Monte Carlo study of the effects of system geometry and antiscatter grids on cone‐beam CT scatter distributions
Open accessAbstract
Purpose: The proliferation of cone‐beam CT (CBCT) has created interest in performance optimization, with x‐ray scatter identified among the main limitations to image quality. CBCT often contends with elevated scatter, but the wide variety of imaging geometry in different CBCT configurations suggests that not all configurations are affected to the same extent. Graphics processing unit (GPU) accelerated Monte Carlo (MC) simulations are employed over a range of imaging geometries to elucidate the factors governing scatter characteristics, efficacy of antiscatter grids, guide system design, and augment development of scatter correction. Methods: A MC x‐ray simulator implemented on GPU was accelerated by inclusion of variance reduction techniques (interaction splitting, forced scattering, and forced detection) and extended to include x‐ray spectra and analytical models of antiscatter grids and flat‐panel detectors. The simulator was applied to small animal (SA), musculoskeletal (MSK) extremity, otolaryngology (Head), breast, interventional C‐arm, and on‐board (kilovoltage) linear accelerator (Linac) imaging, with an axis‐to‐detector distance (ADD) of 5, 12, 22, 32, 60, and 50 cm, respectively. Each configuration was modeled with and without an antiscatter grid and with (i) an elliptical cylinder varying 70–280 mm in major axis; and (ii) digital murine and anthropomorphic models. The effects of scatter were evaluated in terms of the angular distribution of scatter incident upon the detector, scatter‐to‐primary ratio (SPR), artifact magnitude, contrast, contrast‐to‐noise ratio (CNR), and visual assessment. Results: Variance reduction yielded improvements in MC simulation efficiency ranging from ∼17‐fold (for SA CBCT) to ∼35‐fold (for Head and C‐arm), with the most significant acceleration due to interaction splitting (∼6 to ∼10‐fold increase in efficiency). The benefit of a more extended geometry was evident by virtue of a larger air gap—e.g., for a 16 cm diameter object, the SPR reduced from 1.5 for ADD = 12 cm (MSK geometry) to 1.1 for ADD = 22 cm (Head) and to 0.5 for ADD = 60 cm (C‐arm). Grid efficiency was higher for configurations with shorter air gap due to a broader angular distribution of scattered photons—e.g., scatter rejection factor ∼0.8 for MSK geometry versus ∼0.65 for C‐arm. Grids reduced cupping for all configurations but had limited improvement on scatter‐induced streaks and resulted in a loss of CNR for the SA, Breast, and C‐arm. Relative contribution of forward‐directed scatter increased with a grid (e.g., Rayleigh scatter fraction increasing from ∼0.15 without a grid to ∼0.25 with a grid for the MSK configuration), resulting in scatter distributions with greater spatial variation (the form of which depended on grid orientation). Conclusions: A fast MC simulator combining GPU acceleration with variance reduction provided a systematic examination of a range of CBCT configurations in relation to scatter, highlighting the magnitude and spatial uniformity of individual scatter components, illustrating tradeoffs in CNR and artifacts and identifying the system geometries for which grids are more beneficial (e.g., MSK) from those in which an extended geometry is the better defense (e.g., C‐arm head imaging). Compact geometries with an antiscatter grid challenge assumptions of slowly varying scatter distributions due to increased contribution of Rayleigh scatter.
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Traditional scatter estimation relies on variance-reduced MC simulations, which requires a complete MC setup.
“Traditional scatter estimation relies on variancereduced MC simulations [13], which requires a complete MC setup.”
