Performance Characteristics of an Eight-Ring Whole Body PET Scanner
Abstract
The technical characteristics of the multislice whole-body positron emission tomographic scanner (model PC4096-15WB Scanditronix) and its performance parameters are described. Spatial resolution at the center of the field of view was found to be 4.9 mm in-plane and 4.6 mm (cross slices) and 6.0 mm (direct slices) in the axial direction. The sensitivity for true and scattered coincidences is approximately 5,000 cps for direct slices and 7,100 cps for cross slices. At an activity concentration of 37 kBq/ml the system deadtime was approximately 5%. By measuring a uniform phantom with a cold cylindrical insert (5.0 cm diameter), the scatter fraction was found to be approximately 5%. The mean global uniformity over all 15 slices was 6.5%, whereas the local uniformity was found to be 4.3%. No systematic nonuniformities were observed. Finally, various methods for attenuation correction (transmission scan, contour finding, ellipse) were utilized to test their effects on the resulting reconstructed images.
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cited by
Rauch, S. L. (1996) · Archives of General Psychiatry · doi:10.1001/archpsyc.1996.01830050014003 · Cited by 1,056
Rauch, S. L. (1994) · Archives of General Psychiatry · doi:10.1001/archpsyc.1994.03950010062008 · Cited by 880
Topographical representations of mental images in primary visual cortex
Kosslyn, S. M., Thompson, W. L., Klm, I. J., & Alpert, N. M. (1995) · Nature · doi:10.1038/378496a0 · Cited by 701
Visual Mental Imagery Activates Topographically Organized Visual Cortex: PET Investigations
Kosslyn, S. M., Alpert, N. M., Thompson, W. L., Maljkovic, V., Weise, S. B., Chabris, C. F., Hamilton, S. E., Rauch, S. L., & Buonanno, F. S. (1993) · Journal of Cognitive Neuroscience · doi:10.1162/jocn.1993.5.3.263 · Cited by 641
Schacter, D. L., Alpert, N. M., Savage, C. R., Rauch, S. L., & Albert, M. S. (1996) · Proceedings of the National Academy of Sciences · doi:10.1073/pnas.93.1.321 · Cited by 552
Metadata
- Authors
- Kops, E. R., Herzog, H., Schmid, A., Holte, S., & Feinendegen, L. E.
- Venue
- Journal of Computer Assisted Tomography
- Published
- 1990-05
- Volume
- 14
- Issue
- 3
- Pages
- 437-445
- ISSN
- 0363-8715 (ISSN-L)
- OpenAlex
- W2061938237
- Cited by
- 194
- References
- 0
- FWCI
- 9.42
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Performance Characteristics of an Eight-Ring Whole Body PET Scanner. https://app.cochat.ai/papers/10.1097/00004728-199005000-00022
- The paper
- Kops, E. R., Herzog, H., Schmid, A., Holte, S., & Feinendegen, L. E. (1990). Performance Characteristics of an Eight-Ring Whole Body PET Scanner. Journal of Computer Assisted Tomography. https://doi.org/10.1097/00004728-199005000-00022
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.