An Ex Vivo Biomechanical Evaluation of an Inflatable Bone Tamp Used in the Treatment of Compression Fracture

Belkoff, Stephen M., Mathis, John M., Fenton, David C., Scribner, Robert M., Reiley, Mark E., Talmadge, Karen·2001·Spine

Abstract

STUDY DESIGN: Ex vivo biomechanical study using osteoporotic cadaveric vertebral bodies. OBJECTIVES: To determine if the inflatable bone tamp (tamp) restores height to compressed vertebral bodies and to compare the biomechanical properties of isolated, fractured osteoporotic vertebral bodies treated by kyphoplasty (tamp) or vertebroplasty. SUMMARY OF BACKGROUND DATA: Previous biomechanical studies have shown that vertebroplasty increases vertebral body strength and restores vertebral body stiffness, but does not restore vertebral body height lost as a result of compression fracture. METHODS: Compression fractures were experimentally created in 16 osteoporotic VBs assigned to either the tamp or percutaneous vertebroplasty group. The tamp treatment consisted of inserting balloon-like devices into the vertebral body, inflating the bone tamp, and filling the void with Simplex P (Howmedica, Rutherford, NJ) bone cement. The percutaneous vertebroplasty treatment consisted of directly injecting Cranioplastic bone cement (CMW, Blackpool, UK) into the vertebral body. Pre- and posttreatment heights were measured, and the repaired vertebral bodies were recompressed to determine posttreatment strength and stiffness values. RESULTS: The tamp treatment resulted in significant restoration (97%) of vertebral body height lost after compression, whereas percutaneous vertebroplasty treatment resulted in a significantly lower restoration of lost height (30%) (P < 0.05). Both treatments resulted in significantly stronger vertebral bodies relative to their initial state (P < 0.05). The tamp treatment restored vertebral body stiffness to initial values, but the percutaneous vertebroplasty treatment did not (P < 0.05). CONCLUSIONS: Tamp treatment resulted in significantly greater height restoration than did percutaneous vertebroplasty, without loss of vertebral body strength or stiffness.

Overview

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1097/00007632-200101150-00008

Cites

13 of 13 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Show all 13 references

Cited by

See all 305

Metadata

Authors
Belkoff, S. M., Mathis, J. M., Fenton, D. C., Scribner, R. M., Reiley, M. E., & Talmadge, K.
Venue
Spine
Published
2001-01
Volume
26
Issue
2
Pages
151-156
ISSN
0362-2436 (ISSN-L)
OpenAlex
W2128268015
Cited by
305
References
13
FWCI
24.53
Type
journal article
Language
en
Open access
closed

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: An Ex Vivo Biomechanical Evaluation of an Inflatable Bone Tamp Used in the Treatment of Compression Fracture. https://app.cochat.ai/papers/10.1097/00007632-200101150-00008
The paper
Belkoff, S. M., Mathis, J. M., Fenton, D. C., Scribner, R. M., Reiley, M. E., & Talmadge, K. (2001). An Ex Vivo Biomechanical Evaluation of an Inflatable Bone Tamp Used in the Treatment of Compression Fracture. Spine. https://doi.org/10.1097/00007632-200101150-00008

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.