The DOI matches its Crossref record.
Image Quality and Reconstruction Intervals of Dual-Source CT Coronary Angiography: Recommendations for ECG-Pulsing Windowing
Abstract
PURPOSE: To evaluate reconstruction intervals and image quality in dual-source computed tomography (DSCT) coronary angiography for optimal placement of the ECG-pulsing window. MATERIALS AND METHODS: DSCT coronary angiography was performed in 60 patients. Thirteen datasets were reconstructed in 5% increments from 20-80%. Two readers independently assessed image quality of each segment in each percentage-interval, using scores ranging from 1 (no motion artifacts) to 4 (nonevaluable). RESULTS: Mean heart rate (HR) was 69.0 +/- 18.9 beats per minute (bpm) (range, 35-117 bpm). Diagnostic image quality (scores 1-3) was found in 97.8% of all segments (763 of 780). The 70% RR-interval provided best image quality in all patients and all HRs. The narrowest reconstruction window providing diagnostic image quality was 60-70% for HR <60 bpm, 60-80% for 60-70 bpm, 55-80% for 70-80 bpm, and 30-80% for HR >80 bpm. CONCLUSIONS: DSCT coronary angiography provides best image quality for various HRs at 70%. The ECG-pulsing window can be adapted according to the HR while maintaining diagnostic image quality.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cited by
Schroeder, S., Achenbach, S., Bengel, F., Burgstahler, C., Cademartiri, F., de Feyter, P., George, R., Kaufmann, P., Kopp, A. F., Knuuti, J., et al. (2008) · European Heart Journal · doi:10.1093/eurheartj/ehm544 · Cited by 554
Self-citationSCCT guidelines on radiation dose and dose-optimization strategies in cardiovascular CT
Halliburton, S. S., Abbara, S., Chen, M. Y., Gentry, R., Mahesh, M., Raff, G. L., Shaw, L. J., & Hausleiter, J. (2011) · Journal of Cardiovascular Computed Tomography · doi:10.1016/j.jcct.2011.06.001 · Cited by 491
Contemporary carotid imaging: from degree of stenosis to plaque vulnerability
Brinjikji, W., Huston, J., Rabinstein, A. A., Kim, G.-M., Lerman, A., & Lanzino, G. (2016) · Journal of Neurosurgery · doi:10.3171/2015.1.jns142452 · Cited by 401
Low-dose CT coronary angiography in the step-and-shoot mode: diagnostic performance
Scheffel, H., Alkadhi, H., Leschka, S., Plass, A., Desbiolles, L., Guber, I., Krauss, T., Gruenenfelder, J., Genoni, M., Luescher, T. F., et al. (2008) · Heart · doi:10.1136/hrt.2008.149971 · Cited by 291
Self-citationLow kilovoltage cardiac dual-source CT: attenuation, noise, and radiation dose
Leschka, S., Stolzmann, P., Schmid, F. T., Scheffel, H., Stinn, B., Marincek, B., Alkadhi, H., & Wildermuth, S. (2008) · European Radiology · doi:10.1007/s00330-008-0966-1 · Cited by 290
Self-citation
Metadata
- Authors
- Leschka, S., Scheffel, H., Desbiolles, L., Plass, A., Gaemperli, O., Valenta, I., Husmann, L., Flohr, T. G., Genoni, M., Marincek, B., Kaufmann, P. A., & Alkadhi, H.
- Venue
- Investigative Radiology
- Published
- 2007-08
- Volume
- 42
- Issue
- 8
- Pages
- 543-549
- ISSN
- 0020-9996 (ISSN-L)
- OpenAlex
- W1974239818
- Cited by
- 172
- References
- 0
- FWCI
- 23.03
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Image Quality and Reconstruction Intervals of Dual-Source CT Coronary Angiography: Recommendations for ECG-Pulsing Windowing. https://app.cochat.ai/papers/10.1097/rli.0b013e31803b93cf
- The paper
- Leschka, S., Scheffel, H., Desbiolles, L., Plass, A., Gaemperli, O., Valenta, I., Husmann, L., Flohr, T. G., Genoni, M., Marincek, B., Kaufmann, P. A., & Alkadhi, H. (2007). Image Quality and Reconstruction Intervals of Dual-Source CT Coronary Angiography: Recommendations for ECG-Pulsing Windowing. Investigative Radiology. https://doi.org/10.1097/rli.0b013e31803b93cf
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.