Comparison of the Accuracy of Noninvasive Hemoglobin Monitoring by Spectrophotometry (SpHb) and HemoCue® with Automated Laboratory Hemoglobin Measurement
The DOI matches its Crossref record; 21 of the 21 references it cites were found in a registry or an index.
Abstract
Background: The reference method for hemoglobin concentration measurement remains automated analysis in the laboratory. Although point-of-care devices such as the HemoCue® 201+ (HemoCue, Ängelholm, Sweden) provide immediate hemoglobin values, a noninvasive, spectrophotometry-based technology (Radical-7®; Masimo Corp., Irvine, CA) that provides continuous online hemoglobin (SpHb) measurements has been introduced. This clinical study aimed to test the hypothesis that SpHb monitoring was equivalent to that of HemoCue® (the automated hemoglobin measurement in the laboratory taken as a reference method) during acute surgical hemorrhage. Methods: Blood for laboratory analysis was sampled after induction of anesthesia, during surgery according to the requirements of the anesthesiologist, and finally after the transfer of the patient to the recovery room. When each blood sample was taken, capillary samples were obtained for analysis with HemoCue®. SpHb monitoring was performed continuously during surgery. Using the automated hemoglobin measurement in the laboratory as a reference method, the authors tested the hypothesis that SpHb monitoring is equivalent to that of HemoCue®. The agreement between two methods was evaluated by linear regression and Bland and Altman analysis. Results: Eighty-five simultaneous measurements from SpHb, HemoCue®, and the laboratory were obtained from 44 patients. Bland and Altman comparison of SpHb and HemoCue® with the laboratory measurement showed, respectively, bias of -0.02 ± 1.39 g · dl(-1) and -0.17 ± 1.05 g · dl(-1), and a precision of 1.11 ± 0.83 g · dl(-1) and 0.67 ± 0.83 g · dl(-1). Considering an acceptable difference of ± 1.0 g · dl(-1) with the laboratory measurement, the percentage of outliers was significantly higher for SpHb than for HemoCue® (46% vs. 16%, P < 0.05). Conclusions: Taking automated laboratory hemoglobin measurement as a reference, the study shows that SpHb monitoring with Radical-7® gives lower readings than does the HemoCue® for assessment of hemoglobin concentration during hemorrhagic surgery.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
Martin Bland, J., & Altman, D. (1986) · The Lancet · doi:10.1016/s0140-6736(86)90837-8 · Cited by 47,965
(1965) · The Lancet · doi:10.1016/s0140-6736(65)91037-8 · Cited by 16,856
Agreement Between Methods of Measurement with Multiple Observations Per Individual
Bland, J. M., & Altman, D. G. (2007) · Journal of Biopharmaceutical Statistics · doi:10.1080/10543400701329422 · Cited by 2,065 · cited 1×
“Because our design included clustering of measurements within individuals over time performed in the same patient, we used appropriate correction for calculating the limits of agreement.(15) Lastly, we calculated the percentage of outliers for both SpHb monitoring and HemoCue, which were defined as difference values w…”
MethodStandardization of hemoglobinometry II. The hemiglobincyanide method
van Kampen, E. J., & Zijlstra, W. G. (1961) · Clinica Chimica Acta · doi:10.1016/0009-8981(61)90145-0 · Cited by 1,098 · cited 5×
“First, automated analysis performed in the hematology laboratory represents the reference method for the hemoglobin measurement, recommended for several decades by the International Committee for Standardization in Hematology.2,3 The major limitations of this method are the need for a blood sample, the time needed for…”
BackgroundMethodInfluentialEvaluation of "HemoCue," a new device for determining hemoglobin.
von Schenck, H., Falkensson, M., & Lundberg, B. (1986) · Clinical Chemistry · doi:10.1093/clinchem/32.3.526 · Cited by 258 · cited 1×
“89.(8) Recently, a new noninvasive, spectrophotometry-based monitoring technology has been introduced (Radical-7 Pulse CO-Oximeter; Masimo Corp.”
Macknet, M. R., Allard, M., Applegate, R. L., & Rook, J. (2010) · Anesthesia & Analgesia · doi:10.1213/ane.0b013e3181fc74b9 · Cited by 175
Tango, T. (1998) · Statistics in Medicine · doi:10.1002/(sici)1097-0258(19980430)17:8<891::aid-sim780>3.3.co;2-2 · Cited by 156
An azide-methemoglobin method for hemoglobin determination in blood.
Vanzetti, G. (1966) · PubMed · Cited by 141
A comparison between haematological parameters in 'capillary' and venous blood from healthy adults
Daae, L. N. W., Halvorsen, S., Mathisen, P., et al. (1988) · Scandinavian Journal of Clinical and Laboratory Investigation · doi:10.1080/00365518809085796 · Cited by 119
Lardi, A. M., Hirst, C., Mortimer, A. J., & McCollum, C. N. (1998) · Anaesthesia · doi:10.1046/j.1365-2044.1998.00328.x · Cited by 82 · cited 2×
“A small volume of blood (10 l) must be deposited into a single-use cuvette for analysis, after which the HemoCue device displays the hemoglobin concentration in less than 1 min.(6,7) The precision of this method is reported to be 1.”
