Surgical Navigation Technology Based on Augmented Reality and Integrated 3D Intraoperative Imaging: A Spine Cadaveric Feasibility and Accuracy Study
Open accessThe DOI matches its Crossref record; 11 of the 11 references it cites were found in a registry or an index.
Abstract
STUDY DESIGN: A cadaveric laboratory study. OBJECTIVE: The aim of this study was to assess the feasibility and accuracy of thoracic pedicle screw placement using augmented reality surgical navigation (ARSN). SUMMARY OF BACKGROUND DATA: Recent advances in spinal navigation have shown improved accuracy in lumbosacral pedicle screw placement but limited benefits in the thoracic spine. 3D intraoperative imaging and instrument navigation may allow improved accuracy in pedicle screw placement, without the use of x-ray fluoroscopy, and thus opens the route to image-guided minimally invasive therapy in the thoracic spine. METHODS: ARSN encompasses a surgical table, a motorized flat detector C-arm with intraoperative 2D/3D capabilities, integrated optical cameras for augmented reality navigation, and noninvasive patient motion tracking. Two neurosurgeons placed 94 pedicle screws in the thoracic spine of four cadavers using ARSN on one side of the spine (47 screws) and free-hand technique on the contralateral side. X-ray fluoroscopy was not used for either technique. Four independent reviewers assessed the postoperative scans, using the Gertzbein grading. Morphometric measurements of the pedicles axial and sagittal widths and angles, as well as the vertebrae axial and sagittal rotations were performed to identify risk factors for breaches. RESULTS: ARSN was feasible and superior to free-hand technique with respect to overall accuracy (85% vs. 64%, P < 0.05), specifically significant increases of perfectly placed screws (51% vs. 30%, P < 0.05) and reductions in breaches beyond 4 mm (2% vs. 25%, P < 0.05). All morphometric dimensions, except for vertebral body axial rotation, were risk factors for larger breaches when performed with the free-hand method. CONCLUSION: ARSN without fluoroscopy was feasible and demonstrated higher accuracy than free-hand technique for thoracic pedicle screw placement. LEVEL OF EVIDENCE: N/A.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Intraclass correlations: Uses in assessing rater reliability.
Shrout, P. E., & Fleiss, J. L. (1979) · Psychological Bulletin · doi:10.1037/0033-2909.86.2.420 · Cited by 24,532
Accuracy of Pedicular Screw Placement In Vivo
GERTZBEIN, S. D., & ROBBINS, S. E. (1990) · Spine · doi:10.1097/00007632-199001000-00004 · Cited by 1,381 · cited 1×
“Two independent neurosurgeons and two radiologists assessed screw positions using the Gertzbein grading: grade 0 (screw within the pedicle without cortical breach), grade 1 (0–2 mm breach, minor perforation including cortical encroachment), grade 2 (2–4 mm breach, moderate breach), and grade 3 (more than 4 mm breach,…”
MethodShin, B. J., James, A., Njoku, I. U., & Härtl, R. (2012) · Journal of Neurosurgery Spine · doi:10.3171/2012.5.spine11399 · Cited by 401
Tian, N.-F., Huang, Q.-S., Zhou, P., Zhou, Y., Wu, R.-K., Lou, Y., & Xu, H.-Z. (2011) · European Spine Journal · doi:10.1007/s00586-010-1577-5 · Cited by 345 · cited 1×
“Surgical navigation systems coupled with intraoperative volumetric imaging are used for various clinical applications in spine surgery, requiring pedicle screw placement.(1) Current systems typically consist of an infrared-camera tracking unit linked to an intraoperative computed tomography (CT) and a reference frame…”
MethodGautschi, O. P., Schatlo, B., Schaller, K., & Tessitore, E. (2011) · Neurosurgical Focus · doi:10.3171/2011.7.focus11168 · Cited by 268 · cited 3×
“In fact, Gautshi et al.(9) showed in a literature review that no complications (dural lesions, nerve root injuries, etc.”
BackgroundMethods to determine pedicle screw placement accuracy in spine surgery: a systematic review
Aoude, A. A., Fortin, M., Figueiredo, R., Jarzem, P., Ouellet, J., & Weber, M. H. (2015) · European Spine Journal · doi:10.1007/s00586-015-3853-x · Cited by 223 · cited 1×
“75).(22) Our study demonstrated an IA of 0.”
