Comparison of Proteomic Measurements Across Platforms in the Atherosclerosis Risk in Communities (ARIC) Study

Open access
Rooney, Mary R, Chen, Jingsha, Ballantyne, Christie M, Hoogeveen, Ron C, Tang, Olive, Grams, Morgan E, Tin, Adrienne, Ndumele, Chiadi E +5 more·2023·Clinical Chemistry

Abstract

Background: The plasma proteome can be quantified using different types of highly multiplexed technologies, including aptamer-based and proximity-extension immunoassay methods. There has been limited characterization of how these protein measurements correlate across platforms and with absolute measures from targeted immunoassays. Methods: We assessed the comparability of (a) highly multiplexed aptamer-based (SomaScan v4; Somalogic) and proximity-extension immunoassay (OLINK Proseek® v5003; Olink) methods in 427 Atherosclerosis Risk in Communities (ARIC) Study participants (Visit 5, 2011–2013), and (b) 18 of the SomaScan protein measurements against targeted immunoassays in 110 participants (55 cardiovascular disease cases, 55 controls). We calculated Spearman correlations (r) between the different measurements and compared associations with case-control status. Results: There were 417 protein comparisons (366 unique proteins) between the SomaScan and Olink platforms. The average correlation was r = 0.46 (range: −0.21 to 0.97; 79 [19%] with r ≥ 0.8). For the comparison of SomaScan and targeted immunoassays, 6 of 18 assays (growth differentiation factor 15 [GDF15], interleukin-1 receptor-like 1 [ST2], interstitial collagenase [MMP1], adiponectin, leptin, and resistin) had good correlations (r ≥ 0.8), 2 had modest correlations (0.5 ≤ r < 0.8; osteopontin and interleukin-6 [IL6]), and 10 were poorly correlated (r < 0.5; metalloproteinase inhibitor 1 [TIMP1], stromelysin-1 [MMP3], matrilysin [MMP7], C-C motif chemokine 2 [MCP1], interleukin-10 [IL10], vascular cell adhesion protein 1 [VCAM1], intercellular adhesion molecule 1 [ICAM1], interleukin-18 [IL18], tumor necrosis factor [TNFα], and visfatin) overall. Correlations for SomaScan and targeted immunoassays were similar according to case status. Conclusions: There is variation in the quantitative measurements for many proteins across aptamer-based and proximity-extension immunoassays (approximately 1/2 showing good or modest correlation and approximately 1/2 poor correlation) and also for correlations of these highly multiplexed technologies with targeted immunoassays. Design and interpretation of protein quantification studies should be informed by the variation across measurement techniques for each protein.

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1093/clinchem/hvac186

Cites

3 of 3 references verified0 citation errors · 0 not checkable

Statistics

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Claims other papers make about this one

Each is checked against this paper’s own text: does it say what the citing paper says it does?

  • Existing comparisons of SomaScan and Olink platforms have shown that some protein measures are not concordant, with median inter-platform correlation = 0.38-0.50, which challenges the accuracy of protein associations observed.

    “However, existing comparisons of SomaScan and Olink platforms have shown that some protein measures are not concordant (median inter-platform correlation = 0.38–0.50)11–16, which challenges the accuracy of protein associations observed, especially as there is no “gold-standard” reference measure for most proteins.”

    From Cross-ancestry comparison of aptamer and antibody protein measures (2026) · cites this paper as [14]

    The Results (page unavailable) report poor SomaScan–Olink correlations for 47% of protein pairs, and the Discussion cautions that divergent measurements complicate interpretation of biological associations. The Discussion reports a prior comparison with median r=0.38, but this paper’s own median is 0.53, outside the cited 0.38–0.50 range. It also cautions that disagreement does not establish which assay is more accurate. (LLM, conf=0.86)

