Pulse oximetry: fundamentals and technology update

Open access
Nitzan, Meir, Romem, Ayal, Koppel, Robert·2014·Medical Devices: Evidence and Research

Abstract

Oxygen saturation in the arterial blood (SaO2) provides information on the adequacy of respiratory function. SaO2 can be assessed noninvasively by pulse oximetry, which is based on photoplethysmographic pulses in two wavelengths, generally in the red and infrared regions. The calibration of the measured photoplethysmographic signals is performed empirically for each type of commercial pulse-oximeter sensor, utilizing in vitro measurement of SaO2 in extracted arterial blood by means of co-oximetry. Due to the discrepancy between the measurement of SaO2 by pulse oximetry and the invasive technique, the former is denoted as SpO2. Manufacturers of pulse oximeters generally claim an accuracy of 2%, evaluated by the standard deviation (SD) of the differences between SpO2 and SaO2, measured simultaneously in healthy subjects. However, an SD of 2% reflects an expected error of 4% (two SDs) or more in 5% of the examinations, which is in accordance with an error of 3%-4%, reported in clinical studies. This level of accuracy is sufficient for the detection of a significant decline in respiratory function in patients, and pulse oximetry has been accepted as a reliable technique for that purpose. The accuracy of SpO2 measurement is insufficient in several situations, such as critically ill patients receiving supplemental oxygen, and can be hazardous if it leads to elevated values of oxygen partial pressure in blood. In particular, preterm newborns are vulnerable to retinopathy of prematurity induced by high oxygen concentration in the blood. The low accuracy of SpO2 measurement in critically ill patients and newborns can be attributed to the empirical calibration process, which is performed on healthy volunteers. Other limitations of pulse oximetry include the presence of dyshemoglobins, which has been addressed by multiwavelength pulse oximetry, as well as low perfusion and motion artifacts that are partially rectified by sophisticated algorithms and also by reflection pulse oximetry.

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.2147/mder.s47319

Cites

Statistics

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Claims other papers make about this one

Each is checked against this paper’s own text: does it say what the citing paper says it does?

  • Darker skin contains more melanin, which absorbs more green light than lighter skin.

    “likely because darker skin contains more melanin which absorbs more green light than lighter skin.”

    From Investigating sources of inaccuracy in wearable optical heart rate sensors (2020) · cites this paper as [30]

    The “Pulse oximetry – the technique” section says melanin can absorb light in the red and near-infrared regions. It does not address green light or compare melanin content and light absorption in darker versus lighter skin. (LLM, conf=0.82)

    Experimental — accuracy not yet measured
  • Inaccurate PPG heart rate measurements occur up to 15% more frequently in dark skin than in light skin.

    “previous research demonstrated that inaccurate PPG HR measurements occur up to 15% more frequently in dark skin as compared to light skin”

    From Investigating sources of inaccuracy in wearable optical heart rate sensors (2020) · cites this paper as [30]

    The 'Pulse oximetry – the technique' section says melanin can absorb light used in PPG, but it does not compare measurement accuracy by skin tone. The accuracy section concerns oxygen-saturation measurements, not PPG heart-rate errors. Searches for skin pigmentation, heart rate, and 15% found no passage supporting the claimed difference. (LLM, conf=0.76)

    Experimental — accuracy not yet measured

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Cites

No references could be resolved yet.