BackgroundSpinal Navigation and Imaging: History, Trends, and Future
Helm, P. A., Teichman, R. S., Hartmann, S. L., & Simon, D. (2015) · IEEE Transactions on Medical Imaging · doi:10.1109/tmi.2015.2391200 · Cited by 105 · cited 3×
“The ARSN system comprises navigation technology integrated in the flat detector frame, eliminating the need for stand-alone imaging and navigation equipment.(2) Several levels can be planned with the same degree of accuracy, without the need for invasive dynamic reference frame positioning and recalibrations.”
BackgroundMethodInfluentialRivkin, M. A., & Yocom, S. S. (2014) · Neurosurgical FOCUS · doi:10.3171/2014.1.focus13499 · Cited by 100
Learning curve of 3D fluoroscopy image–guided pedicle screw placement in the thoracolumbar spine
Ryang, Y., Villard, J., Obermüller, T., Friedrich, B., Wolf, P., Gempt, J., Ringel, F. A., & Meyer, B. (2014) · The Spine Journal · doi:10.1016/j.spinee.2014.10.003 · Cited by 93 · cited 3×
“Ryang et al.(24) concluded the same by looking at each quarter (average of 40 screws per quarter) for a year time span and found an increase in accuracy in thoracic spine starting from 55% after 40 screws up to 87.”
BackgroundMethodInfluentialLeković, G. P., Potts, E. A., Karahalios, D. G., & Hall, G. D. R. (2007) · Journal of Neurosurgery Spine · doi:10.3171/spi-07/10/393 · Cited by 82
Kraus, M. D., Weiskopf, J., Dreyhaupt, J., Krischak, G. D., & Gebhard, F. (2014) · Global Spine Journal · doi:10.1055/s-0034-1396430 · Cited by 30 · cited 1×
“For example, screw misplacement correlated with the increased distance from the treated vertebral body to the reference frame, infrared cameras obstructed the line of sight, calibration errors occurred, and inadvertently touching the reference frame caused registration error.(6,7,16)”
Background
Cited by
Virtual, augmented, and mixed reality applications in orthopedic surgery
Verhey, J. T., Haglin, J. M., Verhey, E. M., & Hartigan, D. E. (2020) · The International Journal of Medical Robotics and Computer Assisted Surgery · doi:10.1002/rcs.2067 · Cited by 264
Augmented and virtual reality in spine surgery, current applications and future potentials
Ghaednia, H., Fourman, M. S., Lans, A., Detels, K., Dijkstra, H., Lloyd, S., Sweeney, A., Oosterhoff, J. H. F., & Schwab, J. H. (2021) · The Spine Journal · doi:10.1016/j.spinee.2021.03.018 · Cited by 237
Elmi-Terander, A., Burström, G., Nachabe, R., Skulason, H., Pedersen, K., Fagerlund, M., Ståhl, F., Charalampidis, A., Söderman, M., Holmin, S., et al. (2019) · Spine · doi:10.1097/brs.0000000000002876 · Cited by 229
Self-citationAugmented Reality Technology Using Microsoft HoloLens in Anatomic Pathology
Hanna, M. G., Ahmed, I., Nine, J., Prajapati, S., & Pantanowitz, L. (2018) · Archives of Pathology & Laboratory Medicine · doi:10.5858/arpa.2017-0189-oa · Cited by 213
Uppot, R. N., Laguna, B., McCarthy, C. J., De Novi, G., Phelps, A., Siegel, E., & Courtier, J. (2019) · Radiology · doi:10.1148/radiol.2019182210 · Cited by 187
Metadata
- Authors
- Elmi-Terander, A., Skulason, H., Söderman, M., Racadio, J., Homan, R., Babic, D., van der Vaart, N., & Nachabe, R.
- Venue
- Spine
- Published
- 2016-11-01
- Volume
- 41
- Issue
- 21
- Pages
- E1303-E1311
- ISSN
- 0362-2436 (ISSN-L)
- 1528-1159
- OpenAlex
- W2513086572
- Cited by
- 160
- References
- 11
- FWCI
- 12.94
- Type
- journal article
- Language
- en
- Open access
- hybrid
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).