Vertebral Mineral Determination by Quantitative Computed Tomography (QCT): Accuracy of Single and Dual Energy Measurements

Glüer, Claus-C., Reiser, Ulrich J., Davis, Craig A., Rutt, Brian K., Genant, Harry K. · 1988 · Journal of Computer Assisted Tomography
Cited by 94

The DOI matches its Crossref record.

Abstract

Quantitative CT (QCT) studies of trabecular vertebral bone tissue have been carried out in vitro on a GE CT/T 9800 scanner. Results of both single energy (SE) 80 kVp and dual energy (DE) 80/140 kVp QCT data are compared with chemical mineral analysis to determine accuracy. We examined 62 vertebral specimens, from 28 cadavers (19 male and 9 female with an age range of 19-93 years, mean = 60.4). Averaging the results of all vertebral bodies of the same individual for SEQCT versus ashweight, we found a correlation coefficient (r) of 0.94 (p less than 0.0001), a standard error of the estimate (SEE) of 12.2 mg/cm3 (calibrated to K2HPO4), with a coefficient of variation (CV) of 13.2% and an average underestimation of bone mineral content of 18.7 mg/cm3. The corresponding DEQCT results were r = 0.98 (p less than 0.0001), SEE = 7.4 mg/cm3, CV = 7.0%, and an average underestimation of 4.9 mg/cm3. The SE and DE results are correlated with r = 0.98 (p less than 0.0001), SEE = 8.0 mg/cm3, and CV = 8.7%. From our SEQCT data and the results of the chemical analysis of bone mineral and fat content we calculated a fat sensitivity of 7.7 mg/cm3 K2HPO4 per 100 mg/cm3 fat change for our scanner. Using an average fat variability of 87.5 mg/cm3, this leads to a fat-related uncertainty for the normative SEQCT data of 6.7 mg/cm3, which is far lower than the normal biological variation of 29.4 mg/cm3. Using tabulated normative data on fat content versus age and versus mineral content of 188 vertebral specimens from five collaborating centers, we derived a correction algorithm for QCT measurement that reduces our average underestimation to 0.88 mg/cm3 with an SEE of 12.1 mg/cm3. Hence, this correction procedure can be used to estimate the fat corrected absolute mineral density for research purposes or for scanners with high fat sensitivity. For the GE CT/T 9800 scanner, with a relatively low fat to mineral sensitivity at 80 kVp, the correction procedure is generally not recommended for clinical studies since it minimizes the average fat induced error but does not reduce the residual, partially fat related uncertainty. Finally, since the fat related uncertainty is small compared to biological variation, the correlation is high between SEQCT and DEQCT, and the radiation dose is lower and the precision higher for SEQCT, we suggest that most clinical diagnostic studies using the GE CT/T 9800 scanner for bone mineral determination employ SEQCT at 80 kVp.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1097/00004728-198803000-00013

Cites

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cited by

See all 94

Metadata

Authors
Glüer, C.-C., Reiser, U. J., Davis, C. A., Rutt, B. K., & Genant, H. K.
Venue
Journal of Computer Assisted Tomography
Published
1988-03
Volume
12
Issue
2
Pages
242-258
ISSN
0363-8715 (ISSN-L)
OpenAlex
W1965258941
Cited by
94
References
0
FWCI
5.09
Type
journal article
Language
en
Open access
closed

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Vertebral Mineral Determination by Quantitative Computed Tomography (QCT): Accuracy of Single and Dual Energy Measurements. https://app.cochat.ai/papers/10.1097/00004728-198803000-00013
The paper
Glüer, C.-C., Reiser, U. J., Davis, C. A., Rutt, B. K., & Genant, H. K. (1988). Vertebral Mineral Determination by Quantitative Computed Tomography (QCT): Accuracy of Single and Dual Energy Measurements. Journal of Computer Assisted Tomography. https://doi.org/10.1097/00004728-198803000-00013

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.