CoChat

Field Strength and Diffusion Encoding Technique Affect the Apparent Diffusion Coefficient Measurements in Diffusion-Weighted Imaging of the Abdomen

Dale, Brian M., Braithwaite, Adam C., Boll, Daniel T., Merkle, Elmar M. · 2010 · Investigative Radiology
Cited by 161

The DOI matches its Crossref record; 21 of the 21 references it cites were found in a registry or an index.

Abstract

OBJECTIVES: The purpose of this study is to determine what effects a variety of diffusion encoding techniques at 1.5 T and 3 T have on measured abdominal apparent diffusion coefficient (ADC) values obtained in a healthy population. MATERIALS AND METHODS: Sixteen healthy male volunteers were enrolled in this prospective Institutional Review Board-approved study following written informed consent. Imaging was performed on a 1.5 T and a 3 T magnetic resonance system (Siemens, Erlangen) with several abdominal axial diffusion weighted imaging (DWI) acquisitions: an orthogonal diffusion encoding with b-values of 0/400 seconds/mm, and a series of four 3-scan trace weighted acquisitions with b-values of 0/50, 0/400, 0/800, 0/50/400/800 seconds/mm, respectively. The mean ADC values were calculated for 3 regions of interest (ROI) in 5 locations (right hepatic lobe, spleen, pancreatic head, body, and tail). The ADC data were analyzed using a repeated-measures analysis of variance. RESULTS: There was a significant difference between measured ADC values at 1.5 T and 3 T for liver (P < 0.001), but not for pancreas (P = 0.427) or spleen (P = 0.167). There was no significant difference (P > 0.999) in the measured ADC values between the orthogonal encodings and the 3-scan trace weighted encoding with the same b-value. There were significant differences (P < 0.001) between all 4 weighting schemes for the 3-scan trace with the measured ADC decreasing with increasing b-value. CONCLUSION: Measured abdominal ADC values depend on the exact selection of b-value used for encoding for liver, pancreas, and spleen. In addition, the measured ADC values depend on the field strength of the scanner for liver.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1097/rli.0b013e3181c8ceac

Cites

21 of 21 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Show all 21 references

Cited by

Metadata

Authors
Dale, B. M., Braithwaite, A. C., Boll, D. T., & Merkle, E. M.
Venue
Investigative Radiology
Published
2010-02
Volume
45
Issue
2
Pages
104-108
ISSN
0020-9996 (ISSN-L)
OpenAlex
W2081519290
Cited by
161
References
21
FWCI
14.46
Type
journal article
Language
en
Open access
closed

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.