Research works printed but not linked: 71

  1. [1]Guyton AC Hall JE Textbook of Medical Physiology 9th ed Philadelphia WB Saunders 1996
  2. [2]Wagner PD Ventilation, blood flow and gas exchange Mason RJ Broaddus VC Murray JF Nadel JA Murray and Nadel’s Textbook of Respiratory Medicine Philadelphia WB Saunders 2005
  3. [4]Gold WM Pulmonary function testing Mason RJ Broaddus VC Murray JF Nadel JA Murray and Nadel’s Textbook of Respiratory Medicine Philadelphia WB Saunders 2005
  4. [5]Yoshiya I Shimady Y Tanake K Spectrophotometric monitoring of arterial oxygen saturation on the fingertip Med Biol Eng Comput 1980 18 1 27 32 7382587
  5. [9]Mannheimer PD Casciani JR Fein ME Nierlich SL Wavelength selection for low-saturation pulse oximetry IEEE Trans Biomed Eng 1997 44 3 148 158 9216128
  6. [10]Zourabian A Siegel A Chance B Ramanujan N Rode M Boas DA Trans-abdominal monitoring of fetal arterial blood oxygenation using pulse oximetry J Biomed Optics 2000 5 4 391 405 11092427
  7. [11]Kyriacou PA Pulse oximetry in the esophagus Physiol Meas 2006 27 1 R1 R35 16365505
  8. [12]Babchenko A Davidson E Ginosar Y Photoplethysmographic measurement of changes in total and pulsatile tissue blood volume, following sympathetic blockade Physiol Meas 2001 22 2 389 396 11411248
  9. [15]Mannheimer PD The light-tissue interaction of pulse oximetry Anesth Analg 2007 105 6 S10 S17 18048891
  10. [16]Nitzan M Engelberg S Three-wavelength technique for the measurement of oxygen saturation in arterial blood and in venous blood J Biomed Opt 2009 14 2 024046 19405774
  11. [17]Batchelder PB Raley DM Maximizing the laboratory setting for testing devices and understanding statistical output in pulse oximetry Anesth Analg 2007 105 Suppl 6 S85 S94 18048904
  12. [18]Crapo RO Pulmonary function testing Crapo JD Glassroth J Karlinsky JB King TE Baum’s Textbook of Pulmonary Diseases 7th ed Philadelphia Lippincott Williams and Wilkins 2004
  13. [19]Seguin P Le Rouzo A Tanguy M Guillou YM Feuillu A Mallédant Y Evidence for the need of bedside accuracy of pulse oximetry in an intensive care unit Crit Care Med 2000 28 3 703 706 10752818
  14. [20]Van de Louw A Cracco C Cerf C Accuracy of pulse oximetry in the intensive care unit Intensive Care Med 2001 27 10 1606 1613 11685301
  15. [21]Perkins GD McAuley DF Giles S Routledge H Gao F Do changes in pulse oximeter oxygen saturation predict equivalent changes in arterial oxygen saturation? Crit Care 2003 7 4 R67 R71 12930558 PMC270702
  16. [22]Fouzas S Priftis KN Anthrcopoulos MB Pulse oximetry in pediatric practice Pediatrics 2011 128 4 740 752 21930554
  17. [23]Poets CF Southall DP Noninvasive monitoring of oxygenation in infants and children: practical considerations and areas of concern Pediatrics 1994 93 5 737 746 8165071
  18. [25]O’Driscoll BR Howard LS Davison AG BTS guideline for emergency oxygen use in adult patients Thorax 2008 63 Suppl 6 vi1 vi68 18838559
  19. [26]Wilson BJ Cowan HJ Lord JA Zuege DJ Zygun DA The accuracy of pulse oximetry in emergency department patients with severe sepsis and septic shock: a retrospective cohort study BMC Emerg Med 2010 10 9 20444248 PMC2876142
  20. [27]Martin DS Grocott MP Oxygen therapy in critical illness: precise control of arterial oxygenation and permissive hypoxemia Crit Care Med 2013 41 2 423 432 23263574
  21. [28]Saugstad OD Aune D Optimal oxygenation of extremely low birth weight infants: a meta-analysis and systematic review of the oxygen saturation target studies Neonatology 2014 105 1 55 63 24247112
  22. [29]Bohnhorst B Peter CS Poets CF Detection of hyperoxaemia in neonates: data from three new pulse oximeters Arch Dis Child Fetal Neonatal Ed 2002 87 3 F217 F219 12390996 PMC1721482
  23. [30]Salyer JW Neonatal and pediatric pulse oximetry Resp Care 2003 48 4 386 396 12667266
  24. [31]Rosychuk RJ Hudson-Mason A Eklund D Lacaze-Masmonteil T Discrepancies between arterial oxygen saturation and functional oxygen saturation measured with pulse oximetry in very preterm infants Neonatology 2012 101 1 14 19 21791935
  25. [32]Mendelson Y Kent JC Variations in optical absorption spectra of adult and fetal hemoglobins and its effect on pulse oximetry IEEE Trans Biomed Eng 1989 36 8 844 848 2474489
