The DOI matches its Crossref record; 228 of the 228 references it cites were found in a registry or an index; 66 of the 166 sources its claims cite were checked for support.
Active Neural Interface Circuits and Systems for Selective Control of Peripheral Nerves: A Review
Open accessAbstract
Interfaces with peripheral nerves have been widely developed to enable bioelectronic control of neural activity. Peripheral nerve neuromodulation shows great potential in addressing motor dysfunctions, neurological disorders, and psychiatric conditions. The integration of high-density neural electrodes with stimulation and recording circuits poses a challenge in the design of neural interfaces. Recent advances in active electrode strategies have achieved improved reliability and performance by implementing in-situ control, stimulation, and recording of neural fibers. This paper presents an overview of state-of-the-art neural interface systems that comprise a range of neural electrodes, neurostimulators, and bio-amplifier circuits, with a special focus on interfaces for the peripheral nerves. A discussion on the efficacy of active electrode systems and recommendations for future directions conclude this paper.
Overview
CoChat Summary
TL;DR
This paper is a review of active neural interface circuits and systems for selective peripheral nerve control, covering stimulation, recording, artifact suppression, and integration challenges. It argues that placing electronics close to the electrodes improves selectivity, reliability, signal quality, and supports closed-loop neuromodulation.
Key findings
- Passive neural interfaces rely on remote electronic control via cables, but they face limits in selectivity, scaling, interconnect reliability, signal quality, and device size; active interfaces address these by in-situ signal conditioning and control.
- The paper identifies three active interface architectures: stand-alone, tethered, and distributed; it notes that distributed systems can enable flexible multisite, high-density coverage, but can increase surgical complexity and face micromotion challenges.
- For stimulation, the paper reports that peripheral-nerve interfaces commonly use current amplitudes from tens of μA to a few mA and pulse widths up to a few hundred μs, with high compliance voltage often needed for small, high-impedance electrodes.
- For recording, neural spike signals are described as μV-level with bandwidths of 200 Hz to 10 kHz, typically requiring about 40 dB to 60 dB preamplification and high CMRR; the paper also notes that implantable microelectrode arrays have achieved noise levels below 1 μVrms.
Methods
- Narrative review of published active neural interface systems and circuit techniques, focused on peripheral nerve applications.
- Synthesizes literature on electrode types, stimulation driver architectures, recording front-ends, artifact-reduction methods, fabrication/packaging, and wireless power/communication.
- Compares representative state-of-the-art systems in tables, including process nodes, channel counts, area, power/data protocol, and artifact/charge-balancing performance where available.
Limitations
- The paper does not present a new experimental dataset or head-to-head empirical evaluation; this is a review article.
- It states that selecting a suitable active interface is highly application-dependent, and that miniaturization, high-voltage compliance, crosstalk, biocompatibility, chronic stability, and power constraints remain unresolved challenges.
- Evident from the text: many performance claims are drawn from heterogeneous prior studies with different nerves, devices, and metrics, so direct comparisons across systems are limited.
Model-written summary. Check it against the paper before relying on it.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
50 Valid · 7 Not valid · 62 Not checked · 23 Own results and interpretations
Every source the paper cites for this claim was checked and supports it.
In the paper
“The electrode's degree of invasiveness is an important factor in the neural response [5].”
What the sources say
- Supportedcites [5] source
The Electrical section (p. 2) contrasts electrodes placed adjacent to, around, or within a nerve and explains that placement affects activation thresholds and the risk of nerve damage. The Penetrating microelectrodes section (p. 14) further reports differences in neural conduction and motor recruitment between invasive and extraneural electrodes. Together, these support the claim that invasiveness matters to neural response. (LLM, conf=0.87)
plainstates an association- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Supported
The cited sources support only part of this claim — read each source’s verdict below.
In the paper
“Interfaces of the peripheral nerves such as the ultrasonic neural dust, StimDust, and the NeuroTalk active Utah electrode array demonstrated good performance in monitoring and activating the sciatic nerve [42], [52], [53].”
