Clinical Acceptance and Accuracy Assessment of Spinal Implants Guided With SpineAssist Surgical Robot: Retrospective Study
Abstract
STUDY DESIGN: Retrospective, multicenter study of robotically-guided spinal implant insertions. Clinical acceptance of the implants was assessed by intraoperative radiograph, and when available, postoperative computed tomography (CT) scans were used to determine placement accuracy. OBJECTIVE: To verify the clinical acceptance and accuracy of robotically-guided spinal implants and compare to those of unguided free-hand procedures. SUMMARY OF BACKGROUND DATA: SpineAssist surgical robot has been used to guide implants and guide-wires to predefined locations in the spine. SpineAssist which, to the best of the authors' knowledge, is currently the sole robot providing surgical assistance in positioning tools in the spine, guided over 840 cases in 14 hospitals, between June 2005 and June 2009. METHODS: Clinical acceptance of 3271 pedicle screws and guide-wires inserted in 635 reported cases was assessed by intraoperative fluoroscopy, where placement accuracy of 646 pedicle screws inserted in 139 patients was measured using postoperative CT scans. RESULTS: Screw placements were found to be clinically acceptable in 98% of the cases when intraoperatively assessed by fluoroscopic images. Measurements derived from postoperative CT scans demonstrated that 98.3% of the screws fell within the safe zone, where 89.3% were completely within the pedicle and 9% breached the pedicle by up to 2 mm. The remaining 1.4% of the screws breached between 2 and 4 mm, while only 2 screws (0.3%) deviated by more than 4 mm from the pedicle wall. Neurologic deficits were observed in 4 cases yet, following revisions, no permanent nerve damage was encountered, in contrast to the 0.6% to 5% of neurologic damage reported in the literature. CONCLUSION: SpineAssist offers enhanced performance in spinal surgery when compared to free-hand surgeries, by increasing placement accuracy and reducing neurologic risks. In addition, 49% of the cases reported herein used a percutaneous approach, highlighting the contribution of SpineAssist in procedures without anatomic landmarks.
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cited by
Medical Robotics and Computer-Integrated Surgery
Taylor, R. H., Menciassi, A., Fichtinger, G., Fiorini, P., & Dario, P. (2016) · Springer Handbooks · doi:10.1007/978-3-319-32552-1_63 · Cited by 325
Medical Robots: Current Systems and Research Directions
Beasley, R. A. (2012) · Journal of Robotics · doi:10.1155/2012/401613 · Cited by 312
Hyun, S.-J., Kim, K.-J., Jahng, T.-A., & Kim, H.-J. (2017) · Spine · doi:10.1097/brs.0000000000001778 · Cited by 305
Gautschi, O. P., Schatlo, B., Schaller, K., & Tessitore, E. (2011) · Neurosurgical Focus · doi:10.3171/2011.7.focus11168 · Cited by 268
Metadata
- Authors
- Devito, D. P., Kaplan, L., Dietl, R., Pfeiffer, M., Horne, D., Silberstein, B., Hardenbrook, M., Kiriyanthan, G., Barzilay, Y., Bruskin, A., Sackerer, D., Alexandrovsky, V., Stüer, C., Burger, R., Maeurer, J., Gordon, D. G., Schoenmayr, R., Friedlander, A., Knoller, N., Schmieder, K., Pechlivanis, I., Kim, I.-S., Meyer, B., & Shoham, M.
- Venue
- Spine
- Published
- 2010-11
- Volume
- 35
- Issue
- 24
- Pages
- 2109-2115
- ISSN
- 0362-2436 (ISSN-L)
- OpenAlex
- W2088839360
- Cited by
- 407
- References
- 0
- FWCI
- 10.25
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Clinical Acceptance and Accuracy Assessment of Spinal Implants Guided With SpineAssist Surgical Robot: Retrospective Study. https://app.cochat.ai/papers/10.1097/brs.0b013e3181d323ab
- The paper
- Devito, D. P., Kaplan, L., Dietl, R., Pfeiffer, M., Horne, D., Silberstein, B., Hardenbrook, M., Kiriyanthan, G., Barzilay, Y., Bruskin, A., Sackerer, D., Alexandrovsky, V., Stüer, C., Burger, R., Maeurer, J., Gordon, D. G., Schoenmayr, R., Friedlander, A., Knoller, N., . . . Shoham, M. (2010). Clinical Acceptance and Accuracy Assessment of Spinal Implants Guided With SpineAssist Surgical Robot: Retrospective Study. Spine. https://doi.org/10.1097/brs.0b013e3181d323ab
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.