From DeepDRR – A Catalyst for Machine Learning in Fluoroscopy-Guided Procedures (2018) · cites this paper as [13]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Feldkamp, L. A., Davis, L. C., & Kress, J. W. (1984) · Journal of the Optical Society of America A · doi:10.1364/josaa.1.000612 · Cited by 6,363
Fast calculation of the exact radiological path for a three‐dimensional CT array
Siddon, R. L. (1985) · Medical Physics · doi:10.1118/1.595715 · Cited by 1,535
Flat-panel cone-beam computed tomography for image-guided radiation therapy
Jaffray, D. A., Siewerdsen, J. H., Wong, J. W., & Martinez, A. A. (2002) · International Journal of Radiation Oncology*Biology*Physics · doi:10.1016/s0360-3016(02)02884-5 · Cited by 1,331
Self-citationPENELOPE: A code system for Monte Carlo simulation of electron and photon transport
Salvat, F., Fernández-Varea, J. M., & Sempau Roma, J. (2020) · Penelope · doi:10.1787/9d2cc3d5-en · Cited by 1,284
The composition of body tissues
WOODARD, H. Q., & White, D. R. (1986) · British Journal of Radiology · doi:10.1259/0007-1285-59-708-1209 · Cited by 769
An accurate method for computer‐generating tungsten anode x‐ray spectra from 30 to 140 kV
Boone, J. M., & Seibert, J. A. (1997) · Medical Physics · doi:10.1118/1.597953 · Cited by 708
Cone-beam computed tomography with a flat-panel imager: Magnitude and effects of x-ray scatter
Siewerdsen, J. H., & Jaffray, D. A. (2001) · Medical Physics · doi:10.1118/1.1339879 · Cited by 616
Self-citationAn Efficient Method for Generating Discrete Random Variables with General Distributions
Walker, A. J. (1977) · ACM Transactions on Mathematical Software · doi:10.1145/355744.355749 · Cited by 607
Digimouse: a 3D whole body mouse atlas from CT and cryosection data
Dogdas, B., Stout, D., Chatziioannou, A. F., et al. (2007) · Physics in Medicine and Biology · doi:10.1088/0031-9155/52/3/003 · Cited by 533
Dedicated Breast CT: Radiation Dose and Image Quality Evaluation
Boone, J. M., Nelson, T. R., Lindfors, K. K., & Seibert, J. A. (2001) · Radiology · doi:10.1148/radiol.2213010334 · Cited by 466
Badal, A., & Badano, A. (2009) · Medical Physics · doi:10.1118/1.3231824 · Cited by 358
Self-citationX‐ray scatter correction algorithm for cone beam CT imaging
Ning, R., Tang, X., & Conover, D. L. (2004) · Medical Physics · doi:10.1118/1.1711475 · Cited by 323
Siewerdsen, J. H., Moseley, D. J., Burch, S., et al. (2005) · Medical Physics · doi:10.1118/1.1836331 · Cited by 306
Self-citationFlat-detector computed tomography (FD-CT)
Kalender, W. A., & Kyriakou, Y. (2007) · European Radiology · doi:10.1007/s00330-007-0651-9 · Cited by 304
Siewerdsen, J. H., Daly, M. J., Bakhtiar, B., et al. (2005) · Medical Physics · doi:10.1118/1.2148916 · Cited by 284
Self-citationSpektr: A computational tool for x‐ray spectral analysis and imaging system optimization
Siewerdsen, J. H., Waese, A. M., Moseley, D. J., Richard, S., & Jaffray, D. A. (2004) · Medical Physics · doi:10.1118/1.1758350 · Cited by 281
Self-citationMonte Carlo evaluation of kerma at a point for photon transport problems
Williamson, J. F. (1987) · Medical Physics · doi:10.1118/1.596069 · Cited by 242
Grills, I. S., Hugo, G. D., Kestin, L. L., et al. (2007) · International Journal of Radiation Oncology*Biology*Physics · doi:10.1016/j.ijrobp.2007.07.2352 · Cited by 238
Improved scatter correction using adaptive scatter kernel superposition
Sun, M., & Star-Lack, J. M. (2010) · Physics in Medicine and Biology · doi:10.1088/0031-9155/55/22/007 · Cited by 233
Siewerdsen, J. H., Moseley, D. J., Bakhtiar, B., et al. (2004) · Medical Physics · doi:10.1118/1.1819789 · Cited by 227
Self-citationScatter Correction Method for X-Ray CT Using Primary Modulation: Theory and Preliminary Results
Zhu, L., Bennett, N. R., & Fahrig, R. (2006) · IEEE Transactions on Medical Imaging · doi:10.1109/tmi.2006.884636 · Cited by 220
