Performance Characteristics of a New LSO PET/CT Scanner With Extended Axial Field-of-View and PSF Reconstruction

Jakoby, Bjoern W., Bercier, Yanic, Watson, Charles C., Bendriem, Bernard, Townsend, David W. · 2009 · IEEE Transactions on Nuclear Science
Cited by 110

The DOI matches its Crossref record; 1 of the 1 references it cites were found in a registry or an index.

Abstract

A new combined lutetium oxyorthosilicate (LSO) PET/CT scanner with an extended axial field-of-view (FOV) of 21.8 cm has been developed (Biograph TruePoint PET/CT with TrueV; Siemens Molecular Imaging) and introduced into clinical practice. The scanner includes the recently announced point spread function (PSF) reconstruction algorithm. The PET components incorporate four rings of 48 detector blocks, 5.4 cm times 5.4 cm in cross-section. Each block comprises a 13 times 13 matrix of 4 times 4 times 20 mm3elements. Data are acquired with a 4.5 ns coincidence time window and an energy window of 425-650 keV. The physical performance of the new scanner has been evaluated according to the recently revised National Electrical Manufacturers Association (NEMA) NU 2-2007 standard and the results have been compared with a previous PET/CT design that incorporates three rings of block detectors with an axial coverage of 16.2 cm (Biograph TruePoint PET/CT; Siemens Molecular Imaging). In addition to the phantom measurements, patient Noise Equivalent Count Rates (NECRs) have been estimated for a range of patients with different body weights (42-154 kg). The average spatial resolution is the same for both scanners: 4.4 mm (FWHM) and 5.0 mm (FWHM) at 1 cm and 10 cm respectively from the center of the transverse FOV. The scatter fractions of the Biograph TruePoint and Biograph TruePoint TrueV are comparable at 32%. Compared to the three ring design, the system sensitivity and peak NECR with smoothed randoms correction (1R) increase by 82% and 73%, respectively. The increase in sensitivity from the extended axial coverage of the Biograph TruePoint PET/CT with TrueV should allow a decrease in either scan time or injected dose without compromising diagnostic image quality. The contrast improvement with the PSF reconstruction potentially offers enhanced detectability for small lesions.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1109/tns.2009.2015764

Cites

1 of 1 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Cited by

See all 110

Metadata

Authors
Jakoby, B. W., Bercier, Y., Watson, C. C., Bendriem, B., & Townsend, D. W.
Venue
IEEE Transactions on Nuclear Science
Published
2009-06
Volume
56
Issue
3
Pages
633-639
ISSN
0018-9499 (ISSN-L)
OpenAlex
W2122964209
Cited by
110
References
1
FWCI
4.93
Type
journal article
Language
en
Open access
closed

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.