Student or educator? Sign up with your school email and get 6 months of Pro free.
Cited by 216

The DOI matches its Crossref record.

Patellar Component Failure in Cementless Total Knee Arthroplasty

ROSENBERG, AARON G., ANDRIACCHI, THOMAS P., BARDEN, REGINA, GALANTE, JORGE O.·1988·Clinical Orthopaedics and Related Research

Abstract

One hundred twenty-two total knee arthroplasties were performed with porous ingrowth fixation of the patellar component between February 1984 and February 1987. Twelve subsequently have experienced fatigue fracture of the patellar component at the peg-plate junction. All fractured patellar components demonstrated excellent ingrowth and fixation of the porous titanium fiber mesh-peg surfaces with no ingrowth into the porous plate underlying the polyethylene patellar surface. While the initial group of 122 arthroplasties was only slightly greater than one-third male, the preponderance of patellar fractures was in males (ten of 12). Patients with patellar component failure were on average younger and heavier and had a greater range of knee motion than the index group. The average time from implantation to recognition of fracture was 24 months. Biomechanical analysis of the force system about a domed patella demonstrated that loading of the patellofemoral joint results in eccentric loading of the dome surface. Peg fixation not accompanied by fixation of the overlying plate allows these high eccentric forces to load the peg-plate junction in shear with consequent fatigue at the peg-plate junction. These results indicate that the shear forces are sufficiently high to warrant caution in the use of peg-plate systems in which peg in growth without plate ingrowth occurs. Recent studies in dogs show that peg-plate systems consistently demonstrate bone ingrowth into pegs with variable ingrowth into the porous plate. The variable ingrowth in the plate and the expanding indication for total knee arthroplasty to a younger population with increased quadriceps strength and activity level may increase the incidence of failure of patellar components using porous peeplate fixation systems, in which ingrowth does not occur.

Overview

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1097/00003086-198811000-00012

Cites

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cited by

See all 216

Metadata

Authors
ROSENBERG, A. G., ANDRIACCHI, T. P., BARDEN, R., & GALANTE, J. O.
Venue
Clinical Orthopaedics and Related Research
Published
1988-11
Volume
236
Issue
236
Pages
106-114
ISSN
0009-921X (ISSN-L)
OpenAlex
W2169485611
Cited by
216
References
0
FWCI
5.15
Type
journal article
Language
en
Open access
closed

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Patellar Component Failure in Cementless Total Knee Arthroplasty. https://app.cochat.ai/papers/10.1097/00003086-198811000-00012
The paper
ROSENBERG, A. G., ANDRIACCHI, T. P., BARDEN, R., & GALANTE, J. O. (1988). Patellar Component Failure in Cementless Total Knee Arthroplasty. Clinical Orthopaedics and Related Research. https://doi.org/10.1097/00003086-198811000-00012

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.