  26. [33]Whyte RK Jangaard KA Dooley KC From oxygen content to pulse oximetry: completing the picture in the newborn Acta Anaesthesiol Scand Suppl 1995 107 95 100 8599308
  27. [34]Shiao SY Ou CN Validation of oxygen saturation monitoring in neonates Am J Crit Care 2007 16 2 168 178 17322018 PMC1931484
  28. [35]Mahle WT Newburger JW Matherne GP Role of pulse oximetry in examining newborns for congenital heart disease: a scientific statement from the AHA and AAP Pediatrics 2009 124 2 823 836 19581259
  29. [36]Ewer AK Furmston AT Middleton LJ Pulse oximetry as a screening test for congenital heart defects in newborn infants: a test accuracy study with evaluation of acceptability and cost-effectiveness Health Technol Assess 2012 16 2 1 184 22284744
  30. [37]Mahle WT Martin GR Beekman RH 3rd Endorsement of Health and Human Services recommendation for pulse oximetry screening for critical congenital heart disease Pediatrics 2012 129 1 190 192 22201143
  31. [38]Thangaratinam S Brown K Zamora J Khan KS Ewer AK Pulse oximetry screening for critical congenital heart defects (CCHD) in asymptomatic newborns: a systematic review and meta analysis Lancet 2012 379 9835 2459 2464 22554860
  32. [39]Toffaletti J Zijlstra WG Misconceptions in reporting oxygen saturation Anesth Analg 2007 105 6 S5 S9 18048899
  33. [40]Barker SJ Curry J Redford D Morgan S Measurement of carboxyhemoglobin and methemoglobin by pulse oximetry: a human volunteer study Anesthesiology 2006 105 5 892 897 17065881
  34. [41]Annabi EH Barker SJ Severe methemoglobinemia detected by pulse oximetry Anesth Analg 2009 108 3 898 899 19224799
  35. [42]Feiner JR Rollins MD Sall JW Eilers H Au P Bickler PE Accuracy of carboxyhemoglobin detection by pulse CO-oximetry during hypoxemia Anesth Analg 2013 117 4 847 858 23477959 PMC4476500
  36. [43]Touger M Birnbaum A Wang J Chou K Pearson D Bijur P Performance of the RAD-57 pulse CO-oximeter compared with standard laboratory carboxyhemoglobin measurement Ann Emerg Med 2010 56 4 382 388 20605259
  37. [44]Sebbane M Claret PG Mercier G Emergency department management of suspected carbon monoxide poisoning: role of pulse CO-oximetry Respir Care 2013 58 10 1614 1620 23513247
  38. [45]Mendelson Y Ochs BD Noninvasive pulse oximetry utilizing skin reflectance photoplethysmography IEEE Trans Biomed Eng 1988 35 10 798 805 3192229
  39. [46]Agashe GS Coakley J Mannheimer PD Forehead pulse oximetry headband use helps alleviate false low readings likely related to venous pulsation artifact Anesthesiology 2006 105 6 1111 1116 17122573
  40. [47]Berkenbosch JW Tobias JD Comparison of a new forehead reflectance pulse oximeter sensor with a conventional digit sensor in pediatric patients Respir Care 2006 51 7 726 731 16800905
  41. [48]Schallom L Sona C McSweeney M Mazuski J Comparison of forehead and digit oximetry in surgical/trauma patients at risk for decreased peripheral perfusion Heart Lung 2007 36 3 188 194 17509425
  42. [49]Fernandez M Burns K Calhoun B Saramma G Martin B Weaver C Evaluation of a new pulse oximeter sensor Am J Crit Care 2007 16 2 146 152 17322015
  43. [50]Hodgson CL Tuxen DV Holland AE Keating JL Comparison of forehead Max-Fast pulse oximetry sensor with finger sensor at high positive end-expiratory pressure in adult patients with acute respiratory distress syndrome Anaesth Intensive Care 2009 37 6 953 960 20014602
  44. [51]Kyriacou PA Powell S Langford RM Jones DP Esophageal pulse oximetry utilizing reflectance photoplethysmography IEEE Trans Biomed Eng 2002 49 11 1360 1368 12450366
  45. [52]Chen G Zhu Z Liu J Wei W Esophageal pulse oximetry is more accurate and detects hypoxemia earlier than conventional pulse oximetry during general anesthesia Front Med 2012 6 4 406 410 23054501
  46. [53]Brimacombe J Keller C Successful pharyngeal pulse oximetry in low perfusion states Can J Anaesth 2000 47 9 907 909 10989864
  47. [54]Brimacombe J Keller C Margreiter J A pilot study of left tracheal pulse oximetry Anesth Analg 2000 91 4 1003 1006 11004064
  48. [55]Hynson JM Sessler DI Belani K Thermoregulatory vasoconstriction during propofol/nitrous oxide anesthesia in humans: threshold and oxyhemoglobin saturation Anesth Analg 1992 75 6 947 952 1443713