    Experimental — accuracy not yet measured

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Cites

Flagged first, then by citations

Research works printed but not linked: 15

  1. [2]Assarsson E, Lundberg M, Holmquist G, Björkesten J, Bucht Thorsen S, Ekman D, Eriksson A, Homogenous 96-Plex PEA Immunoassay Exhibiting High Sensitivity, Specificity, and Excellent Scalability. PloS One 2014; 9: e95192. 24755770 PMC3995906
  2. [5]Liu RX, Thiessen-Philbrook HR, Vasan RS, Coresh J, Ganz P, Bonventre JV, Kimmel PL, Comparison of proteomic methods in evaluating biomarker-AKI associations in cardiac surgery patients. Transl Res 2021; 238: 49– 62. 34343625 PMC8572170
  3. [6]Kukova LZ, Mansour SG, Coca SG, de Fontnouvelle CA, Thiessen-Philbrook HR, Shlipak MG, El-Khoury JM, Comparison of Urine and Plasma Biomarker Concentrations Measured by Aptamer-Based versus Immunoassay Methods in Cardiac Surgery Patients. J Appl Lab Med 2019; 4: 331– 42. 31659071
  4. [7]Tin A, Yu B, Ma J, Masushita K, Daya N, Hoogeveen RC, Ballantyne CM, Reproducibility and Variability of Protein Analytes Measured Using a Multiplexed Modified Aptamer Assay. J Appl Lab Med 2019; 4: 30– 9. 31639705 PMC6814271
  5. [8]Lopez-Silva C, Surapaneni A, Coresh J, Reiser J, Parikh CR, Obeid W, Grams ME, Comparison of Aptamer-Based and Antibody-Based Assays for Protein Quantification in Chronic Kidney Disease. Clin J Am Soc Nephrol 2022: CJN.11700921. PMC8975030 35197258
  6. [10]Walker KA, Chen J, Zhang J, Fornage M, Yang Y, Zhou L, Grams ME, Large-scale plasma proteomic analysis identifies proteins and pathways associated with dementia risk. Nat Aging 2021; 1: 473– 89. PMC10154040 37118015
  7. [11]Candia J, Cheung F, Kotliarov Y, Fantoni G, Sellers B, Griesman T, Huang J, Assessment of Variability in the SOMAscan Assay. Sci Rep 2017; 7: 14248. 29079756 PMC5660188
  8. [12]Williams SA, Kivimaki M, Langenberg C, Hingorani AD, Casas JP, Bouchard C, Jonasson C, Plasma protein patterns as comprehensive indicators of health. Nat Med 2019; 25: 1851– 7. 31792462 PMC6922049
  9. [13]Bland JM, Altman DG. Measurement error proportional to the mean. BMJ 1996; 313: 106. 8688716 PMC2351517
  10. [14]McKee PA, Castelli WP, McNamara PM, Kannel WB. The natural history of congestive heart failure: the Framingham study. N Engl J Med 1971; 285: 1441– 6. 5122894
  11. [15]Yang J, Brody EN, Murthy AC, Mehler RE, Weiss SJ, DeLisle RK, Ostroff R, Impact of Kidney Function on the Blood Proteome and on Protein Cardiovascular Risk Biomarkers in Patients With Stable Coronary Heart Disease. J Am Heart Assoc 2020; 9: e016463. 32696702 PMC7792282
  12. [16]Sun BB, Maranville JC, Peters JE, Stacey D, Staley JR, Blackshaw J, Burgess S, Genomic atlas of the human plasma proteome. Nature 2018; 558: 73– 9. 29875488 PMC6697541
  13. [17]Petrera A, von Toerne C, Behler J, Huth C, Thorand B, Hilgendorff A, Hauck SM. Multiplatform Approach for Plasma Proteomics: Complementarity of Olink Proximity Extension Assay Technology to Mass Spectrometry-Based Protein Profiling. J Proteome Res 2021; 20: 751– 62. 33253581
  14. [18]Graumann J, Finkernagel F, Reinartz S, Stief T, Brödje D, Renz H, Jansen JM, Multi-platform Affinity Proteomics Identify Proteins Linked to Metastasis and Immune Suppression in Ovarian Cancer Plasma. Front Oncol 2019; 9: 1150-. 31737572 PMC6839336
  15. [19]Wollert KC, Kempf T, Wallentin L. Growth Differentiation Factor 15 as a Biomarker in Cardiovascular Disease. Clin Chem 2017; 63: 140– 51. 28062617

Cited by

Sorted by citations

Authors
Rooney, M. R., Chen, J., Ballantyne, C. M., Hoogeveen, R. C., Tang, O., Grams, M. E., Tin, A., Ndumele, C. E., Zannad, F., Couper, D. J., Tang, W., Selvin, E., & Coresh, J.
Affiliations
  • Johns Hopkins University (US)
  • Baylor College of Medicine (US)
  • University of Mississippi Medical Center (US)
  • Inserm (FR)
  • University of North Carolina at Chapel Hill (US)
  • University of Minnesota (US)
Corresponding author
  • Rooney, Mary R
  • Coresh, Josef
Venue
Clinical Chemistry
Published
2022-12-12
Volume
69
Issue
1
Pages
68-79
ISSN
  • 0009-9147 (ISSN-L)
  • 1530-8561
OpenAlex
W4311200104
Cited by
77
References
3
Citations by year
2026: 27 · 2025: 30 · 2024: 11 · 2023: 9
FWCI
4.16
Type
journal article
Language
en
Open access
Locations
  • Clinical Chemistry (publishedVersion)
  • PubMed (publishedVersion)
  • PubMed Central (acceptedVersion)
  • HAL (submittedVersion, other-oa)
Funders
  • NHLBI
  • NIH
  • Department of Health and Human Services
  • NIDDK
Topics
  • Life Sciences › Biochemistry, Genetics and Molecular Biology › Molecular Biology › Advanced Biosensing Techniques and Applications
  • Physical Sciences › Chemistry › Spectroscopy › Advanced Proteomics Techniques and Applications
  • Health Sciences › Medicine › Physiology › Clinical Laboratory Practices and Quality Control
Keywords
plasma proteome, cardiovascular disease, atherosclerosis, inflammatory markers, matrix metalloproteinases, adipokines, growth differentiation factor 15, ST2, Atherosclerosis Risk in Communities Study
MeSH terms
Humans, Immunoassay, Interleukin-6, Proteomics, Atherosclerosis, Adiponectin

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review