An ASIC for Recording and Stimulation in Stacked Microchannel Neural Interfaces
Open accessAbstract
ASIC is implemented in 0.35 μm high-voltage CMOS technology. Each ASIC is the base for seven microchannels each with three electrodes in a pseudo-tripolar arrangement. Multiplexing allows stimulating or recording from any one of 49 channels, across seven ASICs. Connections to the ASICs are made with a five-line parallel bus. Current controlled biphasic stimulation from 5 to 500 μA has been demonstrated with switching between channels and ASICs. The high-voltage technology gives a compliance of 40 V for stimulation, appropriate for the high impedances within microchannels. High frequency biphasic stimulation, up to 40 kHz is achieved, suitable for reversible high frequency nerve blockades. Recording has been demonstrated with mV level signals; common-mode inputs are differentially distorted and limit the CMRR to 40 dB. The ASIC has been used in vitro in conjunction with an oversize (2 mm diameter) microchannel in phosphate buffered saline, demonstrating attenuation of interference from outside the microchannel and tripolar recording of signals from within the microchannel. By using five-lines for 49 active microchannels the device overcomes limitations when connecting many electrodes in a 3-D miniaturized nerve interface.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Proximity of interface sites on tissue can result in crosstalk between channels due to current spread, which can compromise efficacy of the neural interface solution, requiring electrical or physical isolation between interface channels.
“However, the proximity of interface sites on the tissue can result in crosstalk between channels due to current spread. This can compromise the efficacy of the neural interface solution [77], requiring electrical [78] or physical isolation between the interface channel [71].”
From Active Neural Interface Circuits and Systems for Selective Control of Peripheral Nerves: A Review (2024) · cites this paper as [71]
The citation lists haven’t been fetched yet
References 66 · Cited by 15
- Authors
- Lancashire, H. T., Jiang, D., Demosthenous, A., & Donaldson, N.
- Affiliations
- University College London (GB)
- Venue
- IEEE Transactions on Biomedical Circuits and Systems
- Published
- 2019-04
- Volume
- 13
- Issue
- 2
- Pages
- 259-270
- ISSN
- 1932-4545 (ISSN-L)
- 1940-9990
- OpenAlex
- W2909059685
- Cited by
- 15
- References
- 66
- Citations by year
- 2025: 1 · 2024: 2 · 2023: 2 · 2022: 1 · 2021: 6 · 2020: 1 · 2019: 2
- FWCI
- 1.19
- Type
- journal article
- Language
- en
- Open access
- hybrid
- doi.org
- Locations
- IEEE Transactions on Biomedical Circuits and Systems (publishedVersion, cc-by)
- PubMed (publishedVersion)
- UCL Discovery (University College London) (submittedVersion)
- Funders
- Engineering and Physical Sciences Research Council (EP/M506448/1)
- Topics
- Life Sciences › Neuroscience › Cellular and Molecular Neuroscience › Neuroscience and Neural Engineering
- Physical Sciences › Engineering › Biomedical Engineering › Analog and Mixed-Signal Circuit Design
- Physical Sciences › Engineering › Electrical and Electronic Engineering › CCD and CMOS Imaging Sensors
- Keywords
- ASIC, high-voltage CMOS, biphasic stimulation, neural recording, multiplexing, common-mode rejection, electrode impedance
- MeSH terms
- Brain, Electric Stimulation, Electricity, Electrodes, Electrodes, Implanted, Electronics, Medical, Signal Processing, Computer-Assisted
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review