European Society of Hypertension guidelines for blood pressure monitoring at home: a summary report of the Second International Consensus Conference on Home Blood Pressure Monitoring
Open accessAbstract
This document summarizes the available evidence and provides recommendations on the use of home blood pressure monitoring in clinical practice and in research. It updates the previous recommendations on the same topic issued in year 2000. The main topics addressed include the methodology of home blood pressure monitoring, its diagnostic and therapeutic thresholds, its clinical applications in hypertension, with specific reference to special populations, and its applications in research. The final section deals with the problems related to the implementation of these recommendations in clinical practice.
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
About 10% to 30% of subjects attending clinics due to high blood pressure have white coat hypertension.
“About 10%-30% of subjects attending clinics due to high BP have white coat hypertension”
From 2020 International Society of Hypertension Global Hypertension Practice Guidelines (2020) · cites this paper as [20]
The Diagnostic implication section reports 15–20% white-coat hypertension in the general population, a figure within the claimed range. It does not establish a 10–30% prevalence specifically among people attending clinics because of high blood pressure; the population and stated range differ. No page number was available in our copy. (LLM, conf=0.83)
Experimental — accuracy not yet measuredOut-of-office blood pressure measurements are more closely associated with hypertension-induced organ damage and the risk of cardiovascular events than office measurements.
“more closely associated with hypertension-induced organ damage and the risk of cardiovascular events”
From 2020 International Society of Hypertension Global Hypertension Practice Guidelines (2020) · cites this paper as [20]
The 'Predictive value for morbidity' section reports that home blood pressure predicted cardiovascular events better than office blood pressure in hypertensive patients and patients with chronic kidney disease, but also reports a population study finding no such superiority. The 'Other analyses' section says home measurements predict target-organ damage, while noting that some of those studies did not compare them with office measurements. The claim therefore overgeneralizes the evidence to out-of-office measurements and to both outcomes without qualification. (LLM, conf=0.85)
Experimental — accuracy not yet measuredAbout 10% to 15% of subjects attending clinics due to high blood pressure have masked hypertension.
“and 10%-15% have masked hypertension.”
From 2020 International Society of Hypertension Global Hypertension Practice Guidelines (2020) · cites this paper as [20]
The section ‘Diagnostic implication of home blood pressure monitoring in hypertension’ gives a 10–15% prevalence of masked hypertension in the general population. It does not establish that percentage specifically for people attending clinics because of high blood pressure; the same section reports a different estimate for some treated patients with controlled office blood pressure. No page number is available in our copy. (LLM, conf=0.88)
Experimental — accuracy not yet measuredOther blood pressure monitoring device components in addition to the electronic transducer, such as tubing, connections and cuffs, may deteriorate and may affect accuracy, emphasizing the need for regular maintenance.
“However, other BP monitoring device components in addition to the electronic transducer, such as tubing, connections and cuffs, may deteriorate (air leaks, etc.), and may affect accuracy, which emphasizes the need of regular maintenance.”
From Home blood pressure monitoring: methodology, clinical relevance and practical application: a 2021 position paper by the Working Group on Blood Pressure Monitoring and Cardiovascular Variability of the European Society of Hypertension (2021) · cites this paper as [4]
The Device calibration section says tubing and cuffs may deteriorate and affect accuracy, and recommends periodic checks and replacement when needed. It does not specifically mention connections or air leaks, so those details are not supported by our copy. (LLM, conf=0.84)
Experimental — accuracy not yet measuredUsing a larger cuff, or overcuffing, is common in children and leads to blood pressure underestimation.
“whereas using a larger cuff (overcuffing) is common in children and leads to BP underestimation [4,90,98,99].”
From Home blood pressure monitoring: methodology, clinical relevance and practical application: a 2021 position paper by the Working Group on Blood Pressure Monitoring and Cardiovascular Variability of the European Society of Hypertension (2021) · cites this paper as [4]
The “Cuff and bladder” section says that an oversized bladder may underestimate actual blood pressure and recommends different cuff sizes for children at different ages. It does not say that overcuffing is common in children. The supplied copy does not give a page number for this section. (LLM, conf=0.88)
Experimental — accuracy not yet measured
Using a small bladder, or undercuffing, is common in obese adults and leads to overestimation of blood pressure.
