Validation of the Somnotouch-NIBP noninvasive continuous blood pressure monitor according to the European Society of Hypertension International Protocol revision 2010
Open accessAbstract
OBJECTIVE: The present study aimed to evaluate the accuracy of the Somnotouch-NIBP noninvasive continuous blood pressure monitor according to the European Society of Hypertension International Protocol revision 2010. MATERIALS AND METHODS: Systolic and diastolic blood pressures were sequentially measured in 33 adults (11 women, mean age 63.5±11.9 years) using a mercury sphygmomanometer (two observers) and the Somnotouch-NIBP device (one supervisor). A total of 99 pairs of comparisons were obtained from 33 participants for judgments in two parts with three grading phases. RESULTS: All the validation requirements were fulfilled. The Somnotouch-NIBP device fulfilled the requirements of the part 1 of the validation study. The number of absolute differences between device and observers within 5, 10, and 15 mmHg was 75/99, 90/99, and 96/99, respectively, for systolic blood pressure and 90/99, 99/99, and 99/99, respectively, for diastolic blood pressure. The device also fulfilled the criteria in part 2 of the validation study. Twenty-seven and 31 participants had at least two of the three device-observers differences less than or equal to 5 mmHg for systolic and diastolic blood pressure, respectively. All three device-observer differences were greater than 5 mmHg in two participants for systolic and in one participant for diastolic blood pressure. CONCLUSION: The Somnotouch-NIBP noninvasive continuous blood pressure monitor has passed the requirements of the International Protocol revision 2010, and hence can be recommended for blood pressure monitoring in adults, at least under conditions corresponding to those investigated in our study.
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Figures


Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The cited source says what the claim attributes to it.
In the citing paper
“advances in sensor technology for arterial pulse waveform and speed detection have paved the way for the potential development of devices for cuffless measurement of BP, in the perspective of continuous, beat-by-beat monitoring [103–106]”
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
O'Brien, E. T., Atkins, N., Stergiou, G. S., Karpettas, N., Parati, G., Asmar, R. G., Imai, Y., Wang, J., Mengden, T., & Shennan, A. H. (2010) · Blood Pressure Monitoring · doi:10.1097/mbp.0b013e3283360e98 · Cited by 609
Self-citationGesche, H., Grosskurth, D., Küchler, G., & Patzak, A. R. (2011) · European Journal of Applied Physiology · doi:10.1007/s00421-011-1983-3 · Cited by 472
Ambulatory Blood Pressure and Cardiovascular Outcome in Relation to Perceived Sleep Deprivation
Verdecchia, P., Angeli, F., Borgioni, C., Gattobigio, R., & Reboldi, G. (2007) · Hypertension · doi:10.1161/01.hyp.0000258215.26755.20 · Cited by 133
Ambulatory blood pressure monitoring does not interfere with the haemodynamic effects of sleep.
Parati, G., Pomidossi, G. A., Casadei, R., Malaspina, D., Colombo, A., Ravogli, A., & Mancia, G. (1985) · PubMed · Cited by 36
Self-citationSchmalgemeier, H., Bitter, T., Bartsch, S., Bullert, K., Fischbach, T. J., Eckert, S., Horstkotte, D., & Oldenburg, O. (2011) · Sleep And Breathing · doi:10.1007/s11325-011-0609-7 · Cited by 28
Cited by
Sorted by citations
Management of Hypertension in the Digital Era
Kario, K. (2020) · Hypertension · doi:10.1161/hypertensionaha.120.14742 · Cited by 247
Stergiou, G. S., Mukkamala, R., Avolio, A. P., Kyriakoulis, K. G., Mieke, S., Murray, A., Parati, G., Schutte, A. E., Sharman, J. E., Asmar, R. G., et al. (2022) · Journal of Hypertension · doi:10.1097/hjh.0000000000003224 · Cited by 231
Self-citationEvaluation of the Accuracy of Cuffless Blood Pressure Measurement Devices: Challenges and Proposals
Mukkamala, R., Yavarimanesh, M., Natarajan, K., Hahn, J.-O., Kyriakoulis, K. G., Avolio, A. P., & Stergiou, G. S. (2021) · Hypertension · doi:10.1161/hypertensionaha.121.17747 · Cited by 211
Parati, G., Stergiou, G. S., Bilo, G., Kollias, A., Pengo, M., Ochoa, J. E., Agarwal, R., Asayama, K., Asmar, R., Burnier, M., et al. (2021) · Journal of Hypertension · doi:10.1097/hjh.0000000000002922 · Cited by 178
Self-citationA survey on signals and systems in ambulatory blood pressure monitoring using pulse transit time
Buxi, D., Redouté, J., & Yuce, M. R. (2015) · Physiological Measurement · doi:10.1088/0967-3334/36/3/r1 · Cited by 140
Metadata
- Authors
- Bilo, G., Zorzi, C., Ochoa Munera, J. E., Torlasco, C., Giuli, V., & Parati, G.
- Affiliations
- University of Milan (IT)
- IRCCS Istituto Auxologico Italiano (IT)
- Corresponding author
- Parati, Gianfranco
- Venue
- Blood Pressure Monitoring
- Published
- 2015-10
- Volume
- 20
- Issue
- 5
- Pages
- 291-294
- ISSN
- 1359-5237 (ISSN-L)
- PMC
- PMC4568899
- OpenAlex
- W2096711200
- Cited by
- 143
- References
- 5
- Citations by year
- 2026: 10 · 2025: 8 · 2024: 11 · 2023: 12 · 2022: 20 · 2021: 16 · 2020: 27 · 2019: 14 · 2018: 10 · 2017: 6 · 2016: 7 · 2015: 2
- FWCI
- 6.07
- Type
- journal article
- Language
- en
- Open access
- hybrid
- doi.org
- Licence
- cc-by-nc-nd
- Locations
- Blood Pressure Monitoring (publishedVersion, cc-by-nc-nd)
- PubMed Central (publishedVersion, cc-by-nc-nd)
- PubMed (publishedVersion)
- BOA (University of Milano-Bicocca) (submittedVersion)
- Funders
- Istituto Auxologico Italiano
- Topics
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Blood Pressure and Hypertension Studies
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Health Sciences › Medicine › Surgery › Hemodynamic Monitoring and Therapy
- Keywords
- non-invasive blood pressure monitoring, device validation, systolic blood pressure, diastolic blood pressure, measurement accuracy, mercury sphygmomanometer comparison, adults
- MeSH terms
- Aged, Anthropometry, Arm, Clinical Protocols, Diastole, Electrocardiography, Equipment Design, Female, Heart Rate, Humans, Hypertension, Male, Middle Aged, Plethysmography, Systole, Observer Variation, Blood Pressure Monitors, Blood Pressure Monitoring, Ambulatory, Disease Management, Pulse Wave Analysis
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Validation of the Somnotouch-NIBP noninvasive continuous blood pressure monitor according to the European Society of Hypertension International Protocol revision 2010. https://app.cochat.ai/papers/10.1097/mbp.0000000000000124
- The paper
- Bilo, G., Zorzi, C., Ochoa Munera, J. E., Torlasco, C., Giuli, V., & Parati, G. (2015). Validation of the Somnotouch-NIBP noninvasive continuous blood pressure monitor according to the European Society of Hypertension International Protocol revision 2010. Blood Pressure Monitoring. https://doi.org/10.1097/mbp.0000000000000124