Bone Mineral Density Estimations From Routine Multidetector Computed Tomography: A Comparative Study of Contrast and Calibration Effects
Open accessAbstract
INTRODUCTION: Phantom-based (synchronous and asynchronous) and phantomless (internal tissue calibration based) assessment of bone mineral density (BMD) in routine MDCT (multidetector computed tomography) examinations potentially allows for diagnosis of osteoporosis. Although recent studies investigated the effects of contrast-medium application on phantom-calibrated BMD measurements, it remains uncertain to what extent internal tissue-calibrated BMD measurements are also susceptible to contrast-medium associated density variation. The present study is the first to systemically evaluate BMD variations related to contrast application comparing different calibration techniques. PURPOSE: To compare predicative performance of different calibration techniques for BMD measurements obtained from triphasic contrast-enhanced MDCT. MATERIALS AND METHODS: Bone mineral density was measured on nonenhanced (NE), arterial (AR) and portal-venous (PV) contrast phase MDCT images of 46 patients using synchronous (SYNC) and asynchronous (ASYNC) phantom calibration as well as internal calibration (IC). Quantitative computed tomography (QCT) served as criterion standard. Density variations were analyzed for each contrast phase and calibration technique, and respective linear fitting was performed. RESULTS: Both asynchronous calibration-derived BMD values (NE-ASYNC) and values estimated using IC (NE-IC) on NE MDCT images did reasonably well in predicting QCT BMD (root-mean-square deviation, 8.0% and 7.8%, respectively). Average NE-IC BMD was 2.7% lower when compared with QCT (P = 0.017), whereas no difference could be found for NE-ASYNC (P = 0.957). All average BMD estimates derived from contrast-enhanced scans differed significantly from QCT BMD (all P < 0.005) and led to notable systemic BMD biases (mean difference at least > 6.0 mg/mL). All regression fits revealed a consistent linear dependency (R range, 0.861-0.963). Overall accuracy and goodness of fit tended to decrease from AR to PV contrast phase. Highest precision and best linear fit could be reached using a synchronously scanned phantom (root-mean-square deviation, 9.4% for AR and 14.4% for PV). Both ASYNC and IC estimations performed comparably accurate and precise. CONCLUSIONS: Our data suggest that internal calibration driven BMD measurements derived from contrast-enhanced MDCT need the same amount of post hoc contrast-effect adjustment as measurements using phantom calibration. Adjustment using linear correction equations can correct for systematic bias of bone density variations related to contrast application, irrespective of the calibration technique used.
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Bias of BMD values due to intravenous injection of contrast medium was corrected for using linear correction equations for arterial and portal/venous contrast phases.
“Bias of BMD values due to intravenous injection of contrast medium was corrected for using linear correction equations for arterial and portal/venous contrast phases [23].”
From Automatic opportunistic osteoporosis screening in routine CT: improved prediction of patients with prevalent vertebral fractures compared to DXA (2021) · cites this paper as [23]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Cites
No references could be resolved yet.
Research works printed but not linked: 31
- [1]Nih Consensus Development Panel on Osteoporosis Prevention D, Therapy. Osteoporosis prevention, diagnosis, and therapy. JAMA. 2001; 285: 785– 795. 11176917
- [2]Boonen S Autier P Barette M Functional outcome and quality of life following hip fracture in elderly women: a prospective controlled study. Osteoporos Int. 2004; 15: 87– 94. 14605799
- [3]Papaioannou A Kennedy CC Ioannidis G The impact of incident fractures on health-related quality of life: 5 years of data from the Canadian Multicentre Osteoporosis Study. Osteoporos Int. 2009; 20: 703– 714. 18802659 PMC5101052