Siewerdsen, J. H., & Jaffray, D. A. (2000) · Medical Physics · doi:10.1118/1.1286590 · Cited by 216
Self-citationThe effects of scatter in x‐ray computed tomography
Joseph, P. M., & Spital, R. D. (1982) · Medical Physics · doi:10.1118/1.595111 · Cited by 216
Effect of scattered radiation on image noise in cone beam CT
Endo, M., Tsunoo, T., Nakamori, N., et al. (2001) · Medical Physics · doi:10.1118/1.1357457 · Cited by 212
Grids or air gaps for scatter reduction in digital radiography: A model calculation
Neitzel, U. (1992) · Medical Physics · doi:10.1118/1.596836 · Cited by 206
Scatter correction for cone‐beam CT in radiation therapy
Zhu, L., Xie, Y., Wang, J., et al. (2009) · Medical Physics · doi:10.1118/1.3130047 · Cited by 205
Characterization of scattered radiation in kV CBCT images using Monte Carlo simulations
Jarry, G., Graham, S. A., Moseley, D. J., et al. (2006) · Medical Physics · doi:10.1118/1.2358324 · Cited by 187
The influence of bowtie filtration on cone‐beam CT image quality
Mail, N., Moseley, D. J., Siewerdsen, J. H., et al. (2008) · Medical Physics · doi:10.1118/1.3017470 · Cited by 180
Self-citationDose and image quality for a cone‐beam C‐arm CT system
Fahrig, R., Dixon, R. L., Payne, T., et al. (2006) · Medical Physics · doi:10.1118/1.2370508 · Cited by 176
Coherent scatter in diagnostic radiology
Johns, P. C., & Yaffe, M. J. (1983) · Medical Physics · doi:10.1118/1.595443 · Cited by 169
Compton scatter effects in CT reconstructions
Glover, G. H. (1982) · Medical Physics · doi:10.1118/1.595197 · Cited by 167
X‐ray scatter removal by deconvolution
Seibert, J. A., & Boone, J. M. (1988) · Medical Physics · doi:10.1118/1.596208 · Cited by 165
Zbijewski, W. B., Jean, P. D., Prakash, P., et al. (2011) · Medical Physics · doi:10.1118/1.3611039 · Cited by 151
Self-citationAn efficient Monte Carlo-based algorithm for scatter correction in keV cone-beam CT
Poludniowski, G. G., Evans, P., Hansen, V. N., et al. (2009) · Physics in Medicine and Biology · doi:10.1088/0031-9155/54/12/016 · Cited by 147
Efficient Monte Carlo based scatter artifact reduction in cone-beam micro-CT
Zbijewski, W. B., & Beekman, F. J. (2006) · IEEE Transactions on Medical Imaging · doi:10.1109/tmi.2006.872328 · Cited by 146
Self-citationScatter estimation for a digital radiographic system using convolution filtering
Love, L. A., & Kruger, R. A. (1987) · Medical Physics · doi:10.1118/1.596126 · Cited by 138
Kyriakou, Y., Riedel, T., & Kalender, W. A. (2006) · Physics in Medicine and Biology · doi:10.1088/0031-9155/51/18/008 · Cited by 134
Accelerated Simulation of Cone Beam X-Ray Scatter Projections
Colijn, A. P., & Beekman, F. J. (2004) · IEEE Transactions on Medical Imaging · doi:10.1109/tmi.2004.825600 · Cited by 131
Evaluation of the spatial resolution characteristics of a cone‐beam breast CT scanner
Kwan, A. L. C., Boone, J. M., Yang, K., et al. (2006) · Medical Physics · doi:10.1118/1.2400830 · Cited by 127
Scattered radiation in fan beam imaging systems
Johns, P. C., & Yaffe, M. J. (1982) · Medical Physics · doi:10.1118/1.595076 · Cited by 120
Generalized DQE analysis of radiographic and dual‐energy imaging using flat‐panel detectors
Richard, S., Siewerdsen, J. H., Jaffray, D. A., et al. (2005) · Medical Physics · doi:10.1118/1.1901203 · Cited by 119
Self-citationStatistical limitations in transaxial tomography
Barrett, H. H., Gordon, S. K., & Hershel, R. S. (1976) · Computers in Biology and Medicine · doi:10.1016/0010-4825(76)90068-8 · Cited by 109
Evaluation of x‐ray scatter properties in a dedicated cone‐beam breast CT scanner
Kwan, A. L. C., Boone, J. M., & Shah, N. (2005) · Medical Physics · doi:10.1118/1.1954908 · Cited by 108
On the Estimation of Flux at a Point by Monte Carlo
Kalos, M. H. (1963) · Nuclear Science and Engineering · doi:10.13182/nse63-a26481 · Cited by 103
