Comparison of impedance and inductance ventilation sensors on adults during breathing, motion, and simulated airway obstruction
The DOI matches its Crossref record; 1 of the 1 references it cites were found in a registry or an index.
Abstract
The goal of this study was to compare the relative performance of two noninvasive ventilation sensing technologies on adults during artifacts. We recorded changes in transthoracic impedance and cross-sectional area of the abdomen (abd) and rib cage (rc) using impedance pneumography (IP) and respiratory inductance plethysmography (RIP) on ten adult subjects during natural breathing, motion artifact, simulated airway obstruction, yawning, snoring, apnea, and coughing. We used a pneumotachometer to measure air flow and tidal volume as the standard. We calibrated all sensors during natural breathing, and performed measurements during all maneuvers without changing the calibration parameters. No sensor provided the most-accurate measure of tidal volume for all maneuvers. Overall, the combination of inductance sensors [RIP(sum)] calibrated during an isovolume maneuver had a bias (weighted mean difference) as low or lower than all individual sensors and all combinations of sensors. The IP(rc) sensor had a bias as low or lower than any individual sensor. The cross-correlation coefficient between sensors was high during natural breathing, but decreased during artifacts. The cross correlation between sensor pairs was lower during artifacts without breathing than it was during maneuvers with breathing for four different sensor combinations. We tested a simple breath-detection algorithm on all sensors and found that RIP(sum) resulted in the fewest number of false breath detections, with sensitivity of 90.8% and positive predictivity of 93.6%.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
Martin Bland, J., & Altman, D. (1986) · The Lancet · doi:10.1016/s0140-6736(86)90837-8 · Cited by 47,965
Cited by
Monitoring of Vital Signs with Flexible and Wearable Medical Devices
Khan, Y., Ostfeld, A. E., Lochner, C. M., Pierre, A., & Arias, A. C. (2016) · Advanced Materials · doi:10.1002/adma.201504366 · Cited by 1,408
Critical review of non-invasive respiratory monitoring in medical care
Folke, M., Cernerud, L., Ekström, M., & Hök, B. (2003) · Medical & Biological Engineering & Computing · doi:10.1007/bf02348078 · Cited by 430
Impedance pneumography: Noise as signal in impedance cardiography
Ernst, J. M., Litvack, D. A., Lozano, D. L., Cacioppo, J. T., & Berntson, G. G. (1999) · Psychophysiology · doi:10.1017/s0048577299981003 · Cited by 168
Wearable Devices for Precision Medicine and Health State Monitoring
Jeong, I. C., Bychkov, D., & Searson, P. C. (2019) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2018.2871638 · Cited by 155
“artifacts [89], [90].”
Lanatà, A., Scilingo, E. P., Nardini, E., Loriga, G., Paradiso, R., & De-Rossi, D. (2010) · IEEE Transactions on Information Technology in Biomedicine · doi:10.1109/titb.2009.2037614 · Cited by 111
“4) Statistics: When two or more measurement techniques are compared in order to find which has the best performance, a statistical investigation is necessary in order to provide a specific confidence interval (CI) in which the results fall [ 22 ].”
Metadata
- Authors
- Cohen, K. P., Ladd, W. M., Beams, D. M., Sheers, W. S., Radwin, R. G., Tompkins, W. J., & Webster, J. G.
- Venue
- IEEE Transactions on Biomedical Engineering
- Published
- 1997-07
- Volume
- 44
- Issue
- 7
- Pages
- 555-566
- ISSN
- 0018-9294 (ISSN-L)
- OpenAlex
- W2137821817
- Cited by
- 113
- References
- 1
- FWCI
- 0.77
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Comparison of impedance and inductance ventilation sensors on adults during breathing, motion, and simulated airway obstruction. https://app.cochat.ai/papers/10.1109/10.594896
- The paper
- Cohen, K. P., Ladd, W. M., Beams, D. M., Sheers, W. S., Radwin, R. G., Tompkins, W. J., & Webster, J. G. (1997). Comparison of impedance and inductance ventilation sensors on adults during breathing, motion, and simulated airway obstruction. IEEE Transactions on Biomedical Engineering. https://doi.org/10.1109/10.594896
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.