What the sources say
- Partly supportedcites [42] sourceExperimental — accuracy not yet measured
The Introduction and Results report that ultrasonic neural dust recorded signals from the sciatic nerve in anesthetized rats, and the Discussion says its recording performance was equivalent to conventional wired recordings. The paper does not show neural dust activating the sciatic nerve; stimulation used separate electrodes. The other named devices are associated with other citations in the sentence, not this source. (LLM, conf=0.87)
- Not checked — no full text of the sourcecites [52] source
- Partly supportedcites [53] sourceExperimental — accuracy not yet measured
The Methods (Device design and assembly) describe a wireless microelectrode array using the NeuroTalk interface, and the Main results report reliable activation of the rat sciatic nerve to evoke hindlimb movement. The paper does not demonstrate monitoring sciatic-nerve activity with that array; it reports stimulation and observation of movement. The other named interfaces are not evaluated in this paper. (LLM, conf=0.86)
boostedstates an association- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Partly supported
The cited sources support only part of this claim — read each source’s verdict below.
In the paper
“Passive neural interfaces have proved as effective bioelectronic therapies in managing neurological and psychiatric disorders, as well as treatments for sensory and motor dysfunctions [1].”
What the sources say
- Partly supportedcites [1] sourceExperimental — accuracy not yet measured
The Abstract (page unavailable) says implanted electrical stimulation devices are used to treat neurological disorders, supporting that part of the claim. But the Conclusions and perspective section describes therapies for psychiatric conditions as something future insights will inform, while the Multi-modal interaction section reports restored locomotion only in rodents after spinal cord injury. Our copy does not establish that passive interfaces have proved effective across psychiatric, sensory, and motor dysfunctions. (LLM, conf=0.86)
plainstates a cause- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Partly supported
Every source the paper cites for this claim was checked and supports it.
In the paper
“For example, non-selective vagus nerve stimulation (VNS) results in coughs, headaches, throat pain, nausea and apnea.”
What the sources say
- Supportedcites [3] source
The Introduction identifies cough, headache, throat and tonsil pain, nausea, and apnea among VNS side effects, and says these result primarily from stimulating the whole vagus nerve rather than stimulating it selectively. The section ‘Side Effects of Non-selective VNS’ further discusses throat pain, cough, and acute apnea. (LLM, conf=0.94)
plainstates a cause- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Supported
The cited sources support only part of this claim — read each source’s verdict below.
In the paper
“Mechanical modes of neurostimulation enable non-invasive technologies such as ultrasound and magnetic forces by operating on the basis of mechanotransduction to convert mechanical stimuli into electrophysiological signals through compression, expansion, bending and tension [108].”
What the sources say
- Partly supportedcites [108] sourceExperimental — accuracy not yet measured
Split into 2 atoms — 0 confirmed, 2 partially supported, 0 disputed, 0 misattributed, 0 unverifiable. Expand the claim to see per-atom verdicts.
plainstates a cause- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Partly supported
63 more claims · 22 with their cited sources checked
Create a free account to see all 142 claims and every source verdict.
- It may state two separate claims as one. 46
- Probably not a substantial claim. 37
Citations
Cites
Design of Analog CMOS Integrated Circuits
Razavi, B. (1999) · Cited by 6,609
Neural Stimulation and Recording Electrodes
Cogan, S. F. (2008) · Annual Review of Biomedical Engineering · doi:10.1146/annurev.bioeng.10.061807.160518 · Cited by 2,243 · cited 1×
“This reduces the occurrence of irreversible electrochemical processes at the tissue such as water electrolysis, pH changes, and electrode dissolution or corrosion [102].”
Electrical stimulation of excitable tissue: design of efficacious and safe protocols
Merrill, D. R., Bikson, M., & Jefferys, J. G. R. (2005) · Journal of Neuroscience Methods · doi:10.1016/j.jneumeth.2004.10.020 · Cited by 2,078 · cited 5×
“Faradaic reactions where products are not available for the reverse reaction are not reversed, for example hydrolysis of water or “water window” which places safety limits on electrode-electrolyte interface potentials [101].”
