CoChat

Accutracker II (version 30/23) ambulatory blood pressure monitor: clinical validation using the British Hypertension Society and Association for the Advancement of Medical Instrumentation standards

Taylor, Rod, Chidley, Kay, Goodwin, James, Broeders, Martijn, Kirby, Brian · 1993 · Journal of Hypertension
Cited by 41

The DOI matches its Crossref record.

Abstract

OBJECTIVE: To assess the Accutracker II (version 30/23) ambulatory blood pressure monitor by nationally agreed protocols in order to resolve previous conflicting assessments, and to examine the feasibility of combining these protocols into one study. DESIGN: The protocols of the Association for the Advancement of Medical Instrumentation (AAMI) and the British Hypertension Society (BHS) were used simultaneously. SUBJECTS: Five normotensive subjects were used to assess interobserver variation; 30 subjects took part in the field evaluation and 85 in the laboratory evaluation. The latter subjects were selected to cover a wide range of blood pressures and differences in arm circumference. OUTCOME MEASUREMENTS: Classification of device accuracy according to the criteria of the BHS and AAMI protocols. RESULTS: Among 255 observations the mean +/- SD difference between blood pressure measurements by the Accutracker II and the observers was -2.2 +/- 0.8/-3.5 +/- 0.9 mmHg. There was a small but statistically significant difference between one device and the other two, but all three fulfilled the BHS criteria of 95% of measurements falling within 3 mmHg before and after use. In 30 24-h recordings > 80% of the programmed inflations produced valid recordings. Editing criteria accounted for the majority (55%) of invalid readings; weak Korotkoff sounds, imperfect electrocardiogram signals or movement artefact accounted for the remainder. CONCLUSIONS: Version 30/23 of the Accutracker II fulfilled the AAMI criteria; using the BHS system, it was graded A for systolic and C for diastolic blood pressure. Although both protocols were readily combined into one study, they do not assess exactly the same aspects of blood pressure measurement. Previous conflicting evaluations could be due to differences in applying the protocols or may result from modifications in production models. Posture may affect the rating given by the BHS protocol. Until there is general international agreement on the method used to validate ambulatory blood pressure monitors, simultaneous use of both the BHS and AAMI protocols is recommended. In future assessments or investigational use of any similar instruments the model used should be described precisely.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1097/00004872-199311000-00016

Cites

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cited by

See all 41

Metadata

Authors
Taylor, R., Chidley, K., Goodwin, J., Broeders, M., & Kirby, B.
Venue
Journal of Hypertension
Published
1993-11
Volume
11
Issue
11
Pages
1275???1282
ISSN
0263-6352 (ISSN-L)
OpenAlex
W1996171782
Cited by
41
References
0
FWCI
0.31
Type
journal article
Language
en
Open access
closed

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Accutracker II (version 30/23) ambulatory blood pressure monitor: clinical validation using the British Hypertension Society and Association for the Advancement of Medical Instrumentation standards. https://app.cochat.ai/papers/10.1097/00004872-199311000-00016
The paper
Taylor, R., Chidley, K., Goodwin, J., Broeders, M., & Kirby, B. (1993). Accutracker II (version 30/23) ambulatory blood pressure monitor: clinical validation using the British Hypertension Society and Association for the Advancement of Medical Instrumentation standards. Journal of Hypertension. https://doi.org/10.1097/00004872-199311000-00016

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.