  49. [56]Schramm WM Bartunek A Gilly H Effect of local limb temperature on pulse oximetry and the plethysmographic pulse wave Int J Clin Monit Comput 1997 14 1 17 22 9127780
  50. [57]Talke P Stapelfeldt C Effect of peripheral vasoconstriction on pulse oximetry J Clin Monit Comput 2006 20 5 305 309 16841243
  51. [58]Reddy KA George B Mohan NN Kumar VJ A novel calibration-free method of measurement of oxygen saturation in arterial blood IEEE Instrum Meas Mag 2009 58 5 1699 1705
  52. [59]Franceschini MA Gratton E Fantini S Noninvasive optical method of measuring tissue and arterial saturation: an application to absolute pulse oximetry of the brain Opt Lett 1999 24 12 829 831 18073868
  53. [60]Nitzan M Babchenko A Khanokh B Taitelbaum H The measurement of oxygen saturation in venous blood by dynamic near infrared spectroscopy J Biomed Optics 2000 5 2 155 162 10938779
  54. [61]Zijlstra WG Buursma A Meeuwsen-van der Roest WP Absorption spectra of human fetal and adult oxyhemoglobin, de-oxyhemoglobin, carboxyhemoglobin, and methemoglobin Clin Chem 1991 37 9 1633 1638 1716537
  55. [62]Cope M The Application of Near Infrared Spectroscopy to Non Invasive Monitoring of Cerebral Oxygenation in the Newborn Infant [doctoral thesis] London University College London 1991
  56. [64]Kim JG Liu H Variation of haemoglobin extinction coefficients can cause errors in the determination of haemoglobin concentration measured by near-infrared spectroscopy Phys Med Biol 2007 52 20 6295 6322 17921586
  57. [65]Nitzan M Noach S Tobal E Calibration-free pulse oximetry based on two wavelengths in the infrared – a preliminary study Sensors 2014 14 4 7420 7434 24763216 PMC4029673
  58. [66]Townshend J Taylor BJ Galland B Williams S Comparison of new generation motion-resistant pulse oximeters J Paediatr Child Health 2006 42 6 359 365 16737478
  59. [67]Wijshoff RW Mischi M Woerlee PH Aarts RM Improving pulse oximetry accuracy by removing motion artifacts from photoplethysmograms using relative sensor motion: a preliminary study Adv Exp Med Biol 2013 789 411 417 23852523
  60. [68]Cannesson M Talke P Recent advances in pulse oximetry F1000 Med Rep 2009 1 pii 66 20948714 PMC2948314
  61. [69]Walton ZD Kyriacou PA Silverman DG Shelley KH Measuring venous oxygenation using the photoplethysmograph waveform J Clin Monit Comput 2010 24 4 295 303 20644985
  62. [70]Thiele RH Tucker-Schwartz JM Lu Y Gillies GT Durieux ME Transcutaneous regional venous oximetry: a feasibility study Anesth Analg 2011 112 6 1353 1357 21613200
  63. [72]Wolf M Ferrari M Quaresima V Progress of near-infrared spectroscopy and topography for brain and muscle clinical applications J Biomed Opt 2007 12 6 062104 18163807
  64. [73]Sorensen LC Greisen G Precision of measurement of cerebral tissue oxygenation index using near-infrared spectroscopy in preterm neonates J Biomed Opt 2006 11 5 054005 17092154
  65. [74]De Blasi RA Ferrari M Natali A Conti G Mega A Gasparetto A Noninvasive measurement of forearm blood flow and oxygen consumption by near-infrared spectroscopy J Appl Physiol 1994 76 3 1388 1393 8005887
  66. [75]Wardle SP Yoxall CW Weindling AM Peripheral oxygenation in hypotensive preterm babies Pediatr Res 1999 45 3 343 349 10088652
  67. [76]Yoxall CW Weindling AM Dawani NH Peart I Measurement of cerebral venous oxyhemoglobin saturation in children by near-infrared spectroscopy and partial jugular venous occlusion Pediatr Res 1995 38 3 319 323 7494653
  68. [77]Elwell CE Matcher SJ Tyszczuk L Meek JH Delpy DT Measurement of cerebral venous saturation in adults using near infrared spectroscopy Adv Exp Med Biol 1997 411 453 460 9269462
  69. [78]Franceschini MA Boas DA Zourabian A Near-infrared spiroximetry: noninvasive measurements of venous saturation in piglets and human subjects J Appl Physiol 2002 92 1 372 384 11744680 PMC3786737
  70. [79]Menssen J Colier W Hopman J Liem D de Korte C A method to calculate arterial and venous saturation from near infrared spectroscopy (NIRS) Adv Exp Med Biol 2009 645 135 140 19227462
  71. [80]Wolf M Duc G Keel M Niederer P Continuous noninvasive measurement of cerebral arterial and venous oxygen saturation at the bedside in mechanically ventilated neonates Crit Care Med 1997 25 9 1579 1582 9295835

