American Clinical Neurophysiology Society Guideline 1: Minimum Technical Requirements for Performing Clinical Electroencephalography
Open accessAbstract
Although no single best method exists for recording EEGs under all circumstances, the following standards are considered the minimum for the usual clinical recording of EEGs in all age groups except the very young (see also Guideline 5: Minimum Technical Standards for Pediatric Electroencephalography. Sections where modifications are recommended are delineated by ^for neonates and *for older children). This document discusses minimum requirements; following these requirements alone does not ensure a satisfactory test. Each laboratory should strive for excellence in all aspects of study performance. Recommendations to improve standardization of procedures and to facilitate interchange of recordings and assessment among laboratories in North America have also been included. More detail is provided in recommendations from the International Federation of Clinical Neurophysiology (IFCN).1 1. EQUIPMENT 1.1 ^To display the distribution of EEG activity, it is necessary to record simultaneously from as many regions of the scalp as possible. When too few channels are used, the chances of interpretive errors increase; conversely, when more channels are used, the likelihood of such errors decreases. This is particularly true for transient activity. Sixteen channels of simultaneous recording are considered the minimum number required to show the areas producing most normal and abnormal EEG patterns. Additional channels are often needed for monitoring other physiologic activities and for highlighting activity recorded in specific regions. 1.2 Alternating current (AC) wiring should meet the Underwriters Laboratories standards required for hospital service, as per state and local electrical codes. 1.3 In the usual clinical setting, electrical shielding of the patient and equipment is not necessary, and such shielding need not be installed unless proven necessary. 1.4 Ancillary equipment should include a device for delivering rhythmic, high-intensity flash stimuli to the patient. 1.5 Digital EEG equipment should be used for recording. Digital equipment provides multiple advantages over paper-based EEG recording amplifiers, including greater sensitivity, greater reliability, the possibility of postrecording modification of EEG waveform rendering (gain, filters, montage), and more efficient EEG storage. Digital equipment should conform to the recommendations in Guideline 4: Recording Clinical EEG on Digital Media. 2. ELECTRODES 2.1 ^*Recording electrodes should be free of inherent noise and drift. They should not significantly attenuate signals between 0.5 and 70 Hz. Experimental evidence suggests that silver—silver chloride or gold disk electrodes held on by collodion are the best, but other electrode materials and electrode pastes have been used effectively, especially with contemporary amplifiers having high-input impedances. Electrodes must be disinfected with appropriate procedures and transmission-based precautions taken after recording from patients with contagious diseases (e.g., viral hepatitis, Creutzfeldt–Jakob disease, acquired immunodeficiency syndrome).2 Disposable electrodes may be considered as another option for reducing the risk of iatrogenic infections. 2.2 Needle electrodes are not recommended for routine clinical use. Beyond considerations of patient discomfort and risk of injuries to personnel, these electrodes have higher impedances than appropriately applied cup electrodes, resulting in potentially higher levels of noise. Subdermal needle electrodes (SNE) or wire electrodes (SWE) may be used for prolonged recording of EEG in stuporous or comatose patients in situations where application of cup electrodes is not feasible because of personnel or time constraints. Although impedances are usually higher, they are usually well-matched and stable over long recording periods. Manufacturer's recommendations for insertion, removal, and disposal of each product are available and should be followed. 2.3 ^All 21 electrodes and placements recommended by the IFCN3 should be used. The 10 to 20 System is the only one officially recommended by the IFCN. It is the most commonly used system and should be used universally. The use of the term “modified 10 to 20 System” is undesirable and misleading when it means that head measurements have not been made and placements have been estimated. In this case, the term “estimated 10 to 20 placement” is more appropriate. The term “10 to 10 System” should be used for the extended combinatorial system described in Guideline 2: Guidelines for Standard Electrode Position Nomenclature (For neonates, refer to Guideline 5). With some differences in naming, electrodes in the 10 to 20 System are included in the 10 to 10 System (T3 and T4 in 10–20 system are renamed T7 and T8 in 10–10 system; T5 and T6 in 10–20 system are renamed P7 and P8 in 10–10 system). An adequate number of electrodes are essential to ensure that EEG activity having a small area of representation on the scalp is record
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Report of the committee on methods of clinical examination in electroencephalography
(1958) · Electroencephalography and Clinical Neurophysiology · doi:10.1016/0013-4694(58)90053-1 · Cited by 2,717
Cited by
Clinical utility of EEG in diagnosing and monitoring epilepsy in adults
Tatum, W. O., Rubboli, G., Kaplan, P. W., Mirsatari, S. M., Radhakrishnan, K., Gloss, D., Caboclo, L. O., Drislane, F. W., Koutroumanidis, M., Schomer, D. L., et al. (2018) · Clinical Neurophysiology · doi:10.1016/j.clinph.2018.01.019 · Cited by 314
Self-citationSignal quality and patient experience with wearable devices for epilepsy management
Nasseri, M., Nurse, E., Glasstetter, M., Böttcher, S., Gregg, N. M., Laks Nandakumar, A., Joseph, B., Pal Attia, T., Viana, P. F., Bruno, E., et al. (2020) · Epilepsia · doi:10.1111/epi.16527 · Cited by 90
Nafea, M. S., & Ismail, Z. H. (2025) · AI · doi:10.3390/ai6060120 · Cited by 8
Metadata
- Authors
- Sinha, S. R., Sullivan, L., Sabau, D., San-Juan, D., Dombrowski, K. E., Halford, J. J., Hani, A. J., Drislane, F. W., & Stecker, M. M.
- Venue
- Journal of Clinical Neurophysiology
- Published
- 2016-08
- Volume
- 33
- Issue
- 4
- Pages
- 303-307
- ISSN
- 0736-0258 (ISSN-L)
- OpenAlex
- W2485842390
- Cited by
- 248
- References
- 1
- FWCI
- 9.21
- Type
- journal article
- Language
- en
- Open access
- green
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: American Clinical Neurophysiology Society Guideline 1: Minimum Technical Requirements for Performing Clinical Electroencephalography. https://app.cochat.ai/papers/10.1097/wnp.0000000000000308
- The paper
- Sinha, S. R., Sullivan, L., Sabau, D., San-Juan, D., Dombrowski, K. E., Halford, J. J., Hani, A. J., Drislane, F. W., & Stecker, M. M. (2016). American Clinical Neurophysiology Society Guideline 1: Minimum Technical Requirements for Performing Clinical Electroencephalography. Journal of Clinical Neurophysiology. https://doi.org/10.1097/wnp.0000000000000308
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.