Scatter Correction Method for X-Ray CT Using Primary Modulation: Theory and Preliminary Results

Zhu, Lei, Bennett, N. Robert, Fahrig, Rebecca · 2006 · IEEE Transactions on Medical Imaging
Cited by 220

The DOI matches its Crossref record; 3 of the 3 references it cites were found in a registry or an index.

Abstract

An X-ray system with a large area detector has high scatter-to-primary ratios (SPRs), which result in severe artifacts in reconstructed computed tomography (CT) images. A scatter correction algorithm is introduced that provides effective scatter correction but does not require additional patient exposure. The key hypothesis of the algorithm is that the high-frequency components of the X-ray spatial distribution do not result in strong high-frequency signals in the scatter. A calibration sheet with a checkerboard pattern of semitransparent blockers (a "primary modulator") is inserted between the X-ray source and the object. The primary distribution is partially modulated by a high-frequency function, while the scatter distribution still has dominant low-frequency components, based on the hypothesis. Filtering and demodulation techniques suffice to extract the low-frequency components of the primary and hence obtain the scatter estimation. The hypothesis was validated using Monte Carlo (MC) simulation, and the algorithm was evaluated by both MC simulations and physical experiments. Reconstructions of a software humanoid phantom suggested system parameters in the physical implementation and showed that the proposed method reduced the relative mean square error of the reconstructed image in the central region of interest from 74.2% to below 1%. In preliminary physical experiments on the standard evaluation phantom, this error was reduced from 31.8% to 2.3%, and it was also demonstrated that the algorithm has no noticeable impact on the resolution of the reconstructed image in spite of the filter-based approach. Although the proposed scatter correction technique was implemented for X-ray CT, it can also be used in other X-ray imaging applications, as long as a primary modulator can be inserted between the X-ray source and the imaged object.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1109/tmi.2006.884636

Cites

3 of 3 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Cited by

See all 220

Metadata

Authors
Zhu, L., Bennett, N. R., & Fahrig, R.
Venue
IEEE Transactions on Medical Imaging
Published
2006-12
Volume
25
Issue
12
Pages
1573-1587
ISSN
0278-0062 (ISSN-L)
OpenAlex
W2158776862
Cited by
220
References
3
FWCI
4.88
Type
journal article
Language
en
Open access
closed

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.