The PTB 83-SQUID system for biomagnetic applications in a clinic
The DOI matches its Crossref record.
Abstract
The PTB 83-SQUID (superconducting quantum interference device) system is described which is operated in a clinical environment. Integrated dc SQUID magnetometers with additional positive feedback (APF) are used in order to realize electronic first- or second-order gradiometer configurations. The dewar for the system has a flat bottom. It allows the detection of vertical (B/sub z/) and horizontal (B/sub x,y/) field components: 49 sensors (plus 14 reference magnetometers) are sensitive for B/sub z/ and 14 sensors (plus 6 reference magnetometers) for B/sub x,y/. The system is installed inside a standard shielded room (Vakuumschmelze type AK3b) in the Klinikum Benjamin Franklin, Steglitz, Berlin. A typical white noise level of 2.5 fT//spl radic/(Hz) has been measured in the first-order gradiometer mode. Due to the very low system noise level it became possible for the first time to detect the extremely weak neuromagnetic fields (5-10 fT peak amplitude) generated by the nerve roots deep in the lower back which are evoked by electrical nerve stimulation at the foot (200 Hz system bandwidth, 8000 responses averaged).>
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cited by
Clinical application of magnetocardiography
Fenici, R., Brisinda, D., & Meloni, A. M. (2005) · Expert Review of Molecular Diagnostics · doi:10.1586/14737159.5.3.291 · Cited by 174
“21 Drung D. The PTB 83-SQUID system for biomagnetic applications in a clinic. IEEE Trans. Appl. Supercond. 5, 2112–2117 (1995).”
BackgroundConversion of magnetocardiographic recordings between two different multichannel SQUID devices
Burghoff, M., Nenonen, J., Trahms, L., & Katila, T. (2000) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.846680 · Cited by 51
“ters): The SQUID’s of the 63-channel system developed by the PTB (Berlin, Germany) have a pick-up area of mm and are located on different levels with respect to the flat Dewar bottom [6].”
MethodInfluentialMagnetoneurography: theory and application to peripheral nerve disorders
Mackert, B.-M. (2004) · Clinical Neurophysiology · doi:10.1016/j.clinph.2004.07.028 · Cited by 36
MCG to ECG source differences: Measurements and a two-dimensional computer model study
Weber Dos Santos, R., Kosch, O., Steinhoff, U., Bauer, S., Trahms, L., & Koch, H. (2004) · Journal of Electrocardiology · doi:10.1016/j.jelectrocard.2004.08.036 · Cited by 17
Metadata
- Authors
- Drung, D.
- Venue
- IEEE Transactions on Appiled Superconductivity
- Published
- 1995-06
- Volume
- 5
- Issue
- 2
- Pages
- 2112-2117
- ISSN
- 1051-8223 (ISSN-L)
- OpenAlex
- W2165694677
- Cited by
- 108
- References
- 0
- FWCI
- 5.13
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.