“Using a small bladder (undercuffing) is common in obese adults and leads to overestimation of BP”
From Home blood pressure monitoring: methodology, clinical relevance and practical application: a 2021 position paper by the Working Group on Blood Pressure Monitoring and Cardiovascular Variability of the European Society of Hypertension (2021) · cites this paper as [4]
The “Cuff and bladder” section says a too-small bladder may overestimate BP, and “Obese people” says a regular cuff may do so in obese individuals. The paper identifies inappropriate cuff size as an important measurement difficulty in this group, but does not establish that undercuffing is common or that it invariably leads to overestimation. (LLM, conf=0.84)
Experimental — accuracy not yet measuredOut-of-office blood pressure measurements are more reproducible than office measurements.
“Out-of-office BP measurements (by patients at home or with 24-hour ambulatory blood pressure monitoring [ABPM]) are more reproducible than office measurements,”
From 2020 International Society of Hypertension Global Hypertension Practice Guidelines (2020) · cites this paper as [20]
Experimental — accuracy not yet measuredSubjects with white coat hypertension are at intermediate cardiovascular risk between normotensives and sustained hypertensives.
“White coat hypertension: These subjects are at intermediate cardiovascular risk between normotensives and sustained hypertensives.”
From 2020 International Society of Hypertension Global Hypertension Practice Guidelines (2020) · cites this paper as [20]
The use of office and out-of-office blood pressure measurements identifies individuals with white coat hypertension and those with masked hypertension.
“The use of office and out-of-office (home or ambulatory) BP measurements identifies individuals with white coat hypertension, who have elevated BP only in the office (nonelevated ambulatory or home BP), and those with masked hypertension, who have nonelevated BP in the office but elevated BP out of the office (ambulatory or home). 1,2,17-21,25-27”
From 2020 International Society of Hypertension Global Hypertension Practice Guidelines (2020) · cites this paper as [20]
Experimental — accuracy not yet measuredOut-of-office blood pressure measurements identify the white coat and masked hypertension phenomena.
“and identify the white coat and masked hypertension phenomena (see below).”
From 2020 International Society of Hypertension Global Hypertension Practice Guidelines (2020) · cites this paper as [20]
Patients with masked hypertension are at similar risk of cardiovascular events as sustained hypertensives.
“Masked hypertension: These patients are at similar risk of cardiovascular events as sustained hypertensives.”
From 2020 International Society of Hypertension Global Hypertension Practice Guidelines (2020) · cites this paper as [20]
The clinical application and accuracy of HBPM may be limited by conditions related to the individual, the procedure, the oscillometric technique, or cuff-related issues.
“the clinical application and the accuracy of HBPM may be limited by certain conditions related either to the individual [63–66], to the procedure itself [4,67], to the oscillometric technique for BP measurement or to cuff-related issues [4,65,68–74].”
There is some inconsistency in the recommendations by scientific societies on the issue of miscuffing.
“However, there is some inconsistency in the recommendations by scientific societies on the issue of miscuffing [1,4,90,98–100].”
In the 2008 ESH HBPM recommendations, a threshold for hypertension diagnosis with systolic/diastolic HBP of 135/85 mmHg was proposed, based on a review of evidence including two meta-analyses.
“In the 2008 ESH HBPM recommendations, a threshold for hypertension diagnosis with systolic/diastolic HBP of 135/85 mmHg was proposed [4], based on the review of evidence including two meta-analyses [139,140]”
Current guidelines recommend HBPM as a method for the evaluation of BP in untreated individuals with suspected hypertension and for monitoring long-term BP control in treated hypertensive patients.
“Consequently, current guidelines recommend also HBPM as a method for the evaluation of BP in untreated individuals with suspected hypertension and, more so, for monitoring the long-term BP control in treated hypertensive patients [4,169]”
Compared with the last 2008-2010 Position Papers, the number of articles published in the field of HBPM in the last decade has been on considerable rise.
