The DOI matches its Crossref record; 3 of the 3 references it cites were found in a registry or an index.
Development and validation of an integrated diagnostic algorithm derived from parameters monitored in implantable devices for identifying patients at risk for heart failure hospitalization in an ambulatory setting
Open accessAbstract
BACKGROUND: We developed and validated a heart failure (HF) risk score combining daily measurements of multiple device-derived parameters. METHODS: Heart failure patients from clinical studies with implantable devices were used to form two separate data sets. Daily HF scores were estimated by combining changes in intra-thoracic impedance, atrial fibrillation (AF) burden, rapid rate during AF, %CRT pacing, ventricular tachycardia, night heart rate, heart rate variability, and activity using a Bayesian model. Simulated monthly follow-ups consisted of looking back at the maximum daily HF risk score in the preceding 30 days, categorizing the evaluation as high, medium, or low risk, and evaluating the occurrence of HF hospitalizations in the next 30 days. We used an Anderson-Gill model to compare survival free from HF events in the next 30 days based on risk groups. RESULTS: The development data set consisted of 921 patients with 9790 patient-months of data and 91 months with HF hospitalizations. The validation data set consisted of 1310 patients with 10 655 patient-months of data and 163 months with HF hospitalizations. In the validation data set, 10% of monthly evaluations in 34% of the patients were in the high-risk group. Monthly diagnostic evaluations in the high-risk group were 10 times (adjusted HR: 10.0; 95% CI: 6.4-15.7, P < 0.001) more likely to have an HF hospitalization (event rate of 6.8%) in the next 30 days compared with monthly evaluations in the low-risk group (event rate of 0.6%). CONCLUSION: An HF score based on implantable device diagnostics can identify increased risk for HF hospitalization in the next 30 days.
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
ESC Guidelines for the Diagnosis and Treatment of Acute and Chronic Heart Failure 2012
Authors/Task Force Members, McMurray, J. J. V., Adamopoulos, S., Anker, S. D., Auricchio, A., Böhm, M., Dickstein, K., Falk, V., Filippatos, G., Fonseca, C., et al. (2012) · European Journal of Heart Failure · doi:10.1093/eurjhf/hfs105 · Cited by 2,644
Peterson, P. N., Rumsfeld, J. S., Liang, L., Albert, N. M., Hernandez, A. F., Peterson, E. D., Fonarow, G. C., & Masoudi, F. A. (2010) · Circulation: Cardiovascular Quality and Outcomes · doi:10.1161/circoutcomes.109.854877 · Cited by 568
The CONNECT (Clinical Evaluation of Remote Notification to Reduce Time to Clinical Decision) Trial
Crossley, G. H., Boyle, A., Vitense, H., Chang, Y., & Mead, R. H. (2011) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2010.12.012 · Cited by 526
Self-citation
Cited by
Risk Prediction in Patients With Heart Failure
Rahimi, K., Bennett, D., Conrad, N., Williams, T. M., Basu, J., Dwight, J., Woodward, M., Patel, A., McMurray, J., & MacMahon, S. (2014) · JACC: Heart Failure · doi:10.1016/j.jchf.2014.04.008 · Cited by 410
A Multisensor Algorithm Predicts Heart Failure Events in Patients With Implanted Devices
Boehmer, J. P., Hariharan, R., Devecchi, F. G., Smith, A. L., Molon, G., Capucci, A., An, Q., Averina, V., Stolen, C. M., Thakur, P. H., et al. (2017) · JACC: Heart Failure · doi:10.1016/j.jchf.2016.12.011 · Cited by 393
Avoiding non-responders to cardiac resynchronization therapy: a practical guide
Daubert, C., Behar, N., Martins, R. P., Mabo, P., & Leclercq, C. (2016) · European Heart Journal · doi:10.1093/eurheartj/ehw270 · Cited by 311
Remote management of heart failure using implantable electronic devices
Morgan, J. M., Kitt, S., Gill, J., McComb, J. M., Ng, G. A., Raftery, J., Roderick, P., Seed, A., Williams, S. G., Witte, K. K., et al. (2017) · European Heart Journal · doi:10.1093/eurheartj/ehx227 · Cited by 268
Self-citationLeth, S., Hansen, J., Nielsen, O., & Dinesen, B. (2017) · Sensors · doi:10.3390/s17010211 · Cited by 64
Metadata
- Authors
- Cowie, M. R., Sarkar, S., Koehler, J., Whellan, D. J., Crossley, G. H., Tang, W. H. W., Abraham, W. T., Sharma, V., & Santini, M.
- Venue
- European Heart Journal
- Publisher
- Oxford University Press (OUP)
- Published
- 2013-03-19
- Volume
- 34
- Issue
- 31
- Pages
- 2472-2480
- ISSN
- 0195-668X (ISSN-L)
- 1522-9645
- OpenAlex
- W2021304734
- Cited by
- 164
- References
- 3
- FWCI
- 8.05
- Type
- journal article
- Language
- en
- Open access
- hybrid
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Development and validation of an integrated diagnostic algorithm derived from parameters monitored in implantable devices for identifying patients at risk for heart failure hospitalization in an ambulatory setting. https://app.cochat.ai/papers/10.1093/eurheartj/eht083
- The paper
- Cowie, M. R., Sarkar, S., Koehler, J., Whellan, D. J., Crossley, G. H., Tang, W. H. W., Abraham, W. T., Sharma, V., & Santini, M. (2013). Development and validation of an integrated diagnostic algorithm derived from parameters monitored in implantable devices for identifying patients at risk for heart failure hospitalization in an ambulatory setting. European Heart Journal. https://doi.org/10.1093/eurheartj/eht083
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.