Cited by

Sorted by citations

Authors
Nitzan, M., Romem, A., & Koppel, R.
Affiliations
  • Jerusalem College of Technology (IL)
  • Shaare Zedek Medical Center (IL)
  • Northwell Health (US)
Corresponding author
Nitzan, Meir
Venue
Medical Devices: Evidence and Research
Published
2014-07
Volume
7
Pages
231
ISSN
1179-1470 (ISSN-L)
OpenAlex
W2087851757
Cited by
438
References
0
Citations by year
2026: 30 · 2025: 37 · 2024: 57 · 2023: 66 · 2022: 43 · 2021: 53 · 2020: 43 · 2019: 27 · 2018: 28 · 2017: 18 · 2016: 20 · 2015: 16
FWCI
1.55
Type
journal article
Language
en
Open access
Locations
  • Medical Devices Evidence and Research (publishedVersion, cc-by-nc)
  • PubMed Central (publishedVersion, cc-by-nc)
  • PubMed (publishedVersion)
  • DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion, cc-by-sa)
  • Dove Medical Press (Taylor and Francis Group) (submittedVersion, cc-by-nc)
Topics
  • Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
  • Health Sciences › Medicine › Surgery › Hemodynamic Monitoring and Therapy
  • Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Optical Imaging and Spectroscopy Techniques
Keywords
pulse oximetry, oxygen saturation, arterial oxygen saturation, photoplethysmography, supplemental oxygen, retinopathy of prematurity, preterm infants, critically ill patients, motion artifacts, low perfusion

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review