- [4]Ioannidis G Papaioannou A Hopman WM Relation between fractures and mortality: results from the Canadian Multicentre Osteoporosis Study. CMAJ. 2009; 181: 265– 271. 19654194 PMC2734204
- [5]Adams JE Quantitative computed tomography. Eur J Radiol. 2009; 71: 415– 424. 19682815
- [6]Engelke K Libanati C Liu Y Quantitative computed tomography (QCT) of the forearm using general purpose spiral whole-body CT scanners: accuracy, precision and comparison with dual-energy X-ray absorptiometry (DXA). Bone. 2009; 45: 110– 118. 19345291
- [7]Mao SS Li D Luo Y Application of quantitative computed tomography for assessment of trabecular bone mineral density, microarchitecture and mechanical property. Clin Imaging. 2015; 40: 330– 338. 26602163
- [8]Bolotin HH Sievanen H Grashuis JL Inaccuracies inherent in patient-specific dual-energy X-ray absorptiometry bone mineral density measurements: comprehensive phantom-based evaluation. J Bone Miner Res. 2001; 16: 417– 426. 11204442
- [9]Antonacci MD Hanson DS Heggeness MH Pitfalls in the measurement of bone mineral density by dual energy x-ray absorptiometry. Spine (Phila Pa 1976). 1996; 21: 87– 91. 9122768
- [10]Assessment of fracture risk and its application to screening for postmenopausal osteoporosis. Report of a WHO Study Group. World Health Organ Tech Rep Ser. 1994; 843: 1– 129. 7941614
- [11]Smith MR McGovern FJ Fallon MA Low bone mineral density in hormone-naive men with prostate carcinoma. Cancer. 2001; 91: 2238– 2245. 11413511
- [12]Budoff MJ Malpeso JM Zeb I Measurement of phantomless thoracic bone mineral density on coronary artery calcium CT scans acquired with various CT scanner models. Radiology. 2013; 267: 830– 836. 23440323
- [13]Mueller DK Kutscherenko A Bartel H Phantom-less QCT BMD system as screening tool for osteoporosis without additional radiation. Eur J Radiol. 2011; 79: 375– 381. 20223609
- [14]Budoff MJ Khairallah W Li D Trabecular bone mineral density measurement using thoracic and lumbar quantitative computed tomography. Acad Radiol. 2012; 19: 179– 183. 22112461
- [15]Aubrey J Esfandiari N Baracos VE Measurement of skeletal muscle radiation attenuation and basis of its biological variation. Acta Physiol (Oxf). 2014; 210: 489– 497. 24393306 PMC4309522
- [17]Habashy AH Yan X Brown JK Estimation of bone mineral density in children from diagnostic CT images: a comparison of methods with and without an internal calibration standard. Bone. 2011; 48: 1087– 1094. 21185418 PMC4780050
- [18]Bauer JS Henning TD Mueller D Volumetric quantitative CT of the spine and hip derived from contrast-enhanced MDCT: conversion factors. AJR Am J Roentgenol. 2007; 188: 1294– 1301. 17449773
- [19]Baum T Muller D Dobritz M BMD measurements of the spine derived from sagittal reformations of contrast-enhanced MDCT without dedicated software. Eur J Radiol. 2011; 80: e140– e145. 20851544
- [20]Baum T Muller D Dobritz M Converted lumbar BMD values derived from sagittal reformations of contrast-enhanced MDCT predict incidental osteoporotic vertebral fractures. Calcif Tissue Int. 2012; 90: 481– 487. 22484555
- [21]General Assembly of the World Medical A. World Medical Association Declaration of Helsinki: ethical principles for medical research involving human subjects. J Am Coll Dent. 2014; 81: 14– 18. 25951678
- [22]Gluer CC Blake G Lu Y Accurate assessment of precision errors: how to measure the reproducibility of bone densitometry techniques. Osteoporos Int. 1995; 5: 262– 270. 7492865
- [23]Acu K Scheel M Issever AS Time dependency of bone density estimation from computed tomography with intravenous contrast agent administration. Osteoporos Int. 2014; 25: 535– 542. 23877871
- [24]Wilmink JT Roukema JG van den Burg W Effects of i.v. contrast administration on intraspinal and paraspinal tissues: a CT study. 2. Visual assessment. AJNR Am J Neuroradiol. 1988; 9: 191– 193. 3124570 PMC8331524
- [25]Jorgensen HS Winther S Bottcher M Effect of intravenous contrast on volumetric bone mineral density in patients with chronic kidney disease. J Clin Densitom. 2016. 27174315
- [26]Gruber M Bauer JS Dobritz M Bone mineral density measurements of the proximal femur from routine contrast-enhanced MDCT data sets correlate with dual-energy X-ray absorptiometry. Eur Radiol. 2013; 23: 505– 512. 22932742
