Evaluation of the Accuracy of Cuffless Blood Pressure Measurement Devices: Challenges and Proposals
Open accessAbstract
Several novel cuffless wearable devices and smartphone applications claiming that they can measure blood pressure (BP) are appearing on the market. These technologies are very attractive and promising, with increasing interest among health care professionals for their potential use. Moreover, they are becoming popular among patients with hypertension and healthy people. However, at the present time, there are serious issues about BP measurement accuracy of cuffless devices and the 2021 European Society of Hypertension Guidelines on BP measurement do not recommend them for clinical use. Cuffless devices have special validation issues, which have been recently recognized. It is important to note that the 2018 Universal Standard for the validation of automated BP measurement devices developed by the American Association for the Advancement of Medical Instrumentation, the European Society of Hypertension, and the International Organization for Standardization is inappropriate for the validation of cuffless devices. Unfortunately, there is an increasing number of publications presenting data on the accuracy of novel cuffless BP measurement devices, with inadequate methodology and potentially misleading conclusions. The objective of this review is to facilitate understanding of the capabilities and limitations of emerging cuffless BP measurement devices. First, the potential and the types of these devices are described. Then, the unique challenges in evaluating the BP measurement accuracy of cuffless devices are explained. Studies from the literature and computer simulations are employed to illustrate these challenges. Finally, proposals are given on how to evaluate cuffless devices including presenting and interpreting relevant study results.
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Cuffless devices have accuracy problems not seen in automated cuff devices, including the need for individual calibration and questionable accuracy in tracking BP changes after calibration.
“However, cuffless devices have specific accuracy issues that are not present in the automated cuff BP devices, such as the need for individual user calibration and their questionable accuracy in tracking BP changes after calibration55,56.”
From Blood pressure and its variability: classic and novel measurement techniques (2022) · cites this paper as [55]
The Types of Cuffless BP Measurement Devices section says cuff-calibrated devices require calibration for an individual, but calibration-free devices do not. The Critical Review section says the accuracy of cuff-calibrated devices in tracking BP changes within an individual often remains unclear. Thus the tracking concern is supported, while the need for individual calibration does not apply to cuffless devices generally; page numbers were not provided. (LLM, conf=0.85)
Experimental — accuracy not yet measuredCuffless devices can provide detailed information about true BP profile and behaviour over time and about all types of BP variability.
“Therefore, these devices can provide detailed information on true BP profile and behaviour over time, as well as all types of BPV (including short-term to long-term BPV6,55–57), and have great potential for use in screening, early diagnosis, close monitoring and optimal management of hypertension.”
From Blood pressure and its variability: classic and novel measurement techniques (2022) · cites this paper as [55]
The Potential of Cuffless Blood Pressure Measurement Devices section says these devices could provide a complete evaluation of BP level and behavior across daily circumstances and long periods. But the Critical Review of Data Presentation in Recent Studies cautions that their accuracy in tracking short- or long-term BP changes within an individual often remains unclear. The source therefore supports the potential for detailed monitoring, not the unqualified claim that cuffless devices can characterize all types of BP variability. (LLM, conf=0.85)
Experimental — accuracy not yet measuredCuffless BP technologies in wearables and smartphones can take multiple or continuous measurements during routine activities over days, weeks or months without cuff-induced limb compression.
“Cuffless BP technologies embedded in wearable devices and smartphones can obtain multiple or even continuous measurements during routine activities for days, weeks or months, without the discomfort of cuff-induced limb compression.”
From Blood pressure and its variability: classic and novel measurement techniques (2022) · cites this paper as [55]
“The Potential of Cuffless Blood Pressure Measurement Devices” (page unavailable) describes numerous measurements from wearables and smartphones across daily circumstances and long periods. It does not establish continuous measurement for months or discuss discomfort from limb compression; “Types of Cuffless BP Measurement Devices” (page unavailable) also notes that some devices require periodic arm-cuff calibration. (LLM, conf=0.83)
Experimental — accuracy not yet measuredMost cuffless BP devices require initial calibration for each user and periodic recalibration over time.
“Most of these devices require initial calibration in each user, which is usually performed by inputting a BP value measured using an upper-arm-cuff device or by inputting simple demographic data that are known to correlate with BP level (such as age and sex), with periodic recalibration required over time (every few hours, days or weeks).”
