An Experimental Study of Voltage, Current, and Charge Controlled Stimulation Front-End Circuitry
The DOI matches its Crossref record; 1 of the 1 references it cites were found in a registry or an index.
Abstract
We have developed an experimental 4-channel stimulator prototype capable of performing voltage controlled stimulation (VCS), current controlled stimulation (CCS), and switched capacitor based stimulation (SCS) all on the same platform for neurological experiments. The stimulator allows changing main stimulation parameters such as amplitude, pulse width, pulse frequency, and output channels in real-time either wirelessly or hardwired. The stimulator was used to objectively evaluate the power efficiencies of each stimulation strategy using platinum epilepsy electrodes with 2.3 mm2exposed area in saline solution. We measured 92%, 65%, and 77% efficiencies for VCS, CCS, and SCS, respectively, on a 962Ω - 810 nF resistive-capacitive load. The motivation behind this research is to find the most power efficient stimulation front-end (SFE) circuitry for implantable neurological stimulators (INS).
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Electrical stimulation of excitable tissue: design of efficacious and safe protocols
Merrill, D. R., Bikson, M., & Jefferys, J. G. R. (2005) · Journal of Neuroscience Methods · doi:10.1016/j.jneumeth.2004.10.020 · Cited by 2,078 · cited 1×
“3- Electrode characteristics including size, geometry, material, impedance, surface roughness, and safe charge injection density [7]”
Background
Cited by
The 128-Channel Fully Differential Digital Integrated Neural Recording and Stimulation Interface
Shahrokhi, F., Abdelhalim, K., Serletis, D., Carlen, P. L., & Genov, R. (2010) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2010.2041350 · Cited by 263
A Power-Efficient Wireless System With Adaptive Supply Control for Deep Brain Stimulation
Lee, H.-M., Park, H., & Ghovanloo, M. (2013) · IEEE Journal of Solid-State Circuits · doi:10.1109/jssc.2013.2266862 · Cited by 232
“Conversely, CCS has been widely used because of its precise charge control and safe operation [14].”
BackgroundSelf-citationLee, H.-M., Kwon, K. Y., Li, W., & Ghovanloo, M. (2015) · IEEE Journal of Solid-State Circuits · doi:10.1109/jssc.2014.2355814 · Cited by 161
Self-citationArfin, S. K., & Sarpeshkar, R. (2012) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2011.2166072 · Cited by 133
“Such design leads to even greater energy dissipation.”
BackgroundLyu, H., John, M., Burkland, D., Greet, B., Post, A., Babakhani, A., & Razavi, M. (2020) · Scientific Reports · doi:10.1038/s41598-020-59017-z · Cited by 51
Metadata
- Authors
- Simpson, J., & Ghovanloo, M.
- Venue
- 2007 IEEE International Symposium on Circuits and Systems
- Publisher
- IEEE
- Published
- 2007-05
- Pages
- 325-328
- ISSN
- 0271-4302 (ISSN-L)
- OpenAlex
- W2108585885
- Cited by
- 65
- References
- 1
- FWCI
- 11.37
- Type
- conference paper
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.