Creatinine Measurement: State of the Art in Accuracy and Interlaboratory Harmonization

Greg Miller, W., Myers, Gary L., Ashwood, Edward R., Killeen, Anthony A., Wang, Edward, Thienpont, Linda M., Siekmann, Lothar·2005·Archives of Pathology & Laboratory Medicine

Abstract

Context. —The National Kidney Disease Education Program recommends calculating glomerular filtration rate from serum creatinine concentration. Accurate creatinine measurements are necessary for this calculation. Objective. —To evaluate the state of the art in measuring serum creatinine, as well as the ability of a proficiency testing program to measure bias for individual laboratories and method peer groups. Design. —A fresh-frozen, off-the-clot pooled serum specimen plus 4 conventional specimens were sent to participants in the College of American Pathologists Chemistry Survey for assay of creatinine. Creatinine concentrations were assigned by isotope dilution mass spectrometry reference measurement procedures. Participants. —Clinical laboratories with an acceptable result for all 5 survey specimens (n = 5624). Results. —The fresh frozen serum (FFS) specimen had a creatinine concentration of 0.902 mg/dL (79.7 μmol/L). Mean bias for 50 instrument-method peer groups varied from −0.06 to 0.31 mg/dL (−5.3 to 27.4 μmol/L), with 30 (60%) of 50 peer groups having significant bias ( P < .001). The bias variability was related to instrument manufacturer ( P ≤ .001) rather than method type ( P = .02) with 24 (63%) of 38 alkaline picric acid methods and with 6 (50%) of 12 enzymatic methods having significant biases. Two conventional specimens had creatinine concentrations of 0.795 and 2.205 mg/dL (70.3 and 194.9 μmol/L) and had apparent survey biases significantly different ( P < .001) from that of the FFS specimen for 34 (68%) and 35 (70%) of 50 peer groups, respectively. Conclusions. —Thirty of 50 peer groups had significant bias for creatinine. Bias was primarily associated with instrument manufacturer, not with type of method used. Proficiency testing using a commutable specimen measured participant bias versus a reference measurement procedure and provided trueness surveillance of instrument-method peer groups.

Overview

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.5858/2005-129-297-cmsota

Cites

13 of 13 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Show all 13 references

Cited by

See all 286

Metadata

Authors
Greg Miller, W., Myers, G. L., Ashwood, E. R., Killeen, A. A., Wang, E., Thienpont, L. M., & Siekmann, L.
Venue
Archives of Pathology & Laboratory Medicine
Published
2005-03
Volume
129
Issue
3
Pages
297-304
ISSN
  • 0003-9985 (ISSN-L)
  • 1543-2165
OpenAlex
W1840746889
Cited by
286
References
13
FWCI
13.50
Type
journal article
Language
en
Open access
closed

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Creatinine Measurement: State of the Art in Accuracy and Interlaboratory Harmonization. https://app.cochat.ai/papers/10.5858/2005-129-297-cmsota
The paper
Greg Miller, W., Myers, G. L., Ashwood, E. R., Killeen, A. A., Wang, E., Thienpont, L. M., & Siekmann, L. (2005). Creatinine Measurement: State of the Art in Accuracy and Interlaboratory Harmonization. Archives of Pathology & Laboratory Medicine. https://doi.org/10.5858/2005-129-297-cmsota

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.