Quantitation in Positron Emission Computed Tomography: 6. Effect of Nonuniform Resolution
The DOI matches its Crossref record.
Abstract
The intrinsic resolution (FWHM) of coincident bismuth germanate detector pairs was measured for geometries consistent with positron emission computed tomography systems designed for brain imaging. The resolutions were measured as a function of the position in the field of view (FOV) for single slice and multislice configurations for both the image and axial resolutions. System diameters of 66 to 42 cm were simulated, and intrinsic resolution variation for image and axial dimensions for a 24 cm FOV were 14 and 32% and 24 and 39%, respectively. Measurement of axial resolutions of interplane slices of multiplane systems employing interplane septa showed resolution variations for the 24 cm FOV of as large as a factor of three. A simple model was shown to predict the dominant features of the line shapes. Images of phantoms with small axial dimensions indicate that the variations in resolution, particularly in interplane data, seriously compromise the validity of point to point comparison of information within an image or between single plane and interplane images. Serious variations in axial resolution were shown to create artifacts and degrade the image resolution. Index Terms: Emission computed tomography--Image quality.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cited by
Levin, C. S., & Hoffman, E. J. (1999) · Physics in Medicine and Biology · doi:10.1088/0031-9155/44/3/019 · Cited by 551
Self-citationFessler, J. A., & Rogers, W. L. (1996) · IEEE Transactions on Image Processing · doi:10.1109/83.535846 · Cited by 518
Circulatory and Metabolic Correlates of Brain Function in Normal Humans
Raichle, M. E. (1987) · Comprehensive Physiology · doi:10.1002/cphy.cp010516 · Cited by 323
History and future technical innovation in positron emission tomography
Jones, T., & Townsend, D. (2017) · Journal of Medical Imaging · doi:10.1117/1.jmi.4.1.011013 · Cited by 303
MRI-PET Correlation in Three Dimensions Using a Volume-of-Interest (VOI) Atlas
Evans, A. C., Marrett, S., Torrescorzo, J., Ku, S., & Collins, L. (1991) · Journal of Cerebral Blood Flow & Metabolism · doi:10.1038/jcbfm.1991.40 · Cited by 264
Metadata
- Authors
- Hoffman, E. J., Huang, S.-C., Plummer, D., & Phelps, M. E.
- Venue
- Journal of Computer Assisted Tomography
- Published
- 1982-10
- Volume
- 6
- Issue
- 5
- Pages
- 987-999
- ISSN
- 0363-8715 (ISSN-L)
- OpenAlex
- W2082732984
- Cited by
- 117
- References
- 0
- FWCI
- 2.02
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Quantitation in Positron Emission Computed Tomography: 6. Effect of Nonuniform Resolution. https://app.cochat.ai/papers/10.1097/00004728-198210000-00022
- The paper
- Hoffman, E. J., Huang, S.-C., Plummer, D., & Phelps, M. E. (1982). Quantitation in Positron Emission Computed Tomography: 6. Effect of Nonuniform Resolution. Journal of Computer Assisted Tomography. https://doi.org/10.1097/00004728-198210000-00022
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.