Biocompatibility of Fixation Materials in the Brain
The DOI matches its Crossref record.
Abstract
Recent clinical reports documenting passive intracranial translocation of microplates and microscrews have prompted concerns regarding brain biocompatibility and neurotoxicity of fixation hardware used in craniofacial surgery. Although the effects of commercially pure titanium. Vitallium (cobalt-chromium-molybdenum), stainless steel, and various alloys have been well assessed in bone and soft tissues, there are no comprehensive studies of these materials in the brain. To investigate this, the biocompatibility of titanium, vitallium, and 316L stainless steel was evaluated in the rabbit brain and compared with silicone elastomer shunt tubing, a material that is used commonly as a neurosurgical implant material with well-established brain biocompatibility. Forty-eight juvenile New Zealand White rabbits were randomly assigned to one of three groups and underwent placement of either commercially pure titanium microscrews, vitallium microscrews, or 316L monofilament stainless steel wire into the parietal region. Silicone elastomer strips of similar size were implanted in the contralateral hemisphere of each rabbit. Animals were assessed daily for signs of neurotoxicity. Rabbits in each group were sacrificed at 2, 4, 8, and 26 weeks following implantation. Brains were sectioned at the implantation site and were examined by means of standard hematoxylin and eosin stains and with immunohistochemical markers sensitive to inflammatory changes in the brain. None of the animals showed any behavioral changes or neurologic deficits suggestive of either systemic or localized toxicity from the implant materials. Silicone clastomer was found to cause the least amount of inflammation relative to other materials tested at all sacrifice points, suggesting that as a standard neurosurgical implant material, it is an appropriate control for studies of brain biocompatibility. At 2 weeks, titanium was found to cause the largest inflammatory response in surrounding brain parenchyma based on analysis of markers for microglial proliferation, gliosis, and leukocyte infiltration. At the 26-week endpoint of our study, the biocompatibility of titanium was nearly equal to the biocompatibility of vitallium based on all studied markers of inflammation. A progressive increase in the inflammatory response surrounding stainless steel implants was noted at 8 and 26 weeks. Relative to all materials studied, at 26 weeks the greatest leukocyte response was found with stainless steel implants. Our results indicate that at the 26-week end-point of our study, titanium and vitallium incited a similar inflammatory response in the rabbit brain that was greater than the response found with silicone elastomer, a standard neurosurgical implant material, but less than that found with stainless steel wire, which is commonly recommended as an alternative fixation material.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cited by
Response of brain tissue to chronically implanted neural electrodes
Polikov, V. S., Tresco, P. A., & Reichert, W. M. (2005) · Journal of Neuroscience Methods · doi:10.1016/j.jneumeth.2005.08.015 · Cited by 1,896
“…is complete, and perhaps as long as the material remains in contact with brain tissue (Stensaas and Stensaas, 1976, 1978; Winn t al., 1989; Edell et al., 1992; Menei et al., 1994; Kunkler and Kraig, 1997; Mofid et al., 1997; Emerich et al., 1999; Mokry et al., 2000; Szarowski et al., 2003).”
BackgroundBrain-Controlled Interfaces: Movement Restoration with Neural Prosthetics
Schwartz, A. B., Cui, X. T., Weber, D. J., & Moran, D. W. (2006) · Neuron · doi:10.1016/j.neuron.2006.09.019 · Cited by 792
A Review of Organic and Inorganic Biomaterials for Neural Interfaces
Fattahi, P., Yang, G., Kim, G., & Abidian, M. R. (2014) · Advanced Materials · doi:10.1002/adma.201304496 · Cited by 580
In vivo studies of polypyrrole/peptide coated neural probes
Cui, X. (2003) · Biomaterials · doi:10.1016/s0142-9612(02)00415-5 · Cited by 505
Resorbable Plate Fixation in Pediatric Craniofacial Surgery: Long-term Outcome
Imola, M. J., Hamlar, D. D., Shao, W., Chowdhury, K., & Tatum, S. (2001) · Archives of Facial Plastic Surgery · doi:10.1001/archfaci.3.2.79 · Cited by 163
Metadata
- Authors
- Mofid, M. M., Thompson, R. C., Pardo, C. A., Manson, P. N., & Vander Kolk, C. A.
- Venue
- Plastic & Reconstructive Surgery
- Published
- 1997-07
- Volume
- 100
- Issue
- 1
- Pages
- 14-20
- ISSN
- 0032-1052 (ISSN-L)
- OpenAlex
- W2044217494
- Cited by
- 51
- References
- 0
- FWCI
- 4.35
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Biocompatibility of Fixation Materials in the Brain. https://app.cochat.ai/papers/10.1097/00006534-199707000-00003
- The paper
- Mofid, M. M., Thompson, R. C., Pardo, C. A., Manson, P. N., & Vander Kolk, C. A. (1997). Biocompatibility of Fixation Materials in the Brain. Plastic & Reconstructive Surgery. https://doi.org/10.1097/00006534-199707000-00003
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.