From Blood pressure and its variability: classic and novel measurement techniques (2022) · cites this paper as [55]
The “Types of Cuffless BP Measurement Devices” section says cuff-calibrated devices require periodic arm-cuff measurements, with intervals ranging from hours to weeks. It also identifies calibration-free devices that do not require calibration for each user, and does not establish that most devices require it. The “Simulation study” section treats demographic input as a feature of some calibration-free devices, not as periodic recalibration. (LLM, conf=0.88)
Experimental — accuracy not yet measuredCuffless BP devices use various parameters and techniques and have different intended uses.
“These devices use several different parameters and techniques to assess BP levels, including pulse transit time, pulse wave analysis, photoplethysmography and applanation tonometry, and have different intended uses (such as wearable devices for occasional use or for short-term continuous monitoring in an intensive care unit)55,56.”
From Blood pressure and its variability: classic and novel measurement techniques (2022) · cites this paper as [55]
Standards have been developed for the validation of wearable, cuffless BP devices, although further work is required to refine validation standards to account for issues presented by PPG-based devices.
“Standards have been developed for the validation of wearable, cuffless BP devices [168], [169] although further work is required to refine validation standards to account for the issues presented by PPG-based devices [170].”
From Wearable Photoplethysmography for Cardiovascular Monitoring (2022) · cites this paper as [170]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Cites
Forouzanfar, M. H., Afshin, A., Alexander, L. T., Anderson, H. R., Bhutta, Z. A., Biryukov, S., Brauer, M., Burnett, R., Cercy, K., Charlson, F. J., et al. (2016) · The Lancet · doi:10.1016/s0140-6736(16)31679-8 · Cited by 7,698 · see citing works
Dhamu Kp Ramamoorthy, K. (2016) · Cited by 2,952 · see citing works
Toward ubiquitous blood pressure monitoring via pulse transit time: Theory and practice
Mukkamala, R., Hahn, J.-O., Inan, O. T., Mestha, L. K., Kim, C.-S., Toreyin, H., & Kyal, S. (2015) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2015.2441951 · Cited by 935 · see citing works
Stergiou, G. S., Palatini, P., Parati, G., O’Brien, E., Januszewicz, A., Lurbe, E., Persu, A., Mancia, G., Kreutz, R., & on behalf of the European Society of Hypertension Council and the European Society of Hypertension Working Group on Blood Pressure Monitoring and Cardiovascular Variability (2021) · Journal of Hypertension · doi:10.1097/hjh.0000000000002843 · Cited by 859 · see citing works
Self-citationCuffless Blood Pressure Estimation Algorithms for Continuous Health-Care Monitoring
Kachuee, M., Kiani, M. M., Mohammadzade, H., & Shabany, M. (2017) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2016.2580904 · Cited by 621 · see citing works
Postexercise hypotension. Key features, mechanisms, and clinical significance.
Kenney, M., & Seals, D. R. (1993) · Hypertension · doi:10.1161/01.hyp.22.5.653 · Cited by 505 · see citing works
Cuff-less and Noninvasive Measurements of Arterial Blood Pressure by Pulse Transit Time
Poon, C. C. Y., & Zhang, Y. (2005) · doi:10.1109/iembs.2005.1615827 · Cited by 376 · see citing works
Stergiou, G. S., Alpert, B., Mieke, S., Asmar, R., Atkins, N., Eckert, S., Frick, G., Friedman, B., Graßl, T., Ichikawa, T., et al. (2018) · Journal of Hypertension · doi:10.1097/hjh.0000000000001634 · Cited by 278 · see citing works
Self-citationChowdhury, M. H., Shuzan, M. N. I., Chowdhury, M. E. H., Mahbub, Z. B., Uddin, M. M., Khandakar, A., & Reaz, M. B. I. (2020) · Sensors · doi:10.3390/s20113127 · Cited by 264 · see citing works
