Clinical Comparison of the Icare Tonometer and Goldmann Applanation Tonometry
The DOI matches its Crossref record; 16 of the 16 references it cites were found in a registry or an index.
Abstract
PURPOSE: To compare a new method of intraocular pressure (IOP) measurement, using the Icare tonometer, with Goldmann applanation tonometry (GAT). PATIENTS AND METHODS: Two observers obtained IOP readings in 292 eyes (143 right and 149 left) of 153 subjects, using the Icare without topical anesthetic. A GAT reading was subsequently obtained by a consultant ophthalmologist, without the knowledge of the Icare readings. Central corneal thickness (CCT) was obtained on all eyes with ultrasound pachymetry. Patient comfort after IOP measurement was assessed in a consecutive subset of patients. RESULTS: The intraclass correlation coefficient between the 2 modalities of IOP measurement was r=0.95 for the right and r=0.93 for the left eye. The mean difference (Icare-GAT) between the IOP measured by the 2 methods was 0.4 mm Hg in the right eye (SD 3.0, 95% confidence interval -5.5 to 6.3), and 0.8 mm Hg in the left eye (SD 3.0, confidence interval -4.7 to 6.2). GAT measurements did not vary with CCT [correlation coefficient=0.09 (P=0.25) right and 0.14 (P=0.09) left eyes]. However, IOP measured with Icare tonometry increased with increasing CCT [correlation coefficient=0.16 (P=0.05) right and 0.21 (P=0.01) left eyes]. For every 100-microm increase in CCT, the difference (Icare-GAT) increased by 1 mm Hg. Of the 38 consecutive patients surveyed, 28 (73.7%) rated the Icare more comfortable than GAT, with only 2 (5.3%) rating it less comfortable (P<0.001). CONCLUSIONS: There is good correlation between the 2 methods of IOP measurement, even at extremes of IOP. The Icare instrument was easy to use and recorded rapid and consistent readings with minimal training. It seems to be more comfortable than GAT and obviates the need for topical anesthesia.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
Martin Bland, J., & Altman, D. (1986) · The Lancet · doi:10.1016/s0140-6736(86)90837-8 · Cited by 47,965
(1965) · The Lancet · doi:10.1016/s0140-6736(65)91037-8 · Cited by 16,856
A new induction‐based impact method for measuring intraocular pressure
Kontiola, A. I. (2000) · Acta Ophthalmologica Scandinavica · doi:10.1034/j.1600-0420.2000.078002142.x · Cited by 214 · cited 3×
“IOP is calculated from the measurement of impact duration and/or maximum deceleration.(2) Use of the device is simple and can be rapidly taught.”
BackgroundMethodReproducibility and Clinical Evaluation of Rebound Tonometry
Martinez-de-la-Casa, J. M., Garcia-Feijoo, J., Castillo, A., & Garcia-Sanchez, J. (2005) · Investigative Opthalmology & Visual Science · doi:10.1167/iovs.05-0586 · Cited by 201 · cited 1×
“Martinezde-la-Casa et al(17) found that in terms of pachymetry, the 2 tonometers behaved in a similar manner.(17) In our experience to obtain more accurate readings (ie, no error bars) or readings more consistent with the GAT, extra attention was needed to ensure the probe was held perpendicular to the cornea and the…”
BackgroundComparison of ICare Tonometer with Goldmann Applanation Tonometer in Glaucoma Patients
Brusini, P., Salvetat, M. L., Zeppieri, M., Tosoni, C., & Parisi, L. (2006) · Journal of Glaucoma · doi:10.1097/01.ijg.0000212208.87523.66 · Cited by 187
Comparison of the ICare® rebound tonometer with the Goldmann tonometer in a normal population
Fernandes, P., Díaz‐Rey, J. A., Queirós, A., Gonzalez‐Meijome, J. M., & Jorge, J. (2005) · Ophthalmic and Physiological Optics · doi:10.1111/j.1475-1313.2005.00327.x · Cited by 167 · cited 1×
“Another small study of subjects with normal IOPs concluded that the Icare overestimates the GAT, although with an absolute bias of <3mm Hg in approximately 80%.(16) A study of 103 subjects compared the Icare with the GAT and found no significant difference between the 2 modalities.”
BackgroundNakamura, M., Darhad, U., Tatsumi, Y., Fujioka, M., Nagai-Kusuhara, A., Maeda, H., & Negi, A. (2006) · American Journal of Ophthalmology · doi:10.1016/j.ajo.2006.02.035 · Cited by 155
Clinical evaluation of rebound tonometer
Davies, L. N., Bartlett, H. E., Mallen, E. A. H., & Wolffsohn, J. S. (2006) · Acta Ophthalmologica Scandinavica · doi:10.1111/j.1600-0420.2005.00610.x · Cited by 125 · cited 11×
“However, intersessional repeatability of IOP was poorer with the Icare compared with GAT.(18) Nakamura et al(19) compared IOP measurements using the Icare with GAT, Tonopen XL, and noncontact tonometer, and the influence of CCT in 45 (12 control and 33 glaucomatous or ocular hypertensive) subjects.”
MethodInfluentialStandardizing the Measurement of Intraocular Pressure for Clinical Research
Kass, M. A. (1996) · Ophthalmology · doi:10.1016/s0161-6420(96)30741-0 · Cited by 120
van der Jagt, L., & Jansonius, N. M. (2005) · Ophthalmic and Physiological Optics · doi:10.1111/j.1475-1313.2005.00318.x · Cited by 118 · cited 2×
“In comparing the Tonopen with GAT, some studies reported significant differences between the 2 devices, whereas others failed to show significant differences.(14,15) There are few studies comparing the Icare impact tonometer with the GAT.”
Result
Cited by
Surveillance for ocular hypertension: an evidence synthesis and economic evaluation.
Burr, J., Botello-Pinzon, P., Takwoingi, Y., Hernández, R., Vazquez-Montes, M., Elders, A., Asaoka, R., Banister, K., van der Schoot, J., Fraser, C., et al. (2012) · Health Technology Assessment · doi:10.3310/hta16290 · Cited by 180
Metadata
- Authors
- Pakrou, N., Gray, T., Mills, R., Landers, J., & Craig, J.
- Venue
- Journal of Glaucoma
- Published
- 2008-01
- Volume
- 17
- Issue
- 1
- Pages
- 43-47
- ISSN
- 1057-0829 (ISSN-L)
- OpenAlex
- W1968810861
- Cited by
- 189
- References
- 16
- FWCI
- 6.28
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Clinical Comparison of the Icare Tonometer and Goldmann Applanation Tonometry. https://app.cochat.ai/papers/10.1097/ijg.0b013e318133fb32
- The paper
- Pakrou, N., Gray, T., Mills, R., Landers, J., & Craig, J. (2008). Clinical Comparison of the Icare Tonometer and Goldmann Applanation Tonometry. Journal of Glaucoma. https://doi.org/10.1097/ijg.0b013e318133fb32
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.