Comparison of ICare Tonometer with Goldmann Applanation Tonometer in Glaucoma Patients
The DOI matches its Crossref record; 22 of the 22 references it cites were found in a registry or an index.
Abstract
PURPOSE: To compare the intraocular pressure (IOP) readings taken with the new ICare tonometer and with the Goldmann applanation tonometer (GAT) and to evaluate the influence of central corneal thickness (CCT) on the IOP measurements. PATIENTS AND METHODS: One eye of 178 consecutive patients with primary open-angle glaucoma underwent ultrasonic CCT measurement, followed by IOP evaluation with the GAT and with the ICare tonometer. The deviation of ICare readings from GAT values, corrected according to the Doughty and Zaman formula, was calculated and correlated to CCT by a linear regression model. The agreement between the 2 devices was assessed by use of the Bland-Altman method. RESULTS: The average CCT was 552+/-39 mum. The mean IOP and the mean corrected IOP with GAT were 19.4+/-5.4 mm Hg, and 18.5+/-5.7 mm Hg, respectively. The mean ICare IOP reading was 18.4+/-5.2 mm Hg. The deviations of ICare readings from corrected GAT values were highly correlated with CCT values (r=0.63, P<0.01). Linear regression analysis showed that a CCT change of 10 mum resulted in an ICare reading deviation of 0.7 mm Hg. The Bland-Altman scatter-plot showed a reasonable agreement between the 2 tonometers. CONCLUSIONS: The ICare tonometer can be useful in a routine clinical setting. The IOP readings are quite in accordance with those obtained by GAT. The measurements seemed to be influenced by CCT variations, and thus pachymetry should always be taken into consideration.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
Martin Bland, J., & Altman, D. (1986) · The Lancet · doi:10.1016/s0140-6736(86)90837-8 · Cited by 47,965
(1965) · The Lancet · doi:10.1016/s0140-6736(65)91037-8 · Cited by 16,856
Kass, M. A. (2002) · Archives of Ophthalmology · doi:10.1001/archopht.120.6.701 · Cited by 3,705
Human Corneal Thickness and Its Impact on Intraocular Pressure Measures
Doughty, M. J., & Zaman, M. L. (2000) · Survey of Ophthalmology · doi:10.1016/s0039-6257(00)00110-7 · Cited by 1,362 · cited 1×
“*Correction formula of Doughty and Zaman.(9)”
APPLANATION TONOMETRY AND CENTRAL CORNEAL THICKNESS
EHLERS, N., BRAMSEN, T., & SPERLING, S. (1975) · Acta Ophthalmologica · doi:10.1111/j.1755-3768.1975.tb01135.x · Cited by 931
Sources of error with use of Goldmann-type tonometers
Whitacre, M. M., & Stein, R. (1993) · Survey of Ophthalmology · doi:10.1016/0039-6257(93)90053-a · Cited by 747 · cited 1×
“The accuracy of GAT measurements is dependent on many factors, which, with regard to the cornea, include thickness, curvature, structure, and axial length.(5) Central corneal thickness (CCT), especially, has been shown to have a substantial effect on IOP readings obtained by GAT.”
BackgroundGoldmann, H., & Schmidt, T. (1957) · Ophthalmologica · doi:10.1159/000303213 · Cited by 607
THE EPIDEMIOLOGY OF OPEN-ANGLE GLAUCOMA: A REVIEW
LESKE, M. C. (1983) · American Journal of Epidemiology · doi:10.1093/oxfordjournals.aje.a113626 · Cited by 535 · cited 1×
“G laucoma is a major cause of blindness in the Western countries.(1) It is well known that the major risk factors in the progression of glaucoma include increased intraocular pressure (IOP) levels and increased variation in IOP.”
BackgroundWOLFS, R. C. W., KLAVER, C. C. W., VINGERLING, J. R., GROBBEE, D. E., HOFMAN, A., & de JONG, P. T. V. M. (1997) · American Journal of Ophthalmology · doi:10.1016/s0002-9394(14)71125-0 · Cited by 415
Shimmyo, M., Ross, A. J., Moy, A., & Mostafavi, R. (2003) · American Journal of Ophthalmology · doi:10.1016/s0002-9394(03)00424-0 · Cited by 375
Cited by
Clinical Comparison of the Icare Tonometer and Goldmann Applanation Tonometry
Pakrou, N., Gray, T., Mills, R., Landers, J., & Craig, J. (2008) · Journal of Glaucoma · doi:10.1097/ijg.0b013e318133fb32 · Cited by 189
Surveillance for ocular hypertension: an evidence synthesis and economic evaluation.
Burr, J., Botello-Pinzon, P., Takwoingi, Y., Hernández, R., Vazquez-Montes, M., Elders, A., Asaoka, R., Banister, K., van der Schoot, J., Fraser, C., et al. (2012) · Health Technology Assessment · doi:10.3310/hta16290 · Cited by 180
Assessment of True Intraocular Pressure: The Gap Between Theory and Practical Data
Chihara, E. (2008) · Survey of Ophthalmology · doi:10.1016/j.survophthal.2008.02.005 · Cited by 180
The Influence of Corneal Properties on Rebound Tonometry
Chui, W.-S., Lam, A., Chen, D., & Chiu, R. (2008) · Ophthalmology · doi:10.1016/j.ophtha.2007.03.061 · Cited by 163
Metadata
- Authors
- Brusini, P., Salvetat, M. L., Zeppieri, M., Tosoni, C., & Parisi, L.
- Venue
- Journal of Glaucoma
- Published
- 2006-06
- Volume
- 15
- Issue
- 3
- Pages
- 213-217
- ISSN
- 1057-0829 (ISSN-L)
- OpenAlex
- W1975068023
- Cited by
- 187
- References
- 22
- FWCI
- 6.29
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Comparison of ICare Tonometer with Goldmann Applanation Tonometer in Glaucoma Patients. https://app.cochat.ai/papers/10.1097/01.ijg.0000212208.87523.66
- The paper
- Brusini, P., Salvetat, M. L., Zeppieri, M., Tosoni, C., & Parisi, L. (2006). Comparison of ICare Tonometer with Goldmann Applanation Tonometer in Glaucoma Patients. Journal of Glaucoma. https://doi.org/10.1097/01.ijg.0000212208.87523.66
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.