Fascicle-Selectivity of an Intraneural Stimulation Electrode in the Rabbit Sciatic Nerve
Abstract
The current literature contains extensive research on peripheral nerve interfaces, including both extraneural and intrafascicular electrodes. Interfascicular electrodes, which are in-between these two with respect to nerve fiber proximity have, however, received little interest. In this proof-of-concept study, an interfascicular electrode was designed to be implanted in the sciatic nerve and activate the tibial and peroneal nerves selectively of each other, and it was tested in acute experiments on nine anaesthetized rabbits. The electrode was inserted without difficulty between the fascicles using blunt glass tools, which could easily penetrate the epineurium but not the perineurium. Selective activation of all tibial and peroneal nerves in the nine animals was achieved with high selectivity (Ŝ = 0.98 ± 0.02). Interfascicular electrodes could provide an interesting addition to the bulk of peripheral nerve interfaces available for neural prosthetic devices. Since interfascicular electrodes can be inserted without fully freeing the nerve and have the advantage of not confining the nerve to a limited space, they could, e.g., be an alternative to extraneural electrodes in locations where such surgery is complicated due to blood vessels or fatty tissue. Further studies are, however, necessary to develop biocompatible electrodes and test their stability and safety in chronic experiments.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Interfascicular interfaces have been explored in the past with varying degrees of success.
“Interfascicular interfaces have been explored in the past with varying degrees of success [37-39].”
From Directed stimulation with interfascicular interfaces for peripheral nerve stimulation (2021) · cites this paper as [39]
The SPINE used silicone 'beams' to shield the contacts and confine the electric field to a smaller region.
“The SPINE used silicone 'beams' to shield the contacts and confine the electric field to a smaller region.”
From Directed stimulation with interfascicular interfaces for peripheral nerve stimulation (2021) · cites this paper as [39]
Interfascicular interfaces designed so far have used versions of directional contacts.
“Interfascicular interfaces designed so far have used versions of directional contacts [37-39].”
From Directed stimulation with interfascicular interfaces for peripheral nerve stimulation (2021) · cites this paper as [39]
Two interfascicular electrode designs tested in rabbit experiments confirmed the capability of interfascicular electrodes in interfacing with subpopulations of nerve fibers.
“These studies also confirmed the capability of interfascicular electrodes in interfacing with subpopulations of nerve fibers.”
From Interfaces with the peripheral nervous system for the control of a neuroprosthetic limb: a review (2020) · cites this paper as [82]
The citation lists haven’t been fetched yet
References 33 · Cited by 11
- Authors
- Nielsen, T. N., Sevcencu, C., & Struijk, J. J.
- Affiliations
- Aalborg University (DK)
- Venue
- IEEE Transactions on Biomedical Engineering
- Published
- 2012-01
- Volume
- 59
- Issue
- 1
- Pages
- 192-197
- ISSN
- 0018-9294 (ISSN-L)
- 1558-2531
- OpenAlex
- W1975528452
- Cited by
- 11
- References
- 33
- Citations by year
- 2022: 1 · 2021: 2 · 2020: 1 · 2019: 1 · 2018: 2 · 2017: 1 · 2014: 1 · 2013: 2
- FWCI
- 0.40
- Type
- journal article
- Language
- en
- Open access
- closed
- Locations
- IEEE Transactions on Biomedical Engineering (publishedVersion)
- PubMed (publishedVersion)
- VBN Forskningsportal (Aalborg Universitet) (submittedVersion)
- Topics
- Life Sciences › Neuroscience › Cellular and Molecular Neuroscience › Neuroscience and Neural Engineering
- Life Sciences › Neuroscience › Neurology › Vagus Nerve Stimulation Research
- Physical Sciences › Engineering › Biomedical Engineering › Muscle activation and electromyography studies
- Keywords
- intraneural stimulation, sciatic nerve, tibial nerve, peroneal nerve, peripheral nerve interfaces, neuroprosthetics, neurostimulation, rabbit model
- MeSH terms
- Action Potentials, Animals, Differential Threshold, Electric Stimulation, Electrodes, Implanted, Equipment Design, Nerve Fibers, Rabbits, Sciatic Nerve, Sensitivity and Specificity, Equipment Failure Analysis
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review