High-Accuracy Real-Time Monitoring of Heart Rate Variability Using 24 GHz Continuous-Wave Doppler Radar
Open accessThe DOI matches its Crossref record; 8 of the 8 references it cites were found in a registry or an index.
Abstract
This paper presents a novel algorithm for the estimation of heart rate variability (HRV) features using 24-GHz continuous-wave Doppler radar with quadrature architecture. The proposed algorithm combines frequency and time domain analysis for high-accuracy estimation of beat-to-beat intervals (BBIs). Initially, band pass filtered in-phase (I) and quadrature (Q) radar components are fused into a single combined signal that contains information on the heartbeats. Its frequency domain analysis is used for coarse heart rate estimation. At the same time, the combined signal is processed using a filter bank containing narrowband band pass filters with different center frequencies. One of the band pass filter outputs is selected as the valid output based on the coarse heart rate estimation. Zero crossings in the resulting filter bank output signal represent heartbeats that are used to extract the BBIs. Finally, four HRV features are calculated from the BBIs. The algorithm is tested on real data obtained from recordings on ten human subjects. The mean relative error of extracted BBIs compared to electrocardiogram (ECG) measurement is in the 1.02-2.07% range. Furthermore, two time-domain and two frequency domain HRV features were calculated from the BBIs. The obtained results show a high level of agreement between radar-extracted and ECG-extracted HRV features. Low computation complexity makes this algorithm suitable for real-time monitoring.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
Martin Bland, J., & Altman, D. (1986) · The Lancet · doi:10.1016/s0140-6736(86)90837-8 · Cited by 47,965
A Real-Time QRS Detection Algorithm
Pan, J., & Tompkins, W. J. (1985) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.1985.325532 · Cited by 7,855 · cited 1×
“BBIs are extracted from the ECG using the Pan-Tompkins algorithm [56].”
MethodAn overview of heart rate variability metrics and norms
Shaffer, F., & Ginsberg, J. P. (2017) · Frontiers in Public Health · doi:10.3389/fpubh.2017.00258 · Cited by 7,185
Detecting Stress During Real-World Driving Tasks Using Physiological Sensors
Healey, J. A., & Picard, R. W. (2005) · IEEE Transactions on Intelligent Transportation Systems · doi:10.1109/tits.2005.848368 · Cited by 2,207 · cited 1×
“High-accuracy monitoring of the heart rate variability (HRV) features is required in numerous applications, such as hospital and in-home health care [1], stress and emotions recognition [2], [3], vigilance monitoring [4], anxiety treatments [5], training optimization [6], etc.”
BackgroundRespiratory sinus arrhythmia in humans: how breathing pattern modulates heart rate
Hirsch, J. A., & Bishop, B. (1981) · American Journal of Physiology-Heart and Circulatory Physiology · doi:10.1152/ajpheart.1981.241.4.h620 · Cited by 1,105 · cited 2×
“Additionally, a respiratory sinus arrhythmia (natural changes in heart rate during a breathing cycle) [42] is also modeled by modulating the BBI sequence with the breathing waveform.”
BackgroundMethodAnxiety Disorders are Associated with Reduced Heart Rate Variability: A Meta-Analysis
Chalmers, J. A., Quintana, D. S., Abbott, M. J.-A., & Kemp, A. H. (2014) · Frontiers in Psychiatry · doi:10.3389/fpsyt.2014.00080 · Cited by 1,067
NUNAN, D., SANDERCOCK, G. R. H., & BRODIE, D. A. (2010) · Pacing and Clinical Electrophysiology · doi:10.1111/j.1540-8159.2010.02841.x · Cited by 874
Emotion recognition using wireless signals
Zhao, M., Adib, F., & Katabi, D. (2016) · Proceedings of the 22nd Annual International Conference on Mobile Computing and Networking · doi:10.1145/2973750.2973762 · Cited by 461 · cited 4×
“Radar technology is one of the most promising methods for non-contact heart rate measurement [3], [8]–[39].”
BackgroundMethodInfluential
Cited by
Human Vital Signs Detection Methods and Potential Using Radars: A Review
Kebe, M., Gadhafi, R., Mohammad, B., Sanduleanu, M., Saleh, H., & Al-Qutayri, M. (2020) · Sensors · doi:10.3390/s20051454 · Cited by 205
“Moreover, a millimeter-wave FMCW radar was used in [170] to assess the HR variation of 3 di ff erent individuals sitting at distances between 0.6 m and 1 m.”
InfluentialmmHRV: Contactless Heart Rate Variability Monitoring Using Millimeter-Wave Radio
Wang, F., Zeng, X., Wu, C., Wang, B., & Liu, K. J. R. (2021) · IEEE Internet of Things Journal · doi:10.1109/jiot.2021.3075167 · Cited by 192
“BPFB shows larger IBI errors with 40ms, 42ms, 43ms, 55ms correspondingly.”
MethodInfluential
Metadata
- Authors
- Petrovic, V. L., Jankovic, M. M., Lupsic, A. V., Mihajlovic, V. R., & Popovic-Bozovic, J. S.
- Venue
- IEEE Access
- Published
- 2019
- Volume
- 7
- Pages
- 74721-74733
- ISSN
- 2169-3536 (ISSN-L)
- OpenAlex
- W2951214606
- Cited by
- 165
- References
- 8
- FWCI
- 6.14
- Type
- journal article
- Language
- en
- Open access
- gold
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: High-Accuracy Real-Time Monitoring of Heart Rate Variability Using 24 GHz Continuous-Wave Doppler Radar. https://app.cochat.ai/papers/10.1109/access.2019.2921240
- The paper
- Petrovic, V. L., Jankovic, M. M., Lupsic, A. V., Mihajlovic, V. R., & Popovic-Bozovic, J. S. (2019). High-Accuracy Real-Time Monitoring of Heart Rate Variability Using 24 GHz Continuous-Wave Doppler Radar. IEEE Access. https://doi.org/10.1109/access.2019.2921240
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.