Maltz, J. S., Gangadharan, B., Bose, S., et al. (2008) · IEEE Transactions on Medical Imaging · doi:10.1109/tmi.2008.928922 · Cited by 89
CT arthrography of the wrist using a novel, mobile, dedicated extremity cone-beam CT (CBCT)
Koskinen, S. K., Haapamäki, V. V., Salo, J., et al. (2012) · Skeletal Radiology · doi:10.1007/s00256-012-1516-0 · Cited by 83
Kyprianou, I. S., Rudin, S., Bednarek, D. R., et al. (2005) · Medical Physics · doi:10.1118/1.1844151 · Cited by 82
Self-citationDedicated Cone-Beam Breast CT: Feasibility Study with Surgical Mastectomy Specimens
Yang, W. T., Çarkaci, S., Chen, L., et al. (2007) · American Journal of Roentgenology · doi:10.2214/ajr.07.2403 · Cited by 76
Variance reduction techniques for fast Monte Carlo CBCT scatter correction calculations
Mainegra‐Hing, E., & Kawrakow, I. (2010) · Physics in Medicine and Biology · doi:10.1088/0031-9155/55/16/s05 · Cited by 74
Efficiency of antiscatter grids for flat-detector CT
Kyriakou, Y., & Kalender, W. A. (2007) · Physics in Medicine and Biology · doi:10.1088/0031-9155/52/20/013 · Cited by 68
Absorption and noise in cesium iodide x‐ray image intensifiers
Rowlands, J. A., & Taylor, K. W. (1983) · Medical Physics · doi:10.1118/1.595357 · Cited by 65
X-ray scatter correction for cone-beam CT using moving blocker array
Zhu, L., Strobel, N. K., & Fahrig, R. (2005) · Proceedings of SPIE, the International Society for Optical Engineering/Proceedings of SPIE · doi:10.1117/12.594699 · Cited by 64
Yu, L., Vrieze, T. J., Bruesewitz, M. R., et al. (2010) · American Journal of Roentgenology · doi:10.2214/ajr.09.2951 · Cited by 60
Xu, J., Reh, D. D., Carey, J. P., et al. (2012) · Medical Physics · doi:10.1118/1.4736805 · Cited by 60
Updating of form factor tabulations for coherent scattering of photons in tissues
Tartari, A., Taibi, A., Bonifazzi, C., et al. (2001) · Physics in Medicine and Biology · doi:10.1088/0031-9155/47/1/312 · Cited by 59
X-ray transmission formula for antiscatter grids
Day, G., & Dance, D. R. (1983) · Physics in Medicine and Biology · doi:10.1088/0031-9155/28/12/008 · Cited by 55
Antiscatter grids in mobile C-arm cone-beam CT: Effect on image quality and dose
Schäfer, S., Stayman, J. W., Zbijewski, W. B., et al. (2011) · Medical Physics · doi:10.1118/1.3666947 · Cited by 53
Self-citationAssessment of a New High-Performance Small-Animal X-Ray Tomograph
Vaquero, J. J., Redondo, S., Lage, E., et al. (2008) · IEEE Transactions on Nuclear Science · doi:10.1109/tns.2008.922814 · Cited by 53
Self-citationda Cunha, D. M., Tomal, A., & Poletti, M. E. (2010) · Physics in Medicine and Biology · doi:10.1088/0031-9155/55/15/010 · Cited by 49
Flat panel cone beam computed tomography of the sinuses
Zoumalan, R. A., Lebowitz, R. A., Wang, E., et al. (2009) · Otolaryngology · doi:10.1016/j.otohns.2009.01.010 · Cited by 43
Experimental validation of a rapid Monte Carlo based micro-CT simulator
Colijn, A. P., Zbijewski, W. B., Sasov, A., et al. (2004) · Physics in Medicine and Biology · doi:10.1088/0031-9155/49/18/009 · Cited by 42
Self-citationLiu, X., Shaw, C. C., Wang, T., et al. (2006) · Proceedings of SPIE, the International Society for Optical Engineering/Proceedings of SPIE · doi:10.1117/12.656655 · Cited by 37
Monte Carlo evaluation of scatter mitigation strategies in cone‐beam CT
Lazos, D., & Williamson, J. F. (2010) · Medical Physics · doi:10.1118/1.3488978 · Cited by 34
Monte Carlo modeling of coherent scattering: influence of interference
Leliveld, C. J., Maas, J. G., Bom, V. R., et al. (1996) · IEEE Transactions on Nuclear Science · doi:10.1109/23.552743 · Cited by 34
Potential of software-based scatter corrections in cone-beam volume CT
Bertram, M., Wiegert, J., & Rose, G. (2005) · Proceedings of SPIE, the International Society for Optical Engineering/Proceedings of SPIE · doi:10.1117/12.595032 · Cited by 27
Rayleigh scatter in kilovoltage x-ray imaging: is the independent atom approximation good enough?
