Comparison of the Accuracy of Noninvasive Hemoglobin Monitoring by Spectrophotometry (SpHb) and HemoCue® with Automated Laboratory Hemoglobin Measurement

Lamhaut, Lionel, Apriotesei, Roxana, Combes, Xavier, Lejay, Marc, Carli, Pierre, Vivien, Benoît · 2011 · Anesthesiology
Cited by 160

The DOI matches its Crossref record; 2 of the 2 references it cites were found in a registry or an index.

Abstract

Background: The reference method for hemoglobin concentration measurement remains automated analysis in the laboratory. Although point-of-care devices such as the HemoCue® 201+ (HemoCue, Ängelholm, Sweden) provide immediate hemoglobin values, a noninvasive, spectrophotometry-based technology (Radical-7®; Masimo Corp., Irvine, CA) that provides continuous online hemoglobin (SpHb) measurements has been introduced. This clinical study aimed to test the hypothesis that SpHb monitoring was equivalent to that of HemoCue® (the automated hemoglobin measurement in the laboratory taken as a reference method) during acute surgical hemorrhage. Methods: Blood for laboratory analysis was sampled after induction of anesthesia, during surgery according to the requirements of the anesthesiologist, and finally after the transfer of the patient to the recovery room. When each blood sample was taken, capillary samples were obtained for analysis with HemoCue®. SpHb monitoring was performed continuously during surgery. Using the automated hemoglobin measurement in the laboratory as a reference method, the authors tested the hypothesis that SpHb monitoring is equivalent to that of HemoCue®. The agreement between two methods was evaluated by linear regression and Bland and Altman analysis. Results: Eighty-five simultaneous measurements from SpHb, HemoCue®, and the laboratory were obtained from 44 patients. Bland and Altman comparison of SpHb and HemoCue® with the laboratory measurement showed, respectively, bias of -0.02 ± 1.39 g · dl(-1) and -0.17 ± 1.05 g · dl(-1), and a precision of 1.11 ± 0.83 g · dl(-1) and 0.67 ± 0.83 g · dl(-1). Considering an acceptable difference of ± 1.0 g · dl(-1) with the laboratory measurement, the percentage of outliers was significantly higher for SpHb than for HemoCue® (46% vs. 16%, P < 0.05). Conclusions: Taking automated laboratory hemoglobin measurement as a reference, the study shows that SpHb monitoring with Radical-7® gives lower readings than does the HemoCue® for assessment of hemoglobin concentration during hemorrhagic surgery.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1097/aln.0b013e3182270c22

Cites

2 of 2 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Show all 2 references

Cited by

See all 160

Metadata

Authors
Lamhaut, L., Apriotesei, R., Combes, X., Lejay, M., Carli, P., & Vivien, B.
Venue
Anesthesiology
Published
2011-09-01
Volume
115
Issue
3
Pages
548-554
ISSN
0003-3022 (ISSN-L)
OpenAlex
W2032116821
Cited by
160
References
2
FWCI
16.88
Type
journal article
Language
en
Open access
closed

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Comparison of the Accuracy of Noninvasive Hemoglobin Monitoring by Spectrophotometry (SpHb) and HemoCue® with Automated Laboratory Hemoglobin Measurement. https://app.cochat.ai/papers/10.1097/aln.0b013e3182270c22
The paper
Lamhaut, L., Apriotesei, R., Combes, X., Lejay, M., Carli, P., & Vivien, B. (2011). Comparison of the Accuracy of Noninvasive Hemoglobin Monitoring by Spectrophotometry (SpHb) and HemoCue® with Automated Laboratory Hemoglobin Measurement. Anesthesiology. https://doi.org/10.1097/aln.0b013e3182270c22

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.