“Compared with the last 2008–2010 Position Papers [4], the number of articles published in the field of HBPM in the last decade has been on considerable rise”
A smartphone application from the European Society of Hypertension (ESH CARE app) has been validated and continuously updated.
“A smartphone application from the European Society of Hypertension (ESH CARE app) has been validated and continuously updated.”
From São Paulo call to action for the prevention and control of high blood pressure: 2020 (2019) · cites this paper as [68]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
The citation lists haven’t been fetched yet
References 285 · Cited by 837
Research works printed but not linked: 20
- [154]154. Faria MS, Nunes JP, Ferraz JM, Fernandes J, Praca A, Pestana M, et al. 24-h blood pressure profile early after renal transplantation. Rev Port Cardiol 1995; 14:227-231. 158
- [186]186. Verdecchia P, et al. Reliability of home self-recorded arterial pressure in essential hypertension in relation to the stage of the disease. In: Germano G, editor. Blood pressure recording in the clinical management of hypertension. Rome: Ediziono Pozzi; 1985. pp. 40-42.
- [187]187. Asmar R. Proceedings from the First International Consensus Conference on Self-Blood Pressure Measurement. Blood Press Monit 2000; 5:91-92.
- [188]188. O'Brien E, De Gaudemaris R, Bobrie G, Agabiti Rosei E, Vaisse B, and the participants of the First International Consensus Conference on Blood Pressure Self-Measurement. Devices and validation. Blood Press Monit 2000; 5:93-100.
- [192]192. Mancia G, Ferrari A, Gregorini L, Parati G, Pomidossi G, Bertinieri G, et al. Blood pressure and heart rate variabilities in normotensive and hypertensive human beings. Circ Res 1983; 53:96- 104.
- [196]196. Silverberg DS, Shemesh E, Jaina A. The unsup- ported arm: a cause of falsely raised blood pressure readings. BMJ 1977; 2:1331.
- [200]200. Patyna WD, Borsch B, Patyna M, Mitrovic V. Information deficits concerning blood pressure self- measurement. Dtsch Med Wochenschr 2004; 129:2466- 2469.
- [221]221. Blood Pressure Lowering Treatment Trialists' Collaboration. Blood pressure dependent and inde- pendent effects of agents that inhibit the renin- angiotensin system. J Hypertens 2007; 25:951-958.
- [224]224. Medical Research Council Working Party. MRC trial of treatment of mild hypertension: principal results. BMJ 1985; 291:97-104.
- [226]226. Parati G, Ulian G, Sampieri L, Palatini P, Villani A, Vanasia A, et al., on behalf of the Study on Ambulatory Monitoring of Blood Pressure and Lisinopril Evalu- ation (SAMPLE) Study Group. Attenuation of the 'white coat effect' by antihypertensive treatment and its relationship to regression of target organ damage. Hypertension 2000; 35:614-620.
- [227]227. O'Brien, E. Unmasking hypertension. Hypertension 2005; 45:481-482.
- [234]234. Lauer RM, Anderson AR, Beaglehole R, Burns TL. Factors related to tracking of blood pressure in children. Hypertension 1984; 6:307-314.
- [244]244. Kario K. Clinician's manual on early morning risk management in hypertension. London: Science Press; 2004. pp. 1-56.
- [253]253. Gupta M, Shennan A, Halligan A, Taylor D, de Swiet M. Oscillometric blood pressure monitoring in pregnancy and preeclampsia. Br J Obstetr Gynaecol 1997; 104:350-355.
- [260]260. Penny JA, Shennan AH, Halligan AW, Taylor DJ, de Swiet M, Anthony J. Blood pressure measure- ment in severe preeclampsia. Letter. Lancet 1997; 349:1518.
- [264]264. Shennan AH, Halligan A, Gupta M, Taylor D, de Swiet M. Oscillometric blood pressure measure- ment in severe preeclampsia: validation of the SpaceLabs 90207. Br J Obstet Gynaecol 1996b; 103: 171-173.
