Accuracy of aneroid sphygmomanometers in clinical usage: University of Michigan experience
The DOI matches its Crossref record.
Abstract
BACKGROUND: Aneroid manometers are frequently used to measure blood pressure. Aneroid manometers have moving parts that are subject to fatigue. The accuracy duration of the aneroid devices, like most digital devices, is unstudied. It has been accepted that if the aneroid device does not rest at '0' it is inaccurate, but how often is the device inaccurate when it does rest at '0'? METHODS: A Universal Biometer DPM-III measuring unit was used for all of the measurements at 10 University of Michigan Health System sites. A total of 136 aneroid manometers were tested. Two additional aneroid devices were not tested, as the needle did not start within '0'. Static pressure measurements were made at nine levels for all devices: 50, 80, 90, 100, 120, 150, 200 and 250mmHg. RESULTS: The average difference of the nine pressure settings of the whole group was 0.2+/-0.31 (95% confidence interval 0.1-0.2) mmHg. The largest number of devices that were not calibrated within +/-3mmHg was seen at the 150mmHg setting with six (4.4%) of the devices failing. If an accuracy standard of +/-2mmHg was used, the largest number of devices failed at 250mmHg (22 devices, 16.2%). The largest number of devices that were inaccurate was manufactured 6 years prior to testing and was from two sites. CONCLUSIONS: Aneroid devices were accurate. A yearly calibration programme should be performed and a +/-2mmHg standard should be used. Portable aneroid manometers may need to be more frequently calibrated due to the trauma associated with dropping.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cited by
Pickering, T. G., Hall, J. E., Appel, L. J., Falkner, B. E., Graves, J., Hill, M. N., Jones, D. W., Kurtz, T., Sheps, S. G., & Roccella, E. J. (2005) · Circulation · doi:10.1161/01.cir.0000154900.76284.f6 · Cited by 3,985
Pickering, T. G., Hall, J. E., Appel, L. J., Falkner, B. E., Graves, J., Hill, M. N., Jones, D. W., Kurtz, T., Sheps, S. G., & Roccella, E. J. (2005) · Hypertension · doi:10.1161/01.hyp.0000150859.47929.8e · Cited by 2,051
Principles and Techniques of Blood Pressure Measurement
Ogedegbe, G., & Pickering, T. (2010) · Cardiology Clinics · doi:10.1016/j.ccl.2010.07.006 · Cited by 554
Kallioinen, N., Hill, A., Horswill, M. S., Ward, H. E., & Watson, M. O. (2017) · Journal of Hypertension · doi:10.1097/hjh.0000000000001197 · Cited by 357
Paridon, S. M., Alpert, B. S., Boas, S. R., Cabrera, M. E., Caldarera, L. L., Daniels, S. R., Kimball, T. R., Knilans, T. K., Nixon, P. A., Rhodes, J., et al. (2006) · Circulation · doi:10.1161/circulationaha.106.174375 · Cited by 325
Metadata
- Authors
- Yarows, S. A., & Qian, K.
- Venue
- Blood Pressure Monitoring
- Published
- 2001-04
- Volume
- 6
- Issue
- 2
- Pages
- 101-106
- ISSN
- 1359-5237 (ISSN-L)
- OpenAlex
- W2329420534
- Cited by
- 61
- References
- 0
- FWCI
- 37.42
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Accuracy of aneroid sphygmomanometers in clinical usage: University of Michigan experience. https://app.cochat.ai/papers/10.1097/00126097-200104000-00006
- The paper
- Yarows, S. A., & Qian, K. (2001). Accuracy of aneroid sphygmomanometers in clinical usage: University of Michigan experience. Blood Pressure Monitoring. https://doi.org/10.1097/00126097-200104000-00006
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.