InfluentialTranscranial Magnetic Stimulation: A Primer
Hallett, M. (2007) · Neuron · doi:10.1016/j.neuron.2007.06.026 · Cited by 1,891 · cited 1×
“Primarily developed as a painless therapy for the brain as transcranial magnetic stimulation, this technique has also been used to diagnose peripheral damage through direct stimulation to the nerves [110].”
Enz, C. C., & Temes, G. C. (1996) · Proceedings of the IEEE · doi:10.1109/5.542410 · Cited by 1,877 · cited 1×
“It results in minimum low-frequency noise, while limiting the signal or noise aliasing and maintaining a zero dc gain [175], [176].”
The development and application of optogenetics
Fenno, L., Yizhar, O., & Deisseroth, K. (2011) · Annual Review of Neuroscience · doi:10.1146/annurev-neuro-061010-113817 · Cited by 1,822 · cited 1×
“The emergence of optogenetic technologies has enabled the genetic encoding of neurons with light-gated ion channels such as Channelrhodopsin-2 (ChR2) and Halorhodopsin (HR), that serve as sensory photoreceptors [106].”
A low-power low-noise cmos for amplifier neural recording applications
Harrison, R. R., & Charles, C. (2003) · IEEE Journal of Solid-State Circuits · doi:10.1109/jssc.2003.811979 · Cited by 1,695 · cited 3×
“Provided that C I , C L (cid:2) C F , the high-pass corner, ω H is approximately g m / ( A M C L ) , where g m , A M , and C L represent the OTA transconductance, the midband gain, and the load capacitance, respectively [168].”
InfluentialDeep brain stimulation: current challenges and future directions
Lozano, A. M., Lipsman, N., Bergman, H., Brown, P., Chabardes, S., Chang, J. W., Matthews, K., McIntyre, C. C., Schlaepfer, T. E., Schulder, M., et al. (2019) · Nature Reviews Neurology · doi:10.1038/s41582-018-0128-2 · Cited by 1,450 · cited 1×
“…to manage neurological and psychiatric disorders such as Parkinson’s disease and essential tremor, Alzheimer’s disease, epilepsy, depression via deep brain stimulation [6], and those associated with the spinal cord to treat chronic neuropathic pain, motor dysfunction, and sensory paralysis [7].”
Neuropixels 2.0: A miniaturized high-density probe for stable, long-term brain recordings.
Steinmetz, N. A., Aydin, C., Lebedeva, A., Okun, M., Pachitariu, M., Bauza, M., Beau, M., Bhagat, J., Böhm, C., Broux, M., et al. (2021) · Science · doi:10.1126/science.abf4588 · Cited by 1,173 · cited 1×
“Neural probes enable high-density interfacing with minimum invasiveness using a narrow, rigid shank, which can accommodate hundreds to thousands of pixels or electrode sites for recording [17], [73] or stimulating [74], [75] neural activity.”
Flexible Electronics: Stretchable Electrodes and Their Future
Huang, S., Liu, Y., Zhao, Y., Ren, Z., & Guo, C. F. (2019) · Advanced Functional Materials · doi:10.1002/adfm.201805924 · Cited by 855 · cited 2×
“Biomechanical energy harvesting systems such as triboelectric and piezoelectric generators harness the electrical field resulting from charge separation and mechanical strain, respectively, enabling in-vivo powering from the natural movement of the organs [208], [216].”
Research works printed but not linked: 2
- O’Reardon, 2006, Vagus nerve stimulation (VNS) and treatment of depression: To the brainstem and beyond, Psychiatry (edgmont), 3, 54
- Elansary, 2021, Transmission and 64 2nd-order opamp-less $\Sigma\Delta$ ADCs, 31, 400
Cited by
An Active Microchannel Neural Interface for Implantable Electrical Stimulation and Recording
Habibollahi, M., Jiang, D., Lancashire, H. T., & Demosthenous, A. (2025) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2025.3533612 · Cited by 8
“Fewer integrated devices have been reported for the PNS [17].”