BackgroundMethodA Clinical Evaluation of the Hemocue Haemoglobinometer using Capillary, Venous and Arterial Samples
Chen, P. P., Short, T. G., Leung, D. H. Y., & Oh, T. E. (1992) · Anaesthesia and Intensive Care · doi:10.1177/0310057x9202000419 · Cited by 73
Radtke, H., Polat, G., Kalus, U. J., Salama, A., & Kiesewetter, H. (2005) · Transfusion and Apheresis Science · doi:10.1016/j.transci.2004.11.004 · Cited by 70 · cited 2×
“, we chose an acceptable difference of less than 1 g dl 1 between SpHb or HemoCue versus the reference hemoglobin measurement in the laboratory.(17) The relevancy of this limit should be interpreted with caution because a 1 g dl 1 error obviously does not have similar consequences for a hemoglo-”
MethodPoint of care hematocrit and hemoglobin in cardiac surgery: a review
Myers, G. J., & Browne, J. (2007) · Perfusion · doi:10.1177/0267659107080826 · Cited by 68 · cited 2×
“the physician to obtain the result, and the blood spoliation induced by repeated measurements.(4) Second, the HemoCue is a portable point-of-care device, which displays the hemo-”
BackgroundMethodHemoglobin determination in blood donors
Cable, R. G. (1995) · Transfusion Medicine Reviews · doi:10.1016/s0887-7963(05)80052-5 · Cited by 56
Gehring, H., Dümbgen, L., Peterlein, M., Hagelberg, S., & DIBBELT, L. (2007) · Anesthesia & Analgesia · doi:10.1213/01.ane.0000268713.58174.cc · Cited by 56 · cited 2×
“in mind that both reference device and test device could be responsible for inherent errors.(18) Finally, the expected per-”
BackgroundDetermination of capillary hemoglobin levels using the HemoCue system in intensive care patients
Séguin, P., Kleiber, A., Chanavaz, C., Morçet, J., & Mallédant, Y. (2010) · Journal of Critical Care · doi:10.1016/j.jcrc.2010.08.012 · Cited by 45 · cited 2×
“In critically ill patients, Seguin et al.(27) recently showed a poor agreement between laboratory hemoglobin measurement and HemoCue, especially with capillary blood samples and to a much greater extent for patients with edema.”
BackgroundInfluentialReliability of HemoCue in patients with gastrointestinal bleeding
de Louw, A. V., Lasserre, N., Drouhin, F., Thierry, S., Lécuyer, L., Caen, D., & Tenaillon, A. (2006) · Intensive Care Medicine · doi:10.1007/s00134-006-0461-6 · Cited by 44 · cited 1×
“reported discrepancies of more than 1 g dl 1 in 21% of cases and more than 2 g dl 1 in 4% of cases.(26) In critically ill patients, Seguin et al.”
BackgroundSamama, C., Quezada, R. S., Riou, B., Mourgeon, E., Arock, M., Ankri, A., & VIARS, P. (1994) · Acta Anaesthesiologica Scandinavica · doi:10.1111/j.1399-6576.1994.tb03880.x · Cited by 34 · cited 1×
“Lastly, we calculated the percentage of outliers for both SpHb monitoring and HemoCue, which were defined as difference values with the reference method of the interval 1 g dl (1), as described previously.(16,17) This variable was considered as the main endpoint.”
MethodAuroy, Y., Lienhart, A., PÉQUIGNOT, F., & Benhamou, D. (2007) · Transfusion · doi:10.1111/j.1537-2995.2007.01383.x · Cited by 31
International Committee for Standardization in Haematology of the European Society of Haematology.
FISHER, J. W., Schofield, R., & Porteous, D. D. (1965) · British Journal of Haematology · doi:10.1111/j.1365-2141.1965.tb06600.x · Cited by 21
Iron Deficiency in Health and Disease
Reeves, J. D., Vichinsky, E. P., Addiego, J. E., & Lubin, B. H. (1983) · Advances in Pediatrics · doi:10.1016/s0065-3101(22)00582-5 · Cited by 19 · cited 1×
“5 g dl 1 between duplicate skin puncture samples, with 10% of paired values differing by 1 g dl (1).(28) To limit such variations, Mills and Meadows suggested that HemoCue cuvettes should be filled from bottle capillary samples, rather than a single blood drop taken directly from the finger.”
Background
Cited by
Smartphone app for non-invasive detection of anemia using only patient-sourced photos
Mannino, R. G., Myers, D. R., Tyburski, E. A., Caruso, C., Boudreaux, J., Leong, T., Clifford, G. D., & Lam, W. A. (2018) · Nature Communications · doi:10.1038/s41467-018-07262-2 · Cited by 272
Methods and analyzers for hemoglobin measurement in clinical laboratories and field settings
Whitehead, R. D., Mei, Z., Mapango, C., & Jefferds, M. E. D. (2019) · Annals of the New York Academy of Sciences · doi:10.1111/nyas.14124 · Cited by 188
New technologies in pediatric anesthesia
Holtby, H., Skowno, J. J., Kor, D. J., Flick, R. P., & Uezono, S. (2012) · Pediatric Anesthesia · doi:10.1111/pan.12007 · Cited by 35
Metadata
- Authors
- Lamhaut, L., Apriotesei, R., Combes, X., Lejay, M., Carli, P., & Vivien, B.
- Venue
- Anesthesiology
- Published
- 2011-09-01
- Volume
- 115
- Issue
- 3
- Pages
- 548-554
- ISSN
- 0003-3022 (ISSN-L)
- OpenAlex
- W2032116821
- Cited by
- 160
- References
- 21
- FWCI
- 16.88
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.