Smartphone-Based Blood Pressure Measurement Using Transdermal Optical Imaging Technology
Hong, L., Yang, D., Barszczyk, A., Vempala, N. N., Wei, J., Wu, S. J., Zheng, P. P., Fu, G., Lee, K., & Feng, Z. (2019) · Circulation Cardiovascular Imaging · doi:10.1161/circimaging.119.008857 · Cited by 242 · see citing works
Bilo, G., Zorzi, C., Ochoa Munera, J. E., Torlasco, C., Giuli, V., & Parati, G. (2015) · Blood Pressure Monitoring · doi:10.1097/mbp.0000000000000124 · Cited by 143 · see citing works
Cuffless Single-Site Photoplethysmography for Blood Pressure Monitoring
Hosanee, M., Chan, G., Welykholowa, K., Cooper, R., Kyriacou, P. A., Zheng, D., Allen, J., Abbott, D., Menon, C., Lovell, N. H., et al. (2020) · Journal of Clinical Medicine · doi:10.3390/jcm9030723 · Cited by 136 · see citing works
Multi-Sensor Fusion Approach for Cuff-Less Blood Pressure Measurement
Miao, F., Liu, Z.-D., Liu, J., Wen, B., He, Q., & Li, Y. (2019) · IEEE Journal of Biomedical and Health Informatics · doi:10.1109/jbhi.2019.2901724 · Cited by 136 · see citing works
Picone, D. S., Deshpande, R. A., Schultz, M. G., Fonseca, R., Campbell, N. R. C., Delles, C., Hecht Olsen, M., Schutte, A. E., Stergiou, G., Padwal, R., et al. (2020) · Hypertension · doi:10.1161/hypertensionaha.120.14719 · Cited by 118 · see citing works
Self-citationRuiz‐Rodríguez, J. C., Ruiz‐Sanmartín, A., Ribas, V. J., Caballero, J., García‐Roche, A., Riera, J., Nuvials, X., NADAL, M. D., Solà-Morales, O., Serra, J., et al. (2013) · Intensive Care Medicine · doi:10.1007/s00134-013-2964-2 · Cited by 117 · see citing works
Nachman, D., Gepner, Y., Goldstein, N., Kabakov, E., Ishay, A. B., Littman, R., Azmon, Y., Jaffe, E., & Eisenkraft, A. (2020) · Scientific Reports · doi:10.1038/s41598-020-73172-3 · Cited by 103 · see citing works
Schoettker, P., Degott, J., Hofmann, G., Proença, M., Bonnier, G., Lemkaddem, A., Lemay, M., Schorer, R., Christen, U., Knebel, J., et al. (2020) · Scientific Reports · doi:10.1038/s41598-020-74955-4 · Cited by 101 · see citing works
Recent Research and Developing Trends of Wearable Sensors for Detecting Blood Pressure
Arakawa, T. (2018) · Sensors · doi:10.3390/s18092772 · Cited by 100 · see citing works
Xing, X., Ma, Z., Zhang, M., Zhou, Y., Dong, W., & Song, M. (2019) · Scientific Reports · doi:10.1038/s41598-019-45175-2 · Cited by 93 · see citing works
Validation protocols for blood pressure measuring devices in the 21st century
Stergiou, G. S., Alpert, B. S., Mieke, S., Wang, J., & O'Brien, E. (2018) · The Journal of Clinical Hypertension · doi:10.1111/jch.13294 · Cited by 92 · see citing works
Self-citationVybornova, A., Polychronopoulou, E., Wurzner-Ghajarzadeh, A., Fallet, S., Solà, J., & Wuerzner, G. (2021) · Blood Pressure Monitoring · doi:10.1097/mbp.0000000000000531 · Cited by 87 · see citing works
Development and Validation of a Novel Cuff-Less Blood Pressure Monitoring Device
Watanabe, N., Bando, Y., Kawachi, T., Yamakita, H., Futatsuyama, K., Honda, Y., Yasui, H., Nishimura, K., Kamihara, T., Okumura, T., et al. (2017) · JACC Basic to Translational Science · doi:10.1016/j.jacbts.2017.07.015 · Cited by 87 · see citing works
Mukkamala, R., & Hahn, J. (2017) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2017.2756018 · Cited by 86 · see citing works
Self-citationCuff-Less Methods for Blood Pressure Telemonitoring
Bard, D. M., Joseph, J. I., & van Helmond, N. (2019) · Frontiers in Cardiovascular Medicine · doi:10.3389/fcvm.2019.00040 · Cited by 84 · see citing works
Natarajan, K., Block, R., Yavarimanesh, M., Chandrasekhar, A., Mestha, L. K., Inan, O. T., Hahn, J., & Mukkamala, R. (2021) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2021.3087105 · Cited by 67 · see citing works
