Clinical evaluation of semiautomatic and automatic devices for home blood pressure measurement: comparison between cuff-oscillometric and microphone methods

Imai, Yutaka, Abe, Keishi, Sasaki, Shuichi, Minami, Naoyoshi, Munakata, Masanori, Sakuma, Hiromichi, Hashimoto, Junichiro, Sekino, Hiroshi, Imai, Keiko, Yoshinaga, Kaoru · 1989 · Journal of Hypertension
Cited by 125

The DOI matches its Crossref record.

Abstract

The accuracy and reliability of blood pressure (BP) values were evaluated by comparing values obtained with eight automatic or semiautomatic devices designed for home BP measurement (four microphone devices based on the Korotkoff-sound technique and four cuff-oscillometric devices) with those obtained by the auscultatory method, using a standard mercury sphygmomanometer. Systolic blood pressure (SBP) values obtained using the microphone devices coincided well with those obtained by the auscultatory method. However, these devices produced a certain proportion of errors in the measurement of diastolic blood pressure (DBP), sometimes resulting in recordings at least 25mmHg higher than those obtained by the standard method. The most frequent causes of this phenomenon were an auscultatory (silent) gap and a weak Korotkoff sound after phase IV. A microphone device using a condenser microphone built into the manometer displayed comparatively good acoustic characteristics for determining DBP. All cuff-oscillometric devices demonstrated minimal mean differences and a constant s.d. of mean difference for DBP, with no great differences from the auscultatory method. However, mean differences and s.d.s in SBP measurements using cuff-oscillometric devices were relatively greater than those obtained using some of the microphone devices. Furthermore, the direction of the mean differences in measurements from those obtained with the auscultatory method differed. The error in relation to the auscultatory method tended to be reproducible in the same subjects with both the microphone and the cuff-oscillometric devices. These results indicate that practitioners should select the most appropriate method and/or device by taking into account the factors which may cause measurement error in relation to the auscultatory method in each subject, and should then evaluate, at least once, the difference in BP values obtained using the auscultatory method and using the device. In future, home blood pressure measurement devices for determination of SBP should employ a microphone method, while a method which combines a microphone with a cuff-oscillometric device, thereby compensating for the disadvantage of the Korotkoff-sound signal with the pulse wave signal, should be recommended for measurement of DBP.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1097/00004872-198912000-00009

Cites

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cited by

See all 125

Metadata

Authors
Imai, Y., Abe, K., Sasaki, S., Minami, N., Munakata, M., Sakuma, H., Hashimoto, J., Sekino, H., Imai, K., & Yoshinaga, K.
Venue
Journal of Hypertension
Published
1989-12
Volume
7
Issue
12
Pages
983-990
ISSN
0263-6352 (ISSN-L)
OpenAlex
W2055457636
Cited by
125
References
0
FWCI
0.70
Type
journal article
Language
en
Open access
closed

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Clinical evaluation of semiautomatic and automatic devices for home blood pressure measurement: comparison between cuff-oscillometric and microphone methods. https://app.cochat.ai/papers/10.1097/00004872-198912000-00009
The paper
Imai, Y., Abe, K., Sasaki, S., Minami, N., Munakata, M., Sakuma, H., Hashimoto, J., Sekino, H., Imai, K., & Yoshinaga, K. (1989). Clinical evaluation of semiautomatic and automatic devices for home blood pressure measurement: comparison between cuff-oscillometric and microphone methods. Journal of Hypertension. https://doi.org/10.1097/00004872-198912000-00009

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.