Unexpected Suppression of Immunoassay Results by Cross-Reactivity: Now a Demonstrated Cause for Concern
The DOI matches its Crossref record.
Abstract
More than other analytical procedures, immunoassays have afforded a wealth of knowledge about biochemical physiology over the last 40 years. These assays have posed major analytical challenges, which most likely stem from the delicate balance between chemical equilibrium and our ever-increasing quest for speed of analysis, simplicity of procedure, and improved reliability. In attempts to enhance these criteria, we often face limitations imposed by the fundamental principles of thermodynamics: balancing chemical equilibrium, the underlying kinetics, and that ever-present variable, time. Cross-reactivity in competitive immunoassays can be measured by various methods, including 50% displacement, equal displacement, or the gradient approach (1). Typically, a cross-reactant is conceptually regarded as an interferent causing positive bias in the assay results. When an immunoassay is “rushed” (i.e., the signal is measured well before reaching equilibrium), the effect of the cross-reactant is enhanced, in essence increasing the positive bias (2). That is the expected phenomenon. The unexpected is the observation that cross-reactivity can also lead to suppression in recovery (3), i.e., a negative bias. In other words, the result obtained for a constant amount of the analyte in a sample is lower when the cross-reactant is present than when it is absent. This indicates that rigorous characterization of cross-reactivity requires studies in the presence of the analyte (1). Suppression of assay results induced by cross-reactivity was first described in Clinical Chemistry in 1996 for digoxin immunoassays (4). In retrospect, however, previous observations hinted at this problem. Kanan et al. (5) reported that in plasma obtained from patients suspected of having increased concentrations of digoxin-like immunoreactive factor, recovery of digoxin was lower in one assay than in others. Additional immunoassays have also been shown to be subject to suppression of results caused by cross-reactivity (6)(7). In addition to digoxin-like immunoreactive factor, other cross-reactants, such as progesterone (7), digitoxin (8), or oleandrin (3), have been reported to suppress recovery of digoxin in various immunoassays. A proposed mechanism for the observed suppression of results lies in the physical design (i.e., architecture) of the assay and is fundamentally based on the fact that for binding of small ligands to antibodies, the rates of association are comparable for primary ligand and cross-reactant. It is the rate of dissociation that is different and accounts for the lower binding affinity of the cross-reactant molecule (2). The microparticle enzyme immunoassay (MEIA) for digoxin (Abbott Diagnostics) demonstrates a thermodynamic justification for this phenomenon. In this assay, digoxin in the sample (shown as gray boxes in Fig. 1A) binds to anti-digoxin polyclonal antibodies complexed to microparticles (MP in Fig. 1A). An aliquot of this reaction mixture is transferred to the matrix cell where the microparticle binds to glass fiber. The matrix cell is then washed to remove the unbound material. The enzyme (alkaline phosphatase) complexed to digoxin serves as the tracer binding to unoccupied antibody sites. The instrument measures the conversion of substrate (4-methylumbelliferyl phosphate) to a fluorescent product by the enzyme. The amount of product formed is inversely proportional to the concentration of digoxin present in the sample. A cross-reactant (shown as hatched boxes in Fig. 1B) such as progesterone competes with digoxin for binding to the antibody sites with similar association rate constants. However, during the wash step, the dissociation rate for the cross-reactant is greater than that for the primary ligand and allows more unoccupied sites to become available to bind tracer. The outcome of such a time-dependent competition is reduced recovery attributable to increased tracer binding and product formation. Therefore, the components necessary to create the environment for this to occur are the presence of a cross-reacting substance in the sample and a wash (or separation) step in the assay protocol. Short assay incubation times can exaggerate the extent of interference. What is very relevant is that the phenomenon of suppression caused by cross-reactivity is not limited to any particular cross-reactant or to any particular immunoassay as long as the above-mentioned criteria are met. To our knowledge, other explanations or plausible mechanisms have not been reported. Mechanism of suppression of results by cross-reactivity. (A), format when only the primary ligand (gray boxes) is present. MP denotes a microparticle with anti-digoxin antibodies attached to it. In this assay, the amount of product generated as a result of enzymatic activity is inversely proportional to the concentration of the analyte in the sample. (B), when a cross-reactant (hatched boxes) is also present along with the primary ligand, it will compete for
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cited by
Performance Characteristics of a Novel Tandem Mass Spectrometry Assay For Serum Testosterone
Kushnir, M. M., Rockwood, A. L., Roberts, W. L., Pattison, E. G., Bunker, A. M., Fitzgerald, R. L., & Meikle, A. W. (2006) · Clinical Chemistry · doi:10.1373/clinchem.2005.052167 · Cited by 170
Prasad, S., Lassetter, B., Welker, K. M., & Mehta, P. H. (2019) · Psychoneuroendocrinology · doi:10.1016/j.psyneuen.2019.104373 · Cited by 41
An Optimized Approach for Protein Residue Extraction and Identification from Ceramics After Cooking
Barker, A., Venables, B., Stevens, S. M., Seeley, K. W., Wang, P., & Wolverton, S. (2012) · Journal of Archaeological Method and Theory · doi:10.1007/s10816-011-9120-5 · Cited by 36
Metadata
- Authors
- Valdes, R., & Jortani, S. A.
- Venue
- Clinical Chemistry
- Publisher
- Oxford University Press (OUP)
- Published
- 2002-03-01
- Volume
- 48
- Issue
- 3
- Pages
- 405-406
- ISSN
- 0009-9147 (ISSN-L)
- 1530-8561
- OpenAlex
- W1898565573
- Cited by
- 51
- References
- 0
- FWCI
- 1.84
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Unexpected Suppression of Immunoassay Results by Cross-Reactivity: Now a Demonstrated Cause for Concern. https://app.cochat.ai/papers/10.1093/clinchem/48.3.405
- The paper
- Valdes, R., & Jortani, S. A. (2002). Unexpected Suppression of Immunoassay Results by Cross-Reactivity: Now a Demonstrated Cause for Concern. Clinical Chemistry. https://doi.org/10.1093/clinchem/48.3.405
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.