Self-citationon behalf of the Korean Society of Hypertension, Lee, H., Lee, D., Seo, J.-M., Ihm, S., Kim, K., Cho, E. J., Kim, H. C., Shin, J., Park, S., et al. (2021) · Clinical Hypertension · doi:10.1186/s40885-020-00158-8 · Cited by 53 · see citing works
Boubouchairopoulou, N., Κόλλιας, Α., Chiu, B., Chen, B., Lagou, S., Anestis, P., & Stergiou, G. S. (2017) · Journal of Human Hypertension · doi:10.1038/jhh.2016.101 · Cited by 43 · see citing works
Self-citationLiu, Z.-D., Zhou, B., Li, Y., Tang, M., & Miao, F. (2020) · Frontiers in Physiology · doi:10.3389/fphys.2020.575407 · Cited by 39 · see citing works
Initialization of Pulse Transit Time-Based Blood Pressure Monitors
Mukkamala, R., & Hahn, J. (2019) · Springer eBooks · doi:10.1007/978-3-030-24701-0_10 · Cited by 8 · see citing works
Self-citation
Flagged first, then by citations
Research works printed but not linked: 1
- [27]IEEE Standard Association, “ IEEE 1708a-2019 - IEEE Standard for Wearable, Cuffless Blood Pressure Measuring Devices - Amendment 1,” IEEE Std, 2019. https://standards.ieee.org/standard/1708a-2019.html. Accessed May 24, 2021.
Cited by
Sorted by citations
Mancia, G., Kreutz, R., Brunström, M., Burnier, M., Grassi, G., Januszewicz, A., Muiesan, M. L., Tsioufis, K., Agabiti-Rosei, E., Algharably, E. A. E., et al. (2023) · Journal of Hypertension · doi:10.1097/hjh.0000000000003480 · Cited by 3,473 · see citing works
Self-citationBlood pressure and its variability: classic and novel measurement techniques
Schutte, A. E., Kollias, A., & Stergiou, G. S. (2022) · Nature Reviews Cardiology · doi:10.1038/s41569-022-00690-0 · Cited by 335 · see citing works
Self-citationWearable Photoplethysmography for Cardiovascular Monitoring
Charlton, P. H., Kyriacou, P. A., Mant, J., Marozas, V., Chowienczyk, P., & Alastruey, J. (2022) · Proceedings of the IEEE · doi:10.1109/jproc.2022.3149785 · Cited by 250 · see citing works
Cuffless Blood Pressure Measurement
Mukkamala, R., Stergiou, G. S., & Avolio, A. P. (2022) · Annual Review of Biomedical Engineering · doi:10.1146/annurev-bioeng-110220-014644 · Cited by 180 · see citing works
Self-citation
- Authors
- Mukkamala, R., Yavarimanesh, M., Natarajan, K., Hahn, J.-O., Kyriakoulis, K. G., Avolio, A. P., & Stergiou, G. S.
- Affiliations
- University of Pittsburgh (US)
- Michigan State University (US)
- University of Maryland, College Park (US)
- National and Kapodistrian University of Athens (GR)
- Macquarie University (AU)
- ORCID
- Venue
- Hypertension
- Published
- 2021-11
- Volume
- 78
- Issue
- 5
- Pages
- 1161-1167
- ISSN
- 0194-911X (ISSN-L)
- 1524-4563
- PMC
- PMC8516718
- OpenAlex
- W3200345925
- Cited by
- 211
- References
- 29
- Citations by year
- 2026: 37 · 2025: 43 · 2024: 38 · 2023: 50 · 2022: 42 · 2021: 1
- FWCI
- 2.76
- Type
- journal article
- Language
- en
- Open access
- bronze
- doi.org
- Locations
- Hypertension (publishedVersion)
- PubMed (publishedVersion)
- PubMed Central (acceptedVersion)
- Funders
- National Heart, Lung, and Blood Institute (R01HL146470)
- National Institute of Biomedical Imaging and Bioengineering (U01EB018818)
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Blood Pressure and Hypertension Studies
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Cardiovascular Health and Disease Prevention
- Keywords
- cuffless blood pressure measurement, measurement accuracy
- MeSH terms
- Wearable Electronic Devices, Blood Pressure, Blood Pressure Determination, Humans, Hypertension, Self Care, Sensitivity and Specificity, Signal Processing, Computer-Assisted, Reproducibility of Results, Pulse Wave Analysis
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review