- [265]265. Quinn M. Automated blood pressure measurement devices: a potential source of morbidity and preeclampsia. Am J Obstet Gynecol 1994; 170:1303- 1307.
- [268]268. Jones MA, Kingswood JC, Dallyn PE, Andrew M, Cheetham A, Burwood R, Sharpstone P. Changes in diurnal blood pressure variation and red cell and plasma volumes in patients with renal failure who develop erythropoietin induced hypertension. Clin Nephrol 1995; 44:193-200.
- [281]281. Pickering TG. Self monitoring of blood pressure. In: Ambulatory monitoring and blood pressure varia- bility (part 1). London: Science Press; 1990. p. 8.5. (103
- [282]282. Longo D, Dorigatti F, Palatini P. Masked hyper- tension in adults. Blood Press Monit 2005; 10:307- 310.
- Authors
- Parati, G., Stergiou, G. S., Asmar, R., Bilo, G., de Leeuw, P., Imai, Y., Kario, K., Lurbe, E., Manolis, A., Mengden, T., O'Brien, E., Ohkubo, T., Padfield, P., Palatini, P., Pickering, T., Redon, J., Revera, M., Ruilope, L. M., Shennan, A., Staessen, J. A., Tisler, A., Waeber, B., Zanchetti, A., & Mancia, G.
- Affiliations
- IRCCS Istituto Auxologico Italiano (IT)
- Sotiria General Hospital (GR)
- Tohoku University (JP)
- Jichi Medical University (JP)
- Universitat de València (ES)
- Evangelismos Hospital (GR)
- University Hospital Bonn (DE)
- University College Dublin (IE)
- Western General Hospital (GB)
- University of Padua (IT)
- Columbia University Irving Medical Center (US)
- Research Institute Hospital 12 de Octubre (ES)
- St Thomas' Hospital (GB)
- Universitair Ziekenhuis Leuven (BE)
- Corresponding author
- Parati, Gianfranco
- ORCID
- Venue
- Journal of Hypertension
- Published
- 2008-08
- Volume
- 26
- Issue
- 8
- Pages
- 1505-1526
- ISSN
- 0263-6352 (ISSN-L)
- OpenAlex
- W2025309234
- Cited by
- 837
- References
- 285
- Citations by year
- 2026: 10 · 2025: 13 · 2024: 16 · 2023: 24 · 2022: 37 · 2021: 29 · 2020: 33 · 2019: 57 · 2018: 47 · 2017: 56 · 2016: 66 · 2015: 60 · 2014: 57 · 2013: 68 · 2012: 58
- FWCI
- 49.67
- Type
- journal article
- Language
- en
- Open access
- green
- hdl.handle.net
- Locations
- Journal of Hypertension (publishedVersion)
- PubMed (publishedVersion)
- BOA (University of Milano-Bicocca) (submittedVersion)
- Lirias (KU Leuven) (submittedVersion, other-oa)
- Institutional Repositories DataBase (IRDB) (submittedVersion, cc-by-nc-nd)
- IRIS (submittedVersion)
- Research Portal (King's College London) (submittedVersion)
- SERVAL (Université de Lausanne) (submittedVersion)
- Funders
- University of Connecticut
- University of Cambridge
- Cambridge University Hospitals
- Università degli Studi di Milano-Bicocca
- Universitair Ziekenhuis Gent
- Universiteit Gent
- Istituto Auxologico Italiano
- Ruijin Hospital
- Topics
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Blood Pressure and Hypertension Studies
- Health Sciences › Medicine › Public Health, Environmental and Occupational Health › Clinical practice guidelines implementation
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Keywords
- home blood pressure monitoring, hypertension, diagnostic thresholds, treatment thresholds, special populations, hypertension research, clinical guidelines
- MeSH terms
- Antihypertensive Agents, Europe, Humans, Hypertension, Drug Monitoring, Blood Pressure Monitoring, Ambulatory
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review