CoChat

Insufficiency in a New Temporal-Artery Thermometer for Adult and Pediatric Patients

Suleman, Mohammad-Irfan, Doufas, Anthony G., Akça, Ozan, Ducharme, Michel, Sessler, Daniel I. · 2002 · Anesthesia & Analgesia
Cited by 107

The DOI matches its Crossref record; 2 of the 2 references it cites were found in a registry or an index.

Abstract

UNLABELLED: SensorTouch is a new noninvasive temperature monitor and consists of an infrared scanner that detects the highest temperature on the skin of the forehead, presumably over the temporal artery. The device estimates core temperature (T(core)). We tested the hypothesis that the SensorTouch is sufficiently precise and accurate for routine clinical use. We studied adults (n = 15) and children (n = 16) who developed mild fever, a core temperature of at least 37.8 degrees C, after cardiopulmonary bypass. Temperature was recorded at 15-min intervals throughout recovery with the SensorTouch thermometer and from the pulmonary artery (adults) or bladder (children). Pulmonary artery (T(core)) and SensorTouch (T(st)) temperatures correlated poorly in adults: T(core) = 0.7. T(st) + 13, r(2) = 0.3. Infrared and pulmonary artery temperatures differed by 1.3 +/- 0.6 degrees C; 89% of the adult temperatures thus differed by more than 0.5 degrees C. Bladder and infrared temperatures correlated somewhat better in pediatric patients: T(core) = 0.9. T(st) + 12, r(2) = 0.6. Infrared and bladder temperatures in children differed by only 0.3 degrees C, but the SD of the difference was 0.5 degrees C. Thus, 31% of the values in the infants and children differed by more than 0.5 degrees C. IMPLICATIONS: We evaluated a noninvasive infrared forehead thermometer (SensorTouch) in adult and pediatric cardiac patients. Accuracy was poor in the adults and suboptimal in infants and children.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1097/00000539-200207000-00012

Cites

2 of 2 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Show all 2 references

Cited by

See all 107

Metadata

Authors
Suleman, M.-I., Doufas, A. G., Akça, O., Ducharme, M., & Sessler, D. I.
Venue
Anesthesia & Analgesia
Published
2002-07
Volume
95
Issue
1
Pages
67-71
ISSN
0003-2999 (ISSN-L)
OpenAlex
W2080348733
Cited by
107
References
2
FWCI
2.11
Type
journal article
Language
en
Open access
closed

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Insufficiency in a New Temporal-Artery Thermometer for Adult and Pediatric Patients. https://app.cochat.ai/papers/10.1097/00000539-200207000-00012
The paper
Suleman, M.-I., Doufas, A. G., Akça, O., Ducharme, M., & Sessler, D. I. (2002). Insufficiency in a New Temporal-Artery Thermometer for Adult and Pediatric Patients. Anesthesia & Analgesia. https://doi.org/10.1097/00000539-200207000-00012

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.