Overall accuracy of the BpTRU??????an automated electronic blood pressure device

Mattu, Gurdial S., Heran, Balraj S., Wright, James M. · 2004 · Blood Pressure Monitoring
Cited by 86

The DOI matches its Crossref record; 1 of the 1 references it cites were found in a registry or an index.

Abstract

BACKGROUND: The objective of this report is to combine the data from an earlier adult study with the data from a paediatric study in order to determine the overall accuracy of the BpTRU (BPM-100 model) as compared to the recognized standard, auscultatory mercury sphygmomanometer. DESIGN: The individual blood pressure points recorded for both adult and paediatric studies were compared directly to its corresponding observer reference measurements from data collected and stored from the two separate studies. There were 255 sets of readings in the adult study and 162 sets from the paediatric study, which were combined to make 417 pairs of blood pressure readings for this study. METHODS: The overall observer standard reference mean for the 417 measurements was calculated and the difference between this and the overall mean BPM-100 was calculated with SD and ranges. Measurements within 5, 10 and 15 mmHg agreement were expressed as percentages. RESULTS: A total of 121 subjects were included for this study (85 from the adult study and 36 from the paediatric study). From these, 417 paired measurements were recorded. The mean difference between the BpTRU and the reference standard systolic blood pressure (BP) was 0.47+/-5.40 mmHg with 89.2% measurements within 5 mmHg, 96.4% within 10 mmHg and 99.3% within 15 mmHg. The mean difference between the BpTRU and reference diastolic BP was -2.12+/-5.93 mmHg with 81.1% within 5 mmHg, 92.1% within 10 mmHg and 97.6% within 15 mmHg. CONCLUSION: The BpTRU has been shown to be an accurate non-invasive blood pressure monitoring device in the general population over a wide range of ages (3-83 years). This combined study meets all requirements of the Association of Advancement of Medical Instrumentation and achieved a grade 'A' in the BHS protocol.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1097/00126097-200402000-00009

Cites

1 of 1 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cited by

See all 86

Metadata

Authors
Mattu, G. S., Heran, B. S., & Wright, J. M.
Venue
Blood Pressure Monitoring
Published
2004-02
Volume
9
Issue
1
Pages
47-52
ISSN
1359-5237 (ISSN-L)
OpenAlex
W2010658805
Cited by
86
References
1
FWCI
0.83
Type
journal article
Language
en
Open access
closed

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Overall accuracy of the BpTRU??????an automated electronic blood pressure device. https://app.cochat.ai/papers/10.1097/00126097-200402000-00009
The paper
Mattu, G. S., Heran, B. S., & Wright, J. M. (2004). Overall accuracy of the BpTRU??????an automated electronic blood pressure device. Blood Pressure Monitoring. https://doi.org/10.1097/00126097-200402000-00009

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.