A single-channel implantable microstimulator for functional neuromuscular stimulation
The DOI matches its Crossref record; 2 of the 2 references it cites were found in a registry or an index.
Abstract
This paper describes a single-channel implantable microstimulator for functional neuromuscular stimulation. This device measures 2 x 2 x 10 mm3 and can be inserted into paralyzed muscle groups by expulsion from a hypodermic needle. Power and data to the device are supplied from outside by RF telemetry using an amplitude-modulated 2-MHz RF carrier generated using a high-efficiency class-E transmitter. The transmitted signal carries a 5-b address which selects one of the 32 possible microstimulators. The selected device then delivers up to 2 microC of charge store in a tantalum chip capacitor for up to 200 microseconds (10 mA) into loads of < 800 omega through a high-current thin-film iridium-oxide (IrOx) electrode (approximately 0.3 mm2 in area). A bi-CMOS receiver circuitry is used to: generate two regulated voltage supplies (4.5 and 9 V), recover a 2-MHz clock from the carrier, demodulate the address code, and activate the output current delivery circuitry upon the reception of an external command. The overall power dissipation of the receiver circuitry is 45-55 mW. The implant is hermetically packaged using a custom-made glass capsule.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
The sensations produced by electrical stimulation of the visual cortex
Brindley, G. S., & Lewin, W. S. (1968) · The Journal of Physiology · doi:10.1113/jphysiol.1968.sp008519 · Cited by 1,170
Current Density Profiles of Surface Mounted and Recessed Electrodes for Neural Prostheses
Rubinstein, J. T., Spelman, F. A., Soma, M., & Suesserman, M. F. (1987) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.1987.326007 · Cited by 89
Cited by
A review of smart homes—Present state and future challenges
Chan, M., Estève, D., Escriba, C., & Campo, E. (2008) · Computer Methods and Programs in Biomedicine · doi:10.1016/j.cmpb.2008.02.001 · Cited by 1,007
A low-power integrated circuit for a wireless 100-electrode neural recording system.
Harrison, R. R., Watkins, P. T., Kier, R. J., Lovejoy, R. O., Black, D. J., Greger, B., & Solzbacher, F. (2007) · IEEE Journal of Solid-State Circuits · doi:10.1109/jssc.2006.886567 · Cited by 782
A Low-Power Integrated Circuit for a Wireless 100-Electrode Neural Recording System
Harrison, R., Watkins, P., Kier, R., Lovejoy, R., Black, D., Normann, R., & Solzbacher, F. (2006) · 2006 IEEE International Solid State Circuits Conference - Digest of Technical Papers · doi:10.1109/isscc.2006.1696288 · Cited by 731
A BioMEMS Review: MEMS Technology for Physiologically Integrated Devices
Grayson, A. C. R., Shawgo, R. S., Johnson, A. M., Flynn, N. T., Li, Y., Cima, M. J., & Langer, R. (2004) · Proceedings of the IEEE · doi:10.1109/jproc.2003.820534 · Cited by 562
Design and analysis of an adaptive transcutaneous power telemetry for biomedical implants
Wang, G., Liu, W., Sivaprakasam, M., & Kendir, G. A. (2005) · IEEE Transactions on Circuits and Systems I: Regular Papers · doi:10.1109/tcsi.2005.852923 · Cited by 435
Metadata
- Authors
- Ziaie, B., Nardin, M. D., Coghlan, A. R., & Najafi, K.
- Venue
- IEEE Transactions on Biomedical Engineering
- Published
- 1997
- Volume
- 44
- Issue
- 10
- Pages
- 909-920
- ISSN
- 0018-9294 (ISSN-L)
- OpenAlex
- W2162960604
- Cited by
- 137
- References
- 2
- FWCI
- 1.39
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.