Self-citationCMOS Circuits in the Design and Development of Implantable Medical Devices
Laskar, M. J., Banerjee, A., Prasad, V., & Das, D. (2025) · Biomedical Materials & Devices · doi:10.1007/s44174-024-00267-0 · Cited by 7
“CMOS circuits produce minimal heat due to their efficient power usage, making them safer for long-term implantation [31].”
Biotic‐Abiotic Interface Engineering for Peripheral Nerve Modulation and Repair
Jekal, J., Park, J. T., Kim, E., Lee, Y., & Jang, K. (2025) · Small Methods · doi:10.1002/smtd.202501620 · Cited by 2
Olorocisimo, J. P., Nag, S., Zhang, H., Yang, S. M., Prytula, M., Liu, S., Kanchwala, M., Sun, Y., Zariffa, J., & Genov, R. (2025) · Journal of NeuroEngineering and Rehabilitation · doi:10.1186/s12984-025-01825-3 · Cited by 3
“Current peripheral nerve interfaces are generally grouped into 3 categories, which from least to most invasive are: extraneural, intraneural, and regenerative [1, 18, 19].”
Adhesion Promotion for Non-Hermetic Encapsulations of Implantable Ceramic Printed Circuit Boards
Martens, J., Henneken, L., & Stieglitz, T. (2025) · International IEEE/EMBS Conference on Neural Engineering · doi:10.1109/ner61569.2025.11588992 · Cited by 1
“Thanks to modern medical technology, sensory abilities can also be restored by stimulating nerves of the sensory branches in the peripheral nervous system (PNS) [1, 3].”
Metadata
- Authors
- Habibollahi, M., Jiang, D., Lancashire, H. T., & Demosthenous, A.
- Affiliations
- University College London (GB)
- Venue
- IEEE Transactions on Biomedical Circuits and Systems
- Published
- 2024-10
- Volume
- 18
- Issue
- 5
- Pages
- 954-975
- ISSN
- 1932-4545 (ISSN-L)
- 1940-9990
- OpenAlex
- W4400727368
- Cited by
- 11
- References
- 215
- Citations by year
- 2026: 5 · 2025: 6
- FWCI
- 0.69
- Type
- journal article
- Language
- en
- Open access
- hybrid
- doi.org
- Locations
- IEEE Transactions on Biomedical Circuits and Systems (publishedVersion, cc-by)
- PubMed (publishedVersion)
- UCL Discovery (University College London) (submittedVersion, cc-by)
- Funders
- Engineering and Physical Sciences Research Council (EPSRC), U.K. Research and Innovation (EP/R513143/1)
- Topics
- Life Sciences › Neuroscience › Cellular and Molecular Neuroscience › Neuroscience and Neural Engineering
- Physical Sciences › Engineering › Electrical and Electronic Engineering › Analog and Mixed-Signal Circuit Design
- Life Sciences › Neuroscience › Neurology › Vagus Nerve Stimulation Research
- Keywords
- peripheral nerve interfaces, nerve stimulator, motor impairment, neurological disorders, psychiatric conditions
- MeSH terms
- Animals, Electric Stimulation, Equipment Design, Humans, Peripheral Nerves
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
The claim verdicts this report counts are experimental: their accuracy is not yet measured.
A language model drew the claims from the paper on 2026-10-11, under version 3 of the claim definition (model deepseek-v4.1-flash, prompt extract-v3).
A second model labelled each claim on 2026-10-11 (model jev-1.13-20260917, definition revision gold-guideline-v0.2).
Cited sources are checked with a language model’s help (gpt-6-sol).
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Active Neural Interface Circuits and Systems for Selective Control of Peripheral Nerves: A Review. https://app.cochat.ai/papers/10.1109/tbcas.2024.3430038
- The paper
- Habibollahi, M., Jiang, D., Lancashire, H. T., & Demosthenous, A. (2024). Active Neural Interface Circuits and Systems for Selective Control of Peripheral Nerves: A Review. IEEE Transactions on Biomedical Circuits and Systems. https://doi.org/10.1109/tbcas.2024.3430038
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.