Poludniowski, G. G., Evans, P., & Webb, S. (2009) · Physics in Medicine and Biology · doi:10.1088/0031-9155/54/22/012 · Cited by 23
Jarry, G., Graham, S. A., Jaffray, D. A., et al. (2006) · Proceedings of SPIE, the International Society for Optical Engineering/Proceedings of SPIE · doi:10.1117/12.653803 · Cited by 18
Computational high-resolution heart phantoms for medical imaging and dosimetry simulations
Gu, S., Gupta, R., & Kyprianou, I. S. (2011) · Physics in Medicine and Biology · doi:10.1088/0031-9155/56/18/005 · Cited by 15
Self-citationDevelopment of a patient-specific two-compartment anthropomorphic breast phantom
Prionas, N. D., Burkett, G. W., McKenney, S. E., et al. (2012) · Physics in Medicine and Biology · doi:10.1088/0031-9155/57/13/4293 · Cited by 15
Liu, X., Shaw, C., Altunbas, M. C., et al. (2005) · Medical Physics · doi:10.1118/1.1998386 · Cited by 10
Badal, A., & Badano, A. G. (2011) · Medical Physics · doi:10.1118/1.3611641 · Cited by 9
Self-citationYang, K., Burkett, G. W., & Boone, J. M. (2012) · Proceedings of SPIE, the International Society for Optical Engineering/Proceedings of SPIE · doi:10.1117/12.911093 · Cited by 8
Dose and scatter characteristics of a novel cone beam CT system for musculoskeletal extremities
Zbijewski, W. B., Sisniega, A., Vaquero, J. J., et al. (2012) · Proceedings of SPIE, the International Society for Optical Engineering/Proceedings of SPIE · doi:10.1117/12.911843 · Cited by 6
Self-citationBootsma, G. J., Verhaegen, F., & Jaffray, D. A. (2012) · Medical Physics · doi:10.1118/1.4735742 · Cited by 2
Ultrafast micro-CT for in vivo small animal imaging and industrial applications
Sasov, A. (2004) · Proceedings of SPIE, the International Society for Optical Engineering/Proceedings of SPIE · doi:10.1117/12.559011 · Cited by 2
Flagged first, then by citations
Research works printed but not linked: 1
- Woodcock E., 1965, Proc. Conference on the Application of Computing Methods to Reactor Problems, 557
Cited by
Sorted by citations
DeepDRR – A Catalyst for Machine Learning in Fluoroscopy-Guided Procedures
Unberath, M., Zaech, J.-N., Lee, S. C., Bier, B., Fotouhi, J., Armand, M., & Navab, N. (2018) · Lecture Notes in Computer Science · doi:10.1007/978-3-030-00937-3_12 · Cited by 145
“A well studied example is variance reduction for scatter correction in cone-beam CT, since scatter is of low frequency [13].”
BackgroundMethod
Metadata
- Authors
- Sisniega, A., Zbijewski, W., Badal, A., Kyprianou, I. S., Stayman, J. W., Vaquero, J. J., & Siewerdsen, J. H.
- Affiliations
- Johns Hopkins University (US)
- United States Food and Drug Administration (US)
- Universidad Carlos III de Madrid (ES)
- Corresponding author
- Zbijewski, W.
- ORCID
- Venue
- Medical Physics
- Publisher
- Wiley
- Published
- 2013-04-25
- Volume
- 40
- Issue
- 5
- Pages
- 051915-051915
- ISSN
- 0094-2405 (ISSN-L)
- 2473-4209
- PMC
- PMC3651212
- OpenAlex
- W2017401048
- Cited by
- 117
- References
- 75
- Citations by year
- 2026: 1 · 2025: 7 · 2024: 10 · 2023: 11 · 2022: 5 · 2021: 7 · 2020: 11 · 2019: 16 · 2018: 6 · 2017: 14 · 2016: 10 · 2015: 11 · 2014: 6 · 2013: 2
- FWCI
- 5.34
- Type
- journal article
- Language
- en
- Open access
- bronze
- doi.org
- Licence
- Locations
- Medical Physics (publishedVersion)
- Europe PMC (PubMed Central) (publishedVersion)
- PubMed Central (publishedVersion)
- PubMed (publishedVersion)
- LA Referencia (Red Federada de Repositorios Institucionales de Publicaciones Científicas) (acceptedVersion, other-oa)
- e-Archivo (Carlos III University of Madrid) (submittedVersion)
- Funders
- National Institutes of Health (2R01‐CA‐112163, 2R01-CA-112163)
- Spanish Ministry of Education (TEC2010-21619-C04-01)
- AMIT (TEC2011-28972-C02-01, PI11/00616)
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › Advanced X-ray and CT Imaging
- Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Radiation Dose and Imaging
- Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Medical Imaging Techniques and Applications
- Keywords
- cone-beam computed tomography, X-ray scattering, anti-scatter grid, image geometry, air gap, scatter-to-primary ratio, scatter correction, variance reduction, interaction splitting, cupping artifacts, small animal imaging, musculoskeletal imaging, C-arm imaging, breast imaging
- MeSH terms
- Animals, Computer Graphics, Humans, Monte Carlo Method, Scattering, Radiation, Phantoms, Imaging, Cone-Beam Computed Tomography, Signal-To-Noise Ratio
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review