Forehead Pulse Oximetry: Headband Use Helps Alleviate False Low Readings Likely Related to Venous Pulsation Artifact
Open accessThe DOI matches its Crossref record.
Abstract
Background: This study investigated whether a tensioning headband that applies up to 20 mmHg pressure over a forehead pulse oximetry sensor could improve arterial hemoglobin oxygen saturation reading accuracy in presence of venous pooling and pulsations at the forehead site. Methods: Healthy volunteers were studied breathing room air in supine and various levels of negative incline (Trendelenburg position) using the forehead sensor with the headband adjusted to its maximum and minimum recommended pressure limits. Saturation readings obtained from the forehead sensor with the subjects supine and the headband in place were used as a baseline to compare the effects of negative incline on reading accuracy when using and not using the headband. Occurrences of false low-saturation readings detected by forehead sensors were compared with those from digit sensors. Results: No difference was observed between saturation readings obtained from the forehead sensor in supine and negative incline positions when the headband was applied. Forehead sensor readings obtained while subjects were inclined and the headband was not used were significantly lower (P < 0.05) than the supine readings. There was no statistically significant difference between the digit and forehead sensor in reporting false low-saturation readings when the headband was applied, regardless of body incline. Conclusions: Application of up to 20 mmHg pressure on the forehead pulse oximetry sensor using an elastic tensioning headband significantly reduced reading errors and provided consistent performance when subjects were placed between supine and up to 15 degrees head-down incline (Trendelenburg position).
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cited by
Pulse oximetry: Understanding its basic principles facilitates appreciation of its limitations
Chan, E. D., Chan, M. M., & Chan, M. M. (2013) · Respiratory Medicine · doi:10.1016/j.rmed.2013.02.004 · Cited by 710
Pulse oximetry: fundamentals and technology update
Nitzan, M., Romem, A., & Koppel, R. (2014) · Medical Devices: Evidence and Research · doi:10.2147/mder.s47319 · Cited by 438
“Reflection pulse oximetry, in which the light sources and the photodetector are located on the same surface of the skin, can be applied on any accessible site, and is thus of advantage in low peripheral perfusion conditions.(45,46) The main site used for reflection pulse-oximetry mea-”
Non-contact video-based vital sign monitoring using ambient light and auto-regressive models
Tarassenko, L., Villarroel, M., Guazzi, A., Jorge, J., Clifton, D. A., & Pugh, C. (2014) · Physiological Measurement · doi:10.1088/0967-3334/35/5/807 · Cited by 423
Current progress of photoplethysmography and SPO2 for health monitoring
Tamura, T. (2019) · Biomedical Engineering Letters · doi:10.1007/s13534-019-00097-w · Cited by 247
Pulse Oximetry in Pediatric Practice
Fouzas, S., Priftis, K. N., & Anthracopoulos, M. B. (2011) · Pediatrics · doi:10.1542/peds.2011-0271 · Cited by 219
Metadata
- Authors
- Agashe, G. S., Coakley, J., & Mannheimer, P. D.
- Venue
- Anesthesiology
- Published
- 2006-12-01
- Volume
- 105
- Issue
- 6
- Pages
- 1111-1116
- ISSN
- 0003-3022 (ISSN-L)
- OpenAlex
- W1965466508
- Cited by
- 62
- References
- 0
- FWCI
- 2.93
- Type
- journal article
- Language
- en
- Open access
- bronze
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Forehead Pulse Oximetry: Headband Use Helps Alleviate False Low Readings Likely Related to Venous Pulsation Artifact. https://app.cochat.ai/papers/10.1097/00000542-200612000-00010
- The paper
- Agashe, G. S., Coakley, J., & Mannheimer, P. D. (2006). Forehead Pulse Oximetry: Headband Use Helps Alleviate False Low Readings Likely Related to Venous Pulsation Artifact. Anesthesiology. https://doi.org/10.1097/00000542-200612000-00010
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.