Role of ventricular autocapture function in increasing longevity of DDDR pacemakers: a prospective study
Open accessThe DOI matches its Crossref record.
Abstract
AIMS: Autocapture is an algorithm for automatic adaptation of ventricular output to capture threshold. The aim of this prospective study was to estimate the effects of ventricular Autocapture algorithm on DDD-DDDR pacemaker longevity. METHODS AND RESULTS: Eighty-three patients implanted with a DDD-DDDR pacemaker (Affinity or Entity; St Jude Medical, USA) were enrolled and the Autocapture function was activated pre-discharge. Ventricular pulse duration was randomly programmed at 0.3 or 0.4 ms, with a cross-over at 8-12 weeks and again at 13-14 months. Diagnostic data were retrieved from device memory and by calculating battery current drain from long-term threshold recordings; device longevity was estimated at the following settings: Autocapture with a pulse duration of 0.3 and 0.4 ms, respectively, standard output (3.5 V, 0.4 ms) and conventional low output programming (2.5 V, 0.4 ms). According to a series of assumptions, Autocapture was associated with a 55-60% increase in estimated device longevity compared with standard output programming and a 6-7% increase in longevity compared with low output programming. No significant differences were found between Autocapture programmed with a pulse duration of 0.3 or 0.4 ms. In projections to a 10-year follow-up, use of the Autocapture function resulted in a 42% reduction in pacing-related estimated costs compared with standard output programming at 3.5 V, 0.4 ms. CONCLUSION: Pacing with constant adaptation of ventricular output in dual-chamber devices has the potential to increase generator longevity and to reduce sizeably pacing-related costs compared with standard programming.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cited by
Self-powered cardiovascular electronic devices and systems
Zheng, Q., Tang, Q., Wang, Z. L., & Li, Z. (2021) · Nature Reviews Cardiology · doi:10.1038/s41569-020-0426-4 · Cited by 348
Next-generation pacemakers: from small devices to biological pacemakers
Cingolani, E., Goldhaber, J. I., & Marbán, E. (2018) · Nature Reviews Cardiology · doi:10.1038/nrcardio.2017.165 · Cited by 194
Boriani, G., Savelieva, I., Dan, G.-A., Deharo, J. C., Ferro, C., Israel, C. W., Lane, D. A., La Manna, G., Morton, J., Mitjans, A. M., et al. (2015) · EP Europace · doi:10.1093/europace/euv202 · Cited by 182
Self-citationMonitoring capabilities of cardiac rhythm management devices
Andrikopoulos, G., Tzeis, S., Theodorakis, G., & Vardas, P. (2010) · Europace · doi:10.1093/europace/eup317 · Cited by 5
Metadata
- Authors
- Boriani, G., Rusconi, L., Biffi, M., Pavia, L., Sassara, M., Malfitano, D., Bongiorni, M. G., Padeletti, L., Filice, I., Sanfelici, D., Maffei, P., Vicentini, A., Branzi, A., & on behalf of Autocapture Study Group
- Venue
- EP Europace
- Publisher
- Oxford University Press (OUP)
- Published
- 2006-02-10
- Volume
- 8
- Issue
- 3
- Pages
- 216-220
- ISSN
- 1099-5129 (ISSN-L)
- 1532-2092
- OpenAlex
- W2161191187
- Cited by
- 57
- References
- 0
- FWCI
- 3.72
- Type
- journal article
- Language
- en
- Open access
- green
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.