- [27]Ziemlewicz TJ Maciejewski A Binkley N Direct comparison of unenhanced and contrast-enhanced CT for opportunistic proximal femur bone mineral density measurement: implications for osteoporosis screening. AJR Am J Roentgenol. 2016; 206: 694– 698. 26866336
- [28]Ziemlewicz TJ Maciejewski A Binkley N Opportunistic quantitative CT bone mineral density measurement at the proximal femur using routine contrast-enhanced scans: direct comparison with DXA in 355 adults. J Bone Miner Res. 2016. 27082831
- [29]Lochmüller EM Eckstein F Kaiser D Prediction of vertebral failure loads from spinal and femoral dual-energy X-ray absorptiometry, and calcaneal ultrasound: an in situ analysis with intact soft tissues. Bone. 1998; 23: 417– 424. 9823447
- [30]Lochmüller EM Bürklein D Kuhn V Mechanical strength of the thoracolumbar spine in the elderly: prediction from in situ dual-energy X-ray absorptiometry, quantitative computed tomography (QCT), upper and lower limb peripheral QCT, and quantitative ultrasound. Bone. 2002; 31: 77– 84. 12110416
- [31]Alarkawi D Bliuc D Nguyen TV Contribution of Lumbar Spine BMD to Fracture Risk in Individuals With T-Score Discordance. J Bone Miner Res. 2016; 31: 274– 280. 26241926
- [32]Hansen KE Vallarta-Ast N Krueger D Use of the lowest vertebral body T-score to diagnose lumbar osteoporosis in men: is “cherry picking” appropriate? J Clin Densitom. 2004; 7: 376– 381. 15618597
Cited by
Sorted by citations
Löffler, M. T., Jacob, A., Scharr, A., Sollmann, N., Burian, E., El Husseini, M., Sekuboyina, A., Tetteh, G., Zimmer, C., Gempt, J., et al. (2021) · European Radiology · doi:10.1007/s00330-020-07655-2 · Cited by 120 · see citing works
“Bias of BMD values due to intravenous injection of contrast medium was corrected for using linear correction equations for arterial and portal/venous contrast phases [23].”
MethodSelf-citationLi, X., Li, X., Li, J., Jiao, X., Jia, X., Zhang, X., Fan, G., Yang, J., & Guo, J. (2020) · Clinical Radiology · doi:10.1016/j.crad.2019.11.008 · Cited by 41 · see citing works
Moser, L. J., Klambauer, K., Diaz Machicado, M. C., Frey, D., Mergen, V., Eberhard, M., Nowak, T., Schmidt, B., Flohr, T., Distler, O., et al. (2025) · Investigative Radiology · doi:10.1097/rli.0000000000001159 · Cited by 8 · see citing works
- Authors
- Kaesmacher, J., Liebl, H., Baum, T., & Kirschke, J. S.
- Affiliations
- TUM Klinikum (DE)
- Corresponding author
- Kaesmacher, Johannes
- Venue
- Journal of Computer Assisted Tomography
- Published
- 2017
- Volume
- 41
- Issue
- 2
- Pages
- 217-223
- ISSN
- 0363-8715 (ISSN-L)
- PMC
- PMC5359785
- OpenAlex
- W2542118749
- Cited by
- 50
- References
- 0
- Citations by year
- 2026: 4 · 2025: 6 · 2024: 5 · 2023: 5 · 2022: 6 · 2021: 6 · 2020: 7 · 2019: 4 · 2018: 3 · 2017: 4
- FWCI
- 4.47
- Type
- journal article
- Language
- en
- Open access
- hybrid
- doi.org
- Licence
- cc-by-nc-nd
- Locations
- Journal of Computer Assisted Tomography (publishedVersion, cc-by-nc-nd)
- PubMed Central (publishedVersion, cc-by-nc-nd)
- PubMed (publishedVersion)
- BORIS (University Library Bern) (submittedVersion, cc-by-nc-nd)
- mediaTUM (Technical University of Munich) (submittedVersion)
- Open Access CRIS of the University of Bern
- Funders
- Deutsche Forschungsgemeinschaft (4906/1-1, BA 4906/1-1, BA 4085/2-, DFG BA 4085/2-1 and BA 4906/1-1, BA 4085/2-1)
- H2020 European Research Council (637164)
- Topics
- Health Sciences › Medicine › Orthopedics and Sports Medicine › Bone health and osteoporosis research
- Physical Sciences › Engineering › Biomedical Engineering › Advanced X-ray and CT Imaging
- Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Radiation Dose and Imaging
- Keywords
- bone mineral density, multidetector computed tomography, phantom calibration, osteoporosis diagnosis, quantitative computed tomography
- MeSH terms
- Calibration, Contrast Media, Female, Humans, Male, Middle Aged, Osteoporosis, Radiographic Image Enhancement, Bone Density, Phantoms, Imaging, Multidetector Computed Tomography
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review