDL-C assay (Genzyme Diagnostics) performed on the Beckman SYNCHRON LX 20 (Beckman-Coulter). Briefly, this assay involves a two-reagent process. The first reagent is a detergent that solubilizes only non-LDL lipoprotein particles. The solubilized cholesterol is consumed in a non-color-producing reaction by cholesterol esterase and cholesterol oxidase. The second reagent is then added, which solubilizes the remaining LDL particles and links the enzymatic consumption of LDL-C to a chromogenic coupler. The colored product is measured spectrophotometrically. TC and TG concentrations were measured using enzymatic methods on the SYNCHRON LX System. HDL-C was measured with Beckman-Coulter reagents on the SYNCHRON LX System, which uses a detergent that solubilizes only HDL particles, allowing HDL-C to be measured by a color-producing enzymatic reaction. An estimated LDL-C concentration was obtained using the Friedewald calculation (6): TC − (HDL-C + TG/5). VLDL-C concentrations were estimated using TG/5, and were also calculated using the direct LDL-C measurement in the formula: VLDL-C = TC − (HDL-C + LDL-C). LDL-C and VLDL-C were also measured by the Vertical Auto Profile (VAP) technique (Atherotech) (13). Briefly, VAP uses density gradient ultracentrifugation to separate lipoproteins, with subsequent measurement of cholesterol in each of the resulting density bands. Hemoglobin A1c was measured by HPLC (TOSOH A1C2.2) on diabetic patients (41 of the 52 samples) (14)(15). On all 52 plasma samples, correlation studies were performed comparing LDL-C measurements by the N-geneous LDL-C assay with those from the ultracentrifugation (VAP) technique (13) and Friedewald calculation (6). Linearity of the direct LDL assay was assessed with a patient plasma sample determined to have a LDL-C concentration of 2400 g/L. The following dilutions, in normal saline, were measured: 1:1, 1:2, 1:4, 1:8, 1:16, 1:32, 1:64, and 1:128. Within-run precision studies used a serum-based control (Liquichek Lipid Control level 1; Bio-Rad Laboratories) and the N-geneous LDL-C lyophilized serum calibrator (Genzyme). Between-run precision studies used Bio-Rad Liquichek Lipid Control level 1 and level 2. Within-day precision was assessed on 20 consecutive assays, whereas between-day precision was determined from data obtained in triplicate over 20 separate days. Interference studies were performed on a patient sample with a TG concentration of 28 mmol/L (2490 mg/dL). Direct LDL-C measurements were obtained from the undiluted sample as well as from 1:2, 1:4, 1:8, and 1:16 dilutions of

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1093/clinchem/46.11.1848

Cites

1 of 1 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Cited by

See all 28

Metadata

Authors
Ragland, B. D., Konrad, R. J., Chaffin, C., Robinson, C. A., & Hardy, R. W.
Venue
Clinical Chemistry
Published
2000-11-01
Volume
46
Issue
11
Pages
1848-1851
ISSN
  • 0009-9147 (ISSN-L)
  • 1530-8561
OpenAlex
W2116446603
Cited by
28
References
1
FWCI
1.71
Type
journal article
Language
en
Open access
bronze

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.