Optimum frequency of office blood pressure measurement using an automated sphygmomanometer

Myers, Martin G., Valdivieso, Miguel, Kiss, Alexander · 2008 · Blood Pressure Monitoring
Cited by 57

The DOI matches its Crossref record.

Abstract

OBJECTIVE: To determine the optimum interval between serial blood pressure measurements using an automated BpTRU sphygmomanometer. METHODS: Two groups of 200 patients each had automated office measurements taken using the BpTRU device at either 1-min or 2-min intervals from the start of one reading to the start of the next reading with a 24-h ambulatory blood pressure (ABP) recording being performed. Another series of 50 patients had BpTRU readings taken at 1-min and 2-min intervals before and after 24-h ABP monitoring. The difference between the mean awake ABP and the mean automated office BP readings were compared for recordings taken at 1-min versus 2-min intervals. RESULTS: In the between-patient comparison (n=400), mean awake ABP was similar to automated BP recordings in the examining room at either 1-min or 2-min intervals except for a slightly lower (-4 mmHg) diastolic BP with the 1-min interval (P<0.01 vs. ABP). In the within-patient comparison (n=50), there was no consistent difference between automated BP readings taken in the examining room at 1-min versus 2-min intervals. Overall, the mean automated BP values tended to be slightly lower than the mean awake ABP. CONCLUSION: Automated measurement of BP in the office setting with devices such as the BpTRU can be taken as frequently as every 1 min without affecting the accuracy of the reading. Small differences in BP between the 1 and 2-min settings and between the automated BpTRU and ABP readings were within accepted clinical standards for validation criteria.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1097/mbp.0b013e3283104247

Cites

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cited by

See all 57

Metadata

Authors
Myers, M. G., Valdivieso, M., & Kiss, A.
Venue
Blood Pressure Monitoring
Published
2008-12
Volume
13
Issue
6
Pages
333-338
ISSN
1359-5237 (ISSN-L)
OpenAlex
W2063504216
Cited by
57
References
0
FWCI
2.86
Type
journal article
Language
en
Open access
closed

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Optimum frequency of office blood pressure measurement using an automated sphygmomanometer. https://app.cochat.ai/papers/10.1097/mbp.0b013e3283104247
The paper
Myers, M. G., Valdivieso, M., & Kiss, A. (2008). Optimum frequency of office blood pressure measurement using an automated sphygmomanometer. Blood Pressure Monitoring. https://doi.org/10.1097/mbp.0b013e3283104247

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.