Selection, Preparation, and Characterization of Commutable Frozen Human Serum Pools as Potential Secondary Reference Materials for Lipid and Apolipoprotein Measurements: Study within the Framework of the Dutch Project “Calibration 2000”
The DOI matches its Crossref record; 3 of the 3 references it cites were found in a registry or an index.
Abstract
Background: The Dutch project “Calibration 2000” aims at harmonization of laboratory results via calibration by development of matrix-based secondary reference materials. We considered the selection, preparation, and characterization of 34 potential reference materials (PRMs). Methods: Sixteen PRMs were prepared either strictly according to the NCCLS C37-A protocol or in a less stringent and more convenient way. In addition, 18 commercial, so-called human serum-based calibrators or controls were purchased and tested. Lipoprotein integrity was evaluated by examining the physicochemical characteristics of the materials. Commutability of the PRMs was assessed in 86 Dutch clinical laboratories, using a multicenter split-patient-sample between-field-method (twin-study) design. Normalized residuals of the PRMs with respect to the patient regression lines were calculated; in addition, the extra contribution of each PRM to the total measurement uncertainty (CVNetto) was calculated. On the basis of these results, the most native PRM was selected to investigate its potential to reduce interlaboratory variation and to improve lipid and apolipoprotein standardization. Results: In general, only the NCCLS C37-A-type materials displayed normalized residuals below the decision limit for commutability and had small CVNetto values ranging between 0 and 3.8%. This contrasts with the findings in regularly pooled frozen sera and lyophilized cryoprotected PRMs. In two subsequent external quality assessment surveys, the NCCLS type C37-A materials contributed to reducing the intermethod lipid and (apo)lipoprotein variation to ∼2–4%. Conclusions: NCCLS C37-A materials have a strong potential as secondary reference materials, not only for cholesterol but also for HDL-cholesterol, LDL-cholesterol, triglyceride, and apolipoprotein measurements.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Passing, H., & Bablok, W. (1983) · cclm · doi:10.1515/cclm.1983.21.11.709 · Cited by 1,513 · cited 1×
“The intermethod relationship was assessed using bivariate regression analysis according to Passing and Bablok (19, 20).”
MethodNauck, M., Warnick, G. R., & Rifai, N. (2002) · Clinical Chemistry · doi:10.1093/clinchem/48.2.236 · Cited by 442
Warnick, G. R., & Wood, P. D. (1995) · Clinical Chemistry · doi:10.1093/clinchem/41.10.1427 · Cited by 212
Cited by
Proficiency Testing/External Quality Assessment: Current Challenges and Future Directions
Miller, W. G., Jones, G. R., Horowitz, G. L., & Weykamp, C. (2011) · Clinical Chemistry · doi:10.1373/clinchem.2011.168641 · Cited by 341
Self-citationCreatinine Measurement: State of the Art in Accuracy and Interlaboratory Harmonization
Greg Miller, W., Myers, G. L., Ashwood, E. R., Killeen, A. A., Wang, E., Thienpont, L. M., & Siekmann, L. (2005) · Archives of Pathology & Laboratory Medicine · doi:10.5858/2005-129-297-cmsota · Cited by 286
Langlois, M. R., Chapman, M. J., Cobbaert, C., Mora, S., Remaley, A. T., Ros, E., Watts, G. F., Borén, J., Baum, H., Bruckert, E., et al. (2018) · Clinical Chemistry · doi:10.1373/clinchem.2018.287037 · Cited by 260
Self-citationMiller, W. G., Myers, G. L., Ashwood, E. R., Killeen, A. A., Wang, E., Ehlers, G. W., Hassemer, D., Lo, S. F., Seccombe, D., Siekmann, L., et al. (2008) · Archives of Pathology & Laboratory Medicine · doi:10.5858/2008-132-838-sotait · Cited by 87
Langlois, M. R., Descamps, O. S., van der Laarse, A., Weykamp, C., Baum, H., Pulkki, K., von Eckardstein, A., De Bacquer, D., Borén, J., Wiklund, O., et al. (2014) · Atherosclerosis · doi:10.1016/j.atherosclerosis.2013.12.016 · Cited by 72
“Quality of the Dutch EQA program has been described previously [16,17].”
BackgroundSelf-citationQuality requirements for quantitative clinical chemistry proteomics
(2014) · Translational Proteomics · doi:10.1016/j.trprot.2013.10.001 · Cited by 11
Metadata
- Authors
- Cobbaert, C., Weykamp, C., Baadenhuijsen, H., Kuypers, A., Lindemans, J., & Jansen, R.
- Venue
- Clinical Chemistry
- Publisher
- Oxford University Press (OUP)
- Published
- 2002-09-01
- Volume
- 48
- Issue
- 9
- Pages
- 1526-1538
- ISSN
- 0009-9147 (ISSN-L)
- 1530-8561
- OpenAlex
- W1933615475
- Cited by
- 45
- References
- 3
- FWCI
- 1.65
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.