Practice Parameters for the Use of Auto-Titrating Continuous Positive Airway Pressure Devices for Titrating Pressures and Treating Adult Patients with Obstructive Sleep Apnea Syndrome
Abstract
Continuous positive airway pressure (CPAP) is used to treat patients with the obstructive sleep apnea syndrome (OSAS). The current standard is for an attendant technician to titrate CPAP during full polysomnography to obtain a fixed single pressure. The patient uses CPAP nightly at this fixed single pressure. Recently, devices using new technology that automatically titrate positive airway pressure (APAP) have become available. Such devices continually adjust pressure, as needed, to maintain airway patency (APAP titration). These adjustments can be made with or without attendant technician intervention. Data obtained during APAP titration can be used to provide a fixed single pressure for subsequent treatment. Alternatively, APAP devices can be used in self-adjusting mode for treatment (APAP treatment). A task force of the Standards of Practice Committee of the American Academy of Sleep Medicine reviewed the available literature. Based on this review, the Standards of Practice Committee developed these practice parameters as a guide to the appropriate use of APAP. Recommendations are as follows: 1) A diagnosis of OSAS must be established by an acceptable method. 2) APAP titration and APAP treatment are not currently recommended for patients with congestive heart failure, significant lung disease (e.g., chronic obstructive pulmonary disease), daytime hypoxemia and respiratory failure from any cause, or prominent nocturnal desaturation other than from OSA (e.g., obesity hypoventilation syndrome). In addition, patients who do not snore (either due to palate surgery or naturally) should not be titrated with an APAP device that relies on vibration or sound in the device's algorithm. 3) APAP devices are not currently recommended for split-night studies since none of the reviewed research studies examined this issue. 4) Certain APAP devices may be used during attended titration to identify by polysomnography a single pressure for use with standard CPAP for treatment of OSA. 5) Once an initial successful attended CPAP or APAP titration has been determined by polysomnography, certain APAP devices may be used in the self-adjusting mode for unattended treatment of patients with OSA. 6) Use of unattended APAP to either initially determine pressures for fixed CPAP or for self-adjusting APAP treatment in CPAP naïve patients is not currently established. 7) Patients being treated with fixed CPAP on the basis of APAP titration or being treated with APAP must be followed to determine treatment effectiveness and safety, and 8) a re-evaluation and, if necessary, a standard attended CPAP titration should be performed if symptoms do not resolve or the CPAP or APAP treatment otherwise appears to lack efficacy.
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
The Occurrence of Sleep-Disordered Breathing among Middle-Aged Adults
Young, T., Palta, M., Dempsey, J., Skatrud, J., Weber, S., & Badr, S. (1993) · New England Journal of Medicine · doi:10.1056/nejm199304293281704 · Cited by 10,183
Sullivan, C., Berthon-Jones, M., Issa, F., & Eves, L. (1981) · The Lancet · doi:10.1016/s0140-6736(81)92140-1 · Cited by 2,731
Autoadjusting CPAP Therapy Based on Impedance Efficacy, Compliance and Acceptance
RANDERATH, W. J., SCHRAEDER, O., GALETKE, W., FELDMEYER, F., & RÜHLE, K.-H. (2001) · American Journal of Respiratory and Critical Care Medicine · doi:10.1164/ajrccm.163.3.2006168 · Cited by 143
Cited by
Adult Obstructive Sleep Apnea Task Force of the American Academy of Sleep Medicine, Epstein, L. J., Kristo, D., Strollo, P. J., Friedman, N., Malhotra, A., Patil, S. P., Ramar, K., Rogers, R., Schwab, R. J., et al. (2009) · Journal of Clinical Sleep Medicine · doi:10.5664/jcsm.27497 · Cited by 3,360
Kushida, C. A., Littner, M. R., Morgenthaler, T., Alessi, C. A., Bailey, D., Coleman, J., Friedman, L., Hirshkowitz, M., Kapen, S., Kramer, M., et al. (2005) · Sleep · doi:10.1093/sleep/28.4.499 · Cited by 2,054
Self-citationKushida, C. A., Littner, M. R., Hirshkowitz, M., Morgenthaler, T. I., Alessi, C. A., Bailey, D., Boehlecke, B., Brown, T. M., Coleman, J., Friedman, L., et al. (2006) · Sleep · doi:10.1093/sleep/29.3.375 · Cited by 775
Self-citationTask Force Members, Kushida, C. A., Chediak, A., Berry, R. B., Brown, L. K., Gozal, D., Iber, C., Parthasarathy, S., Quan, S. F., & Rowley, J. A. (2008) · Journal of Clinical Sleep Medicine · doi:10.5664/jcsm.27133 · Cited by 749
Self-citationEvaluation of Positive Airway Pressure Treatment for Sleep Related Breathing Disorders in Adults
Gay, P., Weaver, T., Loube, D., & Iber, C. (2006) · Sleep · doi:10.1093/sleep/29.3.381 · Cited by 614
Metadata
- Authors
- Littner, M., Hirshkowitz, M., Davila, D., Anderson, W. M., Kushida, C. A., Woodson, B. T., Johnson, S. F., & Wise, M. S.
- Venue
- Sleep
- Publisher
- Oxford University Press (OUP)
- Published
- 2002-01
- Volume
- 25
- Issue
- 2
- Pages
- 143-147
- ISSN
- 0161-8105 (ISSN-L)
- 1550-9109
- OpenAlex
- W84053898
- Cited by
- 183
- References
- 3
- FWCI
- 6.80
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Practice Parameters for the Use of Auto-Titrating Continuous Positive Airway Pressure Devices for Titrating Pressures and Treating Adult Patients with Obstructive Sleep Apnea Syndrome. https://app.cochat.ai/papers/10.1093/sleep/25.2.143
- The paper
- Littner, M., Hirshkowitz, M., Davila, D., Anderson, W. M., Kushida, C. A., Woodson, B. T., Johnson, S. F., & Wise, M. S. (2002). Practice Parameters for the Use of Auto-Titrating Continuous Positive Airway Pressure Devices for Titrating Pressures and Treating Adult Patients with Obstructive Sleep Apnea Syndrome. Sleep. https://doi.org